Small incision lenticule extraction (SMILE) is a commonly used corneal refractive procedure for correcting myopia and myopic astigmatism. During astigmatism correction, eye rotation between the upright examination position and the supine surgical position may affect the alignment of the astigmatic treatment axis and may contribute to residual astigmatism after surgery.
Manual limbal marking is a commonly used method to guide cyclotorsion adjustment during refractive surgery. OcuLign is an assisted cyclotorsion control function of the VISUMAX 800 platform that helps identify ocular rotation and guide axis alignment during SMILE. However, direct clinical evidence comparing OcuLign-assisted cyclotorsion control with manual limbal marking during SMILE is limited.
This study is a single-center, prospective, randomized, masked, contralateral-eye non-inferiority trial. Adults aged 18 to 40 years with stable myopia and astigmatism who plan to undergo bilateral SMILE surgery will be enrolled. In each participant, one eye will receive SMILE with OcuLign-assisted cyclotorsion control, and the fellow eye will receive SMILE with manual limbal marking-guided cyclotorsion adjustment. The treatment assigned to the right eye will be determined by randomization.
The primary objective is to compare residual refractive astigmatism at 3 months after surgery between the two methods. Secondary outcomes include visual acuity, refractive predictability and stability, astigmatism vector analysis, higher-order aberrations, contrast sensitivity, patient-reported quality of vision, overall satisfaction, and safety outcomes. Participants will be followed for approximately 6 months after surgery.
Trial opening soon.
Get Notified18 year–40 year
All sexes
Interventional
Not applicable
Guangzhou, Guangdong, 510060, China
Accurate alignment of the astigmatic treatment axis is an important factor in the correction of myopic astigmatism during small incision lenticule extraction (SMILE). Cyclotorsion may occur when a patient changes from the upright preoperative examination position to the supine surgical position, and may also be influenced by fixation, head position, and suction during surgery. Inadequate compensation for ocular rotation may reduce astigmatic correction accuracy and contribute to postoperative residual astigmatism.
Manual limbal marking is a commonly used method for cyclotorsion adjustment in corneal refractive surgery. This method is clinically feasible and widely used, but it depends on preoperative marking and intraoperative manual judgment. Marking clarity, ocular surface conditions, patient cooperation, and surgeon experience may affect the accuracy and reproducibility of the adjustment. OcuLign is an assisted cyclotorsion control function of the VISUMAX 800 platform that helps identify ocular rotation and guide axis alignment during SMILE. This study is designed to compare OcuLign-assisted cyclotorsion control with manual limbal marking-guided cyclotorsion adjustment during SMILE.
This trial uses a prospective, randomized, contralateral-eye non-inferiority design. Each participant will undergo bilateral SMILE surgery. One eye will be assigned to OcuLign-assisted cyclotorsion control, and the fellow eye will be assigned to manual limbal marking-guided cyclotorsion adjustment, according to a prespecified randomization sequence. This paired-eye design is intended to reduce inter-individual variability in refractive characteristics, healing response, and subjective visual perception. Apart from the method of cyclotorsion adjustment, surgical parameters, lenticule design, suction procedure, postoperative medication, and follow-up schedule will be kept consistent between the two eyes.
Participants and postoperative outcome assessors will be masked to the eye-level treatment assignment whenever feasible. The operating surgeon cannot be masked because the assigned cyclotorsion adjustment method must be performed during surgery. Postoperative assessments will be conducted at predefined follow-up visits through approximately 6 months after surgery.
The primary comparison will evaluate whether OcuLign-assisted cyclotorsion control is non-inferior to manual limbal marking-guided adjustment in terms of residual refractive astigmatism at 3 months after surgery. Secondary assessments will evaluate visual acuity, refractive predictability and stability, astigmatism vector parameters, higher-order aberrations, contrast sensitivity, patient-reported quality of vision, satisfaction, and safety. Safety monitoring will include loss of corrected distance visual acuity and postoperative adverse events.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
One eye of each participant will undergo SMILE using device-assisted cyclotorsion control with the OcuLign function of the VISUMAX 800 platform. During surgery, OcuLign will assist in identifying ocular rotation and guiding astigmatic axis alignment. Apart from the cyclotorsion adjustment method, the surgical parameters, lenticule design, suction procedure, postoperative medication, and follow-up schedule will be the same as those used for the fellow eye.
