Comparison Evaluation of Blood Glucose Meter Systems in a Clinical Setting

The purpose of this study was to test the performance of an investigational Bayer Blood Glucose (BG) meter and five additional Blood Glucose Monitoring Systems (BGMSs). The goal was to obtain glucose samples from subjects with diabetes across the meter operating ranges. A capillary sample was collected from each subject at 3 different times during the visit to obtain natural capillary blood samples with a range of glucose concentrations. Some blood samples were glycolyzed to lower the glucose concentration levels and glucose solution was added to other samples to raise glucose concentration levels. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer). All lancings and testing were performed by study staff.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Rainier Clinical Research Center Inc

Renton, Washington, 98057, United States

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Diagnosis of type 1 or type 2 diabetes mellitus
  • 18 years of age or older
  • Willing to follow all study procedures

Exclusion criteria

  • Received an investigational drug or device within the last 30 days
  • Current use or recent exposure to any prescription medication within 24 hours prior to screening that in the opinion of the investigator could have an influence on the subject's ability to participate in the study or on the performance on the test device
  • Positive pregnancy test obtained at screening (test performed for females only, except if surgically sterile or three years past menopause)
  • Hematocrit <32% or >55%
  • Inadequate veins (in the opinion of the investigator or designated study staff) for venous blood withdrawal or intravenous catheter insertion
  • Symptomatic coronary artery disease with a history of angina, or history of a myocardial infarction or coronary intervention (e.g. percutaneous transluminal coronary angioplasty [PTCA], stent placement) or coronary artery bypass graft (CABG) within the past 6 months
  • Cerebrovascular accident within the past 6 months
  • Diagnosis of the following diabetic autonomic neuropathies: symptomatic orthostatic hypotension, heart rate anomalies, gastroparesis
  • Being treated for malignancy, except basal cell or squamous cell skin cancers
  • Major surgical operation within 30 days prior to screening
  • Active alcohol abuse or substance abuse (as judged by the principal investigator)
  • Severe mental illness, mental incapacity, unwillingness or language barriers precluding adequate understanding or cooperation (as judged by principal investigator)
  • Other medical conditions that would pose safety concerns, interfere with study conduct or seriously compromise study integrity, as judged by principal investigator (reason for exclusion will be clearly documented by investigator or designee)
  • Working for a competitive medical device company, or having an immediate family member or someone who is not a family member but is living within the household of someone who works for such a company

Treatment and study plan

Bayer G3 / Tatsu System

Device

Study staff performed Blood Glucose (BG) tests with capillary fingerstick blood collected from subjects with diabetes. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).

Accu-Chek® Aviva Nano Meter/Accu-Chek® Aviva Test Strips

Device

Study staff performed Blood Glucose (BG) tests with capillary fingerstick blood collected from subjects with diabetes. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).

Freestyle Lite® Meter and Test Strips with ZipwikTM tabs

Device

Study staff performed Blood Glucose (BG) tests with capillary fingerstick blood collected from subjects with diabetes. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).

OneTouch® Ultra®2 / OneTouch® Ultra® Blue Test Strips

Device

Study staff performed Blood Glucose (BG) tested with capillary fingerstick blood collected from subjects with diabetes. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).

One Touch® VerioTM Pro/One Touch® VerioTM Test Strips

Device

Study staff performed Blood Glucose (BG) tests with capillary fingerstick blood collected from subjects with diabetes. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).

Truetrack® Meter/Truetrack® Test Strips

Device

Study staff performed Blood Glucose (BG) tests with capillary fingerstick blood collected from subjects with diabetes. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).

Primary outcomes

  1. MAD Mean Absolute Value of the Difference Between Blood Glucose Meter (BGMS) Results and Corresponding YSI Blood Glucose (BG) Results in the Low BG Range(<70 mg/dL)

    Time frame: 6 hours

    Using samples with BG < 70 mg/dL, the Mean Absolute Value of the Differences Between BGM system readings and the YSI laboratory reference values were compared. MAD was calculated from the sum of all |(BG meter)-(BG reference)| assessments, divided by the number of assessments. Each evaluable sample was tested on all 6 BGMS, thus the same number of BG test results was analyzed for each BGMS intervention. Lower MAD value indicates smaller difference between meter value and the reference value. Higher MAD value indicates larger difference between meter value and the reference value.

Secondary outcomes

  1. MAD Mean Absolute Value of the Difference Between Blood Glucose Meter (BGMS) Results and Corresponding YSI Blood Glucose (BG) Results in the BG Range 70 to 180 mg/dL

    Time frame: 6 hours

    Using samples with BG 70 mg/dL to 180 mg/dL, the Mean Absolute Value of the Differences Between BGM system readings and the YSI laboratory reference values were compared. MAD was calculated from the sum of all|(BG meter)-(BG reference)| assessments, divided by the number of assessments. Each evaluable sample was tested on all 6 BGMS, thus the same number of BG test results was analyzed for each BGMS intervention. Lower MAD value indicates smaller difference between meter value and the reference value. Higher MAD value indicates larger difference between meter value and the reference value.

  2. MAD Mean Absolute Value of the Difference Between Blood Glucose Meter (BGMS) Results and Corresponding YSI Blood Glucose (BG) Results in the BG Range >180 mg/dL

    Time frame: 6 hours

    Using samples with BG >180 mg/dL, the Mean Absolute Value of the Differences Between BGM system readings and the YSI laboratory reference values was compared. MAD was calculated from the sum of all |(BG meter)-(BG reference)| assessments, divided by the number of assessments. Each evaluable sample was tested on all 6 BGMS, thus the same number of BG test results was analyzed for each BGMS intervention. Lower MAD value indicates smaller difference between meter value and the reference value. Higher MAD value indicates larger difference between meter value and the reference value.

  3. Standard Deviations of BGMS Differences (Between BGM Meter Readings and the YSI Laboratory Reference Values Across the BG Range of All Evaluable Samples 21 mg/dL to 496 mg/dL)

    Time frame: 6 hours

    Variability of each meter system was determined by calculating the Standard Deviation derived from each meter's differences between Blood Glucose Meter (BGMS)results and corresponding YSI Blood Glucose (BG) results.

  4. Number of Subject Responses 'Strongly Agree' 'Agree' or 'Neutral' With Questionnaire Statements

    Time frame: 1 hour

    Subjects responded to statements about meter accuracy and diabetes management. Subject responses: 'Strongly Agree' 'Agree' 'Neutral' 'Disagree' or 'Strongly Disagree'or No Response. □1ACCURACY HELPS 1A-with my ability to talk with my HCP. 1B-my satisfaction with my self monitoring of diabetes. 1C-with my ability to manage my diabetes. 1D-prevent low BG. 1E-understand how food/exercise affects low BG. □2 I WOULD USE ONLY the meter and strips my insurance company pays for, even if a more accurate meter was available. □3 I USE MY CURRENT METER BECAUSE 3A-my HCP gave it to me. 3B-my insurance company covers the strips. 3C-I think it is the most accurate meter. □4 I WOULD SWITCH meters for a more accurate meter. □5BEING ABLE TO APPLY MORE BLOOD IS IMPORTANT 5A-as I have wasted test strips by not having enough blood to fill the strip 5B-as this would save test strips 5C-I would switch to a meter with this feature

Sponsors and collaborators

Lead sponsor

Ascensia Diabetes Care

Industry

Registry information

Official study title

Clinical Evaluation of Bayer G3 / Tatsu System and BGMs From Other Companies

Important dates

Study start
2012
Primary completion
2012
Study completion
2012
First posted
Jun 8, 2012
Registry last updated
Feb 29, 2016

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)