NCT Number: NCT04360369
Determine if the Tono-Vera Tonometer accurately measures intraocular pressure (IOP). The hypothesis of this test is to confirm the Tono-Vera Tonometer is equivalent to the Goldmann Applanation Tonometer (within +/- 5.0 mmHg).
Looking for future studies?
Notify Me18 year–90 year
All sexes
Interventional
Not applicable
Durham, North Carolina, 27705, United States
The Tono-Vera Tonometer (Reichert, Inc.) is a portable rebound tonometer intended to measure IOP. Subjects will undergo a standard ophthalmologic examination. Subjects will then be measured with four tonometers: Goldmann Applanation Tonometer (C.S.O. SRL), Ocular Response Analyzer G3 (Reichert, Inc.), iCare ic100 tonometer (iCare Finland Oy), and Tono-Vera Tonometer.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Measurement of intraocular pressure (IOP) with Goldmann Applanation Tonometer. Measurement will be used to categorize each subject as having Low IOP (7 to 16 mmHg), Medium IOP (>16 to <23 mmHg), or High IOP (>23 mmHg).
Measurement of IOP with Ocular Response Analyzer G3 and ic100 tonometers.
Measurement of IOP with Tono-Vera Tonometer
Time frame: Through study completion, approximately 4 months.
Subjects will be distributed into Low IOP, Medium IOP, and High IOP groups based on Goldmann Applanation Tonometer measurements. The Tono-Vera Tonometer measurements will be within +/-5.0 mmHg of the Goldmann Applanation Tonometer measurements in all groups.
Time frame: Through study completion, approximately 4 months.
Tono-Vera has two measurement modes: A 6-measurement mode & "Quick" measurement mode. The 6-measurement mode will be used during this study. It will be possible to post-process the results to determine the IOP value that would have been obtained using Quick mode. As such the sponsor will be able to ensure the accuracy of both measurement modes compared to Goldmann Tonometry.
Time frame: Through study completion, approximately 4 months.
The study will collect data on a large number of patients over a wide range of IOP values. The sponsor may use the collected data in a post-hoc analysis to verify that the Tono-Vera Tonometer factory calibration is optimal.
Reichert, Inc.
Industry
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT07604428
Apnea, Eye Diseases
Olomouc, Czechia
View Trial DetailsNCT06689696
Eye Diseases, GLAUCOMA 1, OPEN ANGLE, D (Disorder)
Fullerton, California, United States
View Trial DetailsNCT03057301
Eye Diseases, Glaucoma
Edinburgh, Midlothian, United Kingdom
View Trial DetailsNCT04698876
Behavior, Eye Diseases
Dortmund, Germany
View Trial Details