Comparative Study of TetraGraph and Mechanomiograph Neuromuscular Monitors in Clinical Settings

Background: Neuromuscular monitoring plays a critical role in reducing postoperative residual neuromuscular blockade (PRNB), a significant risk factor for respiratory complications. Despite the availability of various monitoring techniques, the validation of newer devices remains an ongoing challenge. This study compares the performance of the electromyography (EMG)-based Tetragraph neuromuscular monitor with the gold standard mechanomyography (MMG) device, focusing on their accuracy and reliability in clinical settings.

Methods: Twenty cases were conducted during general surgeries requiring neuromuscular relaxation. Ulnar nerve was stimulated via the Tetragraph which detected the compound muscle action potential (CMAP) of adductor pollicis muscle. Simultaneously on the same arm the isometric force of the same stimulated muscle was registrated by the MMG and displayed in the Labchart 8 program. Bland-Altman analysis was used to describe the agreement between devices during distinct phases of neuromuscular blockade. The primary endpoint of the study was the comparison of TOF values of MMG and EMG during induction. In recovery, two groups were made from TOFRs: below and above the recommended muscle recovery to exclude PRNB (TOFR≥90%) (Fuchs-Buder 2023). Additionally, in deeper neuromuscular blockade Train-of-Four Count (TOFC), and Post-Tetanic Count (PTC) values were also analysed.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Observational

Primary location

University of Debrecen, Department of Anesthesiology and Intensive Care

Debrecen, Hajdú-Bihar, 4030, Hungary

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Age 18 or over.
  • ASA (American Society of Anesthesiology) physical status I-III.
  • Informed written consent.
  • Surgery requires the use of a moderate-duration muscle relaxant.

Exclusion criteria

  • History of neuromuscular disease (e.g. stroke with hemi symptoms, myasthenia gravis, severe polyneuropathy).
  • Medication affecting neuromuscular transmission.
  • Open wound or rash due to electrode position
  • Expected difficult intubation.
  • Pregnancy, breast-feeding.
  • Implanted pacemaker

Treatment and study plan

Mechanomiograph: The "gold standard" of quantitative monitoring

Device

The "gold standard" of quantitative monitoring is mechanomyography (MMG), which measures the force of contraction of the adductor pollicis muscle following ulnar nerve stimulation. MMG responses are accurate and reproducible. However, the complex and bulky devices are only suitable for research use, are not suitable for clinical use and are not commercially available.

Electromiograph

Device

Another type of monitor is the electromyograph, which can be seen as an alternative to mechanomyography because of its accuracy. This technique measures muscle activity as a summation of the action potentials of muscle fibres. This activity is proportional to the strength of the muscle contraction. The most commonly innervated nerve is also the ulnar nerve, which innervates the abductor digiti minimi and the first dorsalis interosseus muscles. During the measurements, electrical signals from these muscles are detected. EMG has several advantages over other monitoring techniques. It does not require immobilisation of the hand, thumb immobility is not a problem, no preload is required, and hypothermia does not affect the hand as much as other neuromuscular monitors. The TetraGraph (Senzime AB, Sweden) is a portable EMG-based neuromuscular minitor for which the manufacturers received marketing approval from the U.S. Food and Drug Administration (FDA) in 2019.

Primary outcomes

  1. concordance of the two devices

    Time frame: Intraoperativ

    The primary endpoint of the study is the comparison of TOF values measured by mechanomyography and electromyography, and the examination of the measurement concordance of the two devices.

Sponsors and collaborators

Lead sponsor

University of Debrecen

Other

Registry information

Acronym: EMGvsMMG

Important dates

Study start
2024
Primary completion
2024
Study completion
2024
First posted
Jan 24, 2025
Registry last updated
Jan 24, 2025

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.