Other names: OcuLign-assisted cyclotorsion control, VISUMAX 800 OcuLign
The fellow eye of each participant will undergo SMILE using manual limbal marking-guided cyclotorsion adjustment. Limbal reference marks will be made in the upright preoperative position, and intraoperative axis alignment will be manually adjusted according to the limbal marks and the surgical reference direction. Apart from the cyclotorsion adjustment method, the surgical parameters, surgical procedure, postoperative medication, and follow-up schedule will be the same as those used for the contralateral eye.
Time frame: 3 months after surgery
Residual refractive astigmatism will be measured as the absolute manifest cylinder power in diopters at 3 months after SMILE. The primary analysis will compare residual refractive astigmatism between eyes treated with OcuLign-assisted cyclotorsion control and eyes treated with manual limbal marking-guided cyclotorsion adjustment to evaluate non-inferiority.
Time frame: 1 month, 3 months, and 6 months after surgery
Astigmatic correction predictability will be assessed as the proportion of eyes with postoperative residual manifest cylinder within predefined thresholds, including ≤0.25 D, ≤0.50 D, and ≤1.00 D.
Time frame: 3 months after surgery
Postoperative astigmatic correction will be assessed using the Alpins vector analysis method based on refractive cylinder and axis data. This outcome will be used to describe the overall accuracy of astigmatism correction after surgery.
Time frame: 1 month, 3 months, and 6 months after surgery
Uncorrected distance visual acuity and corrected distance visual acuity will be measured using standardized visual acuity charts and recorded as logMAR visual acuity. Visual acuity-derived performance will also be summarized using the safety index and efficacy index, calculated from the same visual acuity measurements. The safety index is defined as postoperative corrected distance visual acuity divided by preoperative corrected distance visual acuity, and the efficacy index is defined as postoperative uncorrected distance visual acuity divided by preoperative corrected distance visual acuity.
Time frame: 1 month, 3 months, and 6 months after surgery
Postoperative manifest refraction will include sphere, cylinder, and spherical equivalent measured by standard manifest refraction.
Time frame: 1 month, 3 months, and 6 months after surgery
Refractive accuracy will be assessed as the proportion of eyes with postoperative spherical equivalent within predefined ranges of the target refraction, including within ±0.25 D, ±0.50 D, and ±1.00 D.
Time frame: 1 month to 6 months after surgery
Refractive stability will be assessed as the change in manifest spherical equivalent refraction between postoperative follow-up visits.
Time frame: 1 month, 3 months, and 6 months after surgery
Higher-order wavefront aberrations will be measured using wavefront aberrometry under standardized conditions to assess postoperative optical quality.
Time frame: 3 months after surgery
Contrast sensitivity will be measured under standardized conditions using the CSV-1000 contrast sensitivity testing system at multiple spatial frequencies, including 3, 6, 12, and 18 cycles per degree. The area under the log contrast sensitivity function may be calculated as a summary measure when appropriate.
Time frame: 3 months after surgery
Subjective visual symptoms will be assessed using the Quality of Vision questionnaire, which evaluates the frequency, severity, and bothersome nature of visual disturbances such as glare, halos, and double vision. Higher scores indicate worse subjective visual quality.
Time frame: 3 months after surgery
Overall visual satisfaction will be assessed using a 5-point Likert rating scale. Scores range from 1 to 5, with higher scores indicating greater satisfaction.
Zhongshan Ophthalmic Center, Sun Yat-sen University
Other
Comparison of Astigmatic Correction Outcomes Following SMILE With OcuLign-Assisted Cyclotorsion Control and Manual Limbal Marking: A Prospective Randomized Contralateral-Eye Noninferiority Study
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT07512102
Eye Diseases, Myopia
Guangzhou, Guangdong, China
View Trial DetailsNCT07512115
Eye Diseases, Myopia
Guangzhou, Guangdong, China
View Trial DetailsNCT05790928
Eye Diseases, Hyperopia
Tallahassee, Florida, United States
View Trial DetailsNCT04013789
Asthenopia, Eye Diseases
Bloomington, Indiana, United States
View Trial Details