The primary objective of the study is to demonstrate the accuracy in healthy adults with Vital USA Device by comparison to a non-invasive (auscultatory) reference sphygmomanometer.
The procedure, data collection methods and data analysis that are outlined in the protocol follow the standard: International Standards Organization (ISO) 81060-2:2018 Non-invasive Sphygmomanometers - Part 2: Clinical validation of automated measurement type.
Looking for future studies?
Notify Me18 year and older
All sexes
Interventional
Not applicable
Louisville, Colorado, 80027, United States
This study will include up to 100 adults ≥ 18 years of age. The final subject database will contain no fewer than 85 subjects with a minimum of 255 paired observations. At least 90% of the subjects will contribute 3 paired observations, if necessary the balance of the data will be added from additional subjects.
An interim analysis will be conducted at the end of Part 1 data collection and will consist of simple statistics that will compare the test device tot he reference cuff non-invasive (auscultatory) reference sphygmomanometer.
Part 2 data collection will be conducted after Part 1 data collection interim report is reviewed and additional data collection is approved by the client. It will test an additional 55 subjects (plus extra subjects as replacements if necessary).
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
The end goal is to provide Non-invasive Blood Pressure (NIBP) accuracy data to support validation of the Vital Detect blood pressure monitor.
Time frame: 1 Hour
The primary objective of this study is to provide a validation for non-invasive blood pressure measurement accuracy for the Vital Detect blood pressure monitor when compared to the Reference Sphygmomanometer. Data taken was based on dual observers (Criterion 1(C1), Criterion 2 (C2) for both Systolic and Diastolic determinations. Since the same arm sequential method was used, data for a given subject were averaged to create the reference blood pressure determination against the acceptance criteria. The device under test (DUT) passed or failed the determination based on the mean error and standard deviation of the determination. The acceptance criteria:
Criterion 1:
Mean Error Standard Deviation (Systolic / Diastolic) (Systolic / Diastolic) </= 5.0 mmHg </= 8.0 mmHg
Criterion 2:
Standard Deviation (Systolic) (Diastolic) </= 5.33 mmHg </= 6.09 mmHg
Vital USA, Inc.
Industry
Multiparameter Monitor Non-Invasive Blood Pressure Validation Study
Acronym: NIBP
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT04817137
Blood Pressure
Boston, Massachusetts, United States
View Trial DetailsNCT07713576
Blood Pressure
Chenzhou, Hunan, China
View Trial DetailsNCT06550297
Blood Pressure, Cardiovascular Diseases
New York, United States
View Trial DetailsNCT07317648
Blood Pressure
York, England, United Kingdom
View Trial Details