NCT Number: NCT04030416

Comparing Medical Body Composition Analyzer Seca mBCA 555 With Seca mBCA 515

The mBCA 515 was compared to the gold standard methods MRI, DXA, ADP, D2O- and NaBr-dilution in previous clinical investigations (BCA-01, BCA-02, BCA-20) to generate prediction equations for calculating body composition (FFM, SMM, ECW, TBW and VAT) in different populations. Furthermore, normal ranges for body composition were generated based on the previous study BCA-03 conducted at the UKE.

The mBCA 555 is a new body composition analyzer measuring bioimpedance in standing position developed by seca.

In order to apply prediction equations and normal ranges from previous clinical investigations BCA-01, BCA-02, BCA-03 and BCA-20 to the new mBCA 555 it shall be compared to the mBCA 515.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Observational

Primary location

University Medical Center Hamburg-Eppendorf - Institute of Transfusion Medicine

Hamburg, 20246, Germany

About this study

The use of bioelectrical impedance analysis (BIA) is widespread both in healthy subjects and patients. BIA allows the determination of the fat-free mass (FFM), skeletal muscle mass (SMM), fat mass (FM), extracellular water (ECW) and total body water (TBW) when using appropriate population, age or pathology-specific BIA equations and established procedures. BIA equations are based on validation studies where BIA is compared with reference methods such as Air Displacement Plethysmography (ADP), Dual-Energy X-Ray Absorptiometry (DXA), Magnetic Resonance Imaging (MRI) and isotope dilution.

The predicate device seca mBCA 515 was compared to the gold standard methods MRI, DXA, ADP, D2O- and NaBr-dilution in previous clinical investigations in Kiel (Germany) and New York City (USA) to generate prediction equations for calculating body composition (FFM, SMM, ECW, TBW and VAT) in different populations. Furthermore, normal ranges for body composition were generated in a previous study conducted at the University Hospital Hamburg-Eppendorf.

The mBCA 555 is a new body composition analyzer measuring bioimpedance in standing position developed by seca.

Primary objective is to test whether the mBCA 515 prediction equations can be applied to the new mBCA 555. If necessary, to proof substantial equivalence, prediction equations should be adjusted for the mBCA 555.

Secondary objective is to compare Impedance, phase angle, reactance and resistance for different frequencies in order to clarify if normal ranges for BIVA and phase angle generated with the mBCA 515 need to be adjusted for the new mBCA 555.

Bioimpedance was measured using the seca mBCA 555 and mBCA 515. Body height was measured using the seca 274.

Who can participate

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • male and female subjects,
  • 18 years and older,
  • subjects need to be able to sign the informed consent and privacy policy

Exclusion criteria

  • acute disease
  • pregnancy
  • intake of diuretics
  • edema diagnosed by inspection and palpation of lower limbs
  • paralysis e.g. after a stroke
  • neurodegenerative diseases e.g. ALS
  • tumors in treatment
  • amputation
  • electronic implants e.g. pacemaker
  • probands who cannot provide an ICF by themselves
  • probands who might be dependent from the sponsor or the investigation site
  • current alcohol abuse
  • active prostheses
  • electronic life-support systems, e.g. artificial heart, artificial lung
  • portable electronic medical devices, e.g. ECG devices or infusion pumps
  • metallic implants

Treatment and study plan

Primary outcomes

  1. SMM

    Time frame: 10min

    Skeletal muscle mass, bias of mBCA 555 in comparison to mBCA 515

Secondary outcomes

  1. Phase Angle

    Time frame: 10 minutes

    Phase angle at 50 kHz, bias of mBCA 555 in comparison to mBCA 515

Other outcomes

  1. R50RS

    Time frame: 5 minutes

    resistance right body side, 50 kHz, bias of mBCA 555 in comparison to mBCA 515

  2. R5RS

    Time frame: 5 minutes

    resistance right body side, 5 kHz, bias of mBCA 555 in comparison to mBCA 515

Sponsors and collaborators

Lead sponsor

Seca GmbH & Co. Kg.

Industry

Registry information

Official study title

Clinical Accuracy Validation of the Medical Body Composition Analyzer Seca mBCA 555 in Comparison to Seca mBCA 515

Important dates

Study start
2019
Primary completion
2019
Study completion
2019
First posted
Jul 24, 2019
Registry last updated
Feb 14, 2022

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.