Raja Permaisuri Bainun Hospital
Ipoh, Perak, 30450, Malaysia
NCT Number: NCT03344874
The objective of this study is to compare the relative capability of Stethee® (new wireless digital stethoscope) and 3M™ Littmann® Classic III™ (conventional stethoscope) in the identification of manikin-simulated heart sounds by medical doctors. A randomised, non-blinded, two-period crossover, non-inferiority design will be used to compare the acoustic performance of the two stethoscopes. We aim to determine if Stethee® is as effective as 3M™ Littmann® Classic III™ for auscultation of simulated heart sounds, and whether Stethee® could be recommended for use in clinical practice based on a non-inferior comparison with 3M™ Littmann® Classic III™.
Looking for future studies?
Notify MeAll sexes
Interventional
Not applicable
Ipoh, Perak, 30450, Malaysia
This is a randomised, non-blinded, two-period crossover, non-inferiority study. Stethee®, a new wireless digital stethoscope, is designed to minimise ambient noise associated with electronic stethoscopes. To date, its acoustic capability in comparison to the conventional stethoscope in the auscultation of heart sounds is not known. Therefore, it is important that Stethee®'s capability as a stethoscope with sound amplification be evaluated alongside the conventional stethoscope prior to its use in the clinical setting. As acoustic performance is the primary endpoint, a non-inferiority design is proposed to enable the comparison of the mutual functionalities of the two stethoscopes. The rationale for a non-inferiority design is that: (a) electronic stethoscopes have shown a marginal benefit over conventional stethoscopes in terms of acoustic performance; (b) electronic stethoscopes have several functional advantages over conventional stethoscopes: the ability to analyse heart sounds and record heart sounds, the availability of a database for storage of heart sounds for remote analysis; and (c) electronic stethoscopes were rated better than conventional stethoscopes in terms of the overall quality of auscultation.
The relative capability of Stethee® is compared to that of 3M™ Littmann® Classic III™ by the identification of heart sounds simulated by a manikin, which functions as a surrogate patient. A random sample of non-specialist doctors stratified by health centres in Perak, Malaysia, who meet eligibility criteria and provide written consent to participate, will be randomised to initial testing with Stethee® or 3M™ Littmann® Classic III™ to perform cardiac auscultation on a manikin and identify a set of 10 heart sounds. After a 10-minute break, they will repeat the auscultation of the same 10 simulated heart sounds but played in a different sequence with the alternate stethoscope. Following the completion of both tests, the doctors will be asked to complete a questionnaire that asks about their preference for either Stethee® or 3M™ Littmann® Classic III™ with respect to ease of use, audio clarity, and diagnostic accuracy. At the end of the study, answer sheets and preference questionnaires will be collected for grading and analysis.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Stethee® is a new wireless digital stethoscope.
3MTM Littmann® Classic IIITM is the conventional stethoscope.
Time frame: Measured immediately after the test
The proportion of correctly identified heart sounds between groups will be compared. The test consist of 10 cardiac conditions (represented by 10 manikin-simulated heart sounds). Subjects (medical doctors) are required to auscultate each heart sound and write the complete answer on the answer sheet. The 'all-or-nothing' marking scheme is used, i.e. answers must exactly match the marking scheme.
Time frame: Measured immediately after the test
User preference for the two stethoscopes with respect to ease of use, audio clarity, and diagnostic accuracy, will be compared between groups. This will be estimated as a proportion preferring Stethee® for each domain.
Time frame: Measured immediately after the test
The performance of subgroup of doctors in specialist hospitals and doctors in non-specialist hospitals or primary care clinics will be compared. This will be estimated as a proportion difference of correctly identified heart sounds for each doctor subgroup.
Time frame: Measured immediately after the test
The performance of the two stethoscopes in identifying heart sounds of different levels of difficulty (normal, easy, moderately difficult, difficult) will be compared. This is estimated as a proportion difference of correctly identified heart sounds for each type of heart sound.
Clinical Research Centre, Malaysia
Other
Comparison of Stethee® to the Conventional Stethoscope for Auscultation of Simulated Heart Sounds: A Randomised, Crossover, Non-Inferiority Trial (STELIT)
Acronym: STELIT
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT03188705
Arterial Occlusive Diseases, Arteriosclerosis
Lancaster, Pennsylvania, United States
View Trial DetailsNCT07725445
Arrhythmias, Cardiac, Cardiac Arrhythmias
Seoul, South Korea
View Trial DetailsNCT00241735
Cardiovascular Diseases, Coronary Disease
View Trial DetailsNCT04876963
Arrhythmias, Cardiac, Cardiovascular Diseases
Copenhagen, Denmark
View Trial Details