Terumo Nanopass® 34G pen needle
DeviceInsulin pen needle
NCT Number: NCT03878745
This is a subject single blinded, block randomized, prospective, single-visit, multi-center study to compare user experiences with BD Nano™ PRO pen needle vs. the thinner commercially available Terumo Nanopass® pen needle. The study will include a minimum of 55 Japanese American study subjects with Type 1 or Type 2 diabetes.
Looking for future studies?
Notify Me18 year–75 year
All sexes
Interventional
Not applicable
Mills-Peninsula Medical Center Dorothy L. and James E Frank Diabetes Research Institute, San Mateo, California, United States
Study conduct will consist of one 60 to 120-minute Site visit in which pre-set doses of saline will be abdominally delivered by Subjects via a reusable insulin pen device. All pen needles will be attached by Study Staff and pen needle outer cover and inner shield will be removed for subjects. Subjects are to perform 12 injections into the abdomen (6 pairs of injections). Pairs of injections will be evaluated and each pair will contain one BD Nano™ PRO and one Terumo Nanopass® pen needle. The order of the two pen needles in each pair will be randomized.
The intent of this comparative use study is to determine whether Japanese American patients' experiences are different when using the BD Nano™ PRO 4mm x 32G extra thin wall, 5-bevel pen needle vs the Terumo Nanopass® 4mm x 34G pen needle in the Japanese American population. These experiences include the perceived force to deliver dose, the ability to deliver the full dose (measured by leakage), injection pain and bending.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Insulin pen needle
Insulin pen needle
Time frame: Scores were collected immediately after each paired injection
Injection pain as measured by a relative 150mm visual analog scale. Injection pain as measured by a relative visual analog scale. s. User preference is assessed through a single question reported on a 150mm relative VAS scale with the BD Nano™ PRO pen needle Pen labeled at +75 mm and the Comparator pen needle labeled at -75mm. On this scale, zero represents no preference to either pen needle. Relative VAS scores range from -75mm to 75mm; positive scores reflect preference forBD Nano™ PRO and negative scores reflect preference for the comparator.
Scores from each of the paired injections (330 pairs for all participants) were pooled and a mean was calculated. The two-sided 95% confidence interval was calcuated for the average relative rating. A linear model was used to adjust for the order effect.
If the lower bound of the 95% CI is > -10cm, non-inferiority could be concluded. If the lower bound of the 95% CI is >0, superiority can also be concluded.
Time frame: Scores were collected immediately after each paired injection
Subject perceived force required to deliver dose measured by a relative 5 point Likert scale. The scale ranged from -2 to 2, where positive scores indicated less thumb force required for the BD Nano™ PRO, and negative scores indicated less thumb force for Terumo Nanopass®. Scores from each of the paired injections (330 pairs for all participants) were pooled and a mean was calculated.
Time frame: Immediately after injections were completed
Participants reported a visual score of needle bending. A score of at least 2, corresponding to >10 degrees of bending, was considered an event of a bent needle. The percentage of of bending events (yes/no) was calculated for each groups and the mean difference calculated along with a 95% confidence interval.
Time frame: Immediately after injection
After saline delivery equivalent to 30U of U100 insulin (0.3mL) and subject removal of pen needle from body, study staff used the provided materials and scale to absorb leakage from the pen needle tip and injection site to measure the amount of leakage. Leakage over 0.015g was counted as an event. The percentage of of leakage events (yes/no) was calculated for each groups and the mean difference calculated along with a 95% confidence interval.
Time frame: Immediately after injections were completed
Time from when injection button fully depressed to when pen needle removed from body. Mean total injection time was calculated for both the BD Nano™ PRO and the Terumo Nanopass®. Scores from each injection (330 for each needle type) were pooled and a mean was calculated. The total injection time for the BD Nano™ PRO and Terumo Nanopass® were compared and the mean difference reported with a 95% confidence interval.
Time frame: Immediately after injections were completed
Needle breakage was defined as the patient-end metal cannula being separated into two pieces. The percentage of of breaking events (yes/no) was calculated for each groups and the mean difference calculated along with a 95% confidence interval.
Becton, Dickinson and Company
Industry
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT00837122
Cardiovascular Diseases, Diabetes
Accra, Ghana
View Trial DetailsNCT03899883
Diabetes, Diabetes Complications
Aurora, Colorado, United States
View Trial DetailsNCT00065676
Body Weight, Diabetes
Bethesda, Maryland, United States
View Trial DetailsNCT00862433
Body Weight, Diabetes
Bethesda, Maryland, United States
View Trial Details