Mayo Clinic
Rochester, Minnesota, 55905, United States
NCT Number: NCT02542761
The purpose of this study was to compare the functional performance of individuals with transtibial amputation using two types of prosthetic foot designs: carbon fiber vs. fiberglass composite.
Looking for future studies?
Notify Me21 year and older
All sexes
Interventional
Not applicable
Rochester, Minnesota, 55905, United States
Recently, a new type of prosthetic foot has appeared on the market. This device is composed of a fiberglass composite material. Knowledge is lacking regarding the performance characteristics of this new device. Comprehensive studies are needed to form a solid basis for prosthetic prescription. The current study sought to understand the experience of community-living, transtibial amputees using this prosthetic foot. Specifically, the biomechanical performance of this device were compared to existing conventional dynamic elastic response (DER) technology in a controlled laboratory setting. The investigators hypothesized that the fiberglass composite material provided more energy return and improved ankle kinematics performance.
The study design was a repeated measures cross-over trial whereby only the prosthetic foot was changed. Each subject was tested using their current carbon-fiber energy storage and return prosthetic foot (CFPF) and the fiberglass composite energy storage and return prosthetic foot (Rush, Ability Dynamics) (FPF). Half of the subjects began the study on the CFPF while the other half began on the FPF. All types of CFPF were used in this study. Each subject was given an acclimation period (about 4 weeks) before testing, which was consistent with other similar studies. The same socket and suspension were used throughout the study in order to eliminate these confounding variables.
A 10 camera, high resolution motion capture system with a set of 51 reflective markers was used to capture whole-body motion. Three-dimensional marker trajectory data was collected at 120 Hz and filtered using a fourth-order Butterworth low-pass filter with a cutoff frequency of 8 Hz. The standard Helen Hayes marker set and some additional markers were applied to the subject. Additional markers included an anterior pylon marker, medial pylon marker, lateral pylon marker, right and left medial calcaneus markers, and right and left lateral calcaneus markers. In addition, left and right medial knee markers were used for establishing the knee joint centers and were then removed for the walking trails. Subjects wore standard laboratory athletic shoes for all walking trials. All of the markers associated with the foot were placed on the outside of the subject's shoes.
Following the application of the reflective marker set, the subject performed tests while walking over level ground at a self-selected and normalized speed as well as up and down a 10 degree inclined ramp. The normalized speed controlled for leg length by normalizing to a Froude (Fr) number of 0.25 where Fr = v^2/gl, and v is the walking speed, g is the gravitational constant, and l is the leg length using the greater trochanter height as leg length. Timing gates were used to control the walking speed. Simultaneously, ground reaction force data was collected from force plates at a sampling rate of 600 Hz. Data from these force plates was time-synchronized with the motion cameras. The ramp had a force plate embedded within the ramp.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
The Rush foot is a fiberglass composite energy storage and return prosthetic foot.
Other names: Rush Foot
All types of currently commercially available carbon fiber energy storage and return prosthetic feet will be considered appropriate.
Other names: Otto Bock Triton, Ossur Variflex, Ossur Variflex EVO, Ossur Reflex Shock, Freedom Renegade, Freedom Pacifica, Freedom Thrive with Vertical Shock, Freedom Highlander, Freedom Agilix
Time frame: After approximate 4 week acclimation period
The peak ankle dorsiflexion is the peak ankle backward flexion or bending when walking.
Time frame: After approximate 4 week acclimation period
Peak ankle plantar flexor moment means the peak force of the movement of the foot in which the foot or toes flex downward toward the sole. The unit of measurement for this variable is Nm/kg = 1 nanometer / (kilogram unit).
Time frame: After approximate 4 week acclimation period
Joint power (P) is the "dot product" of the moment (M) at the joint and the angular velocity (w) of the distal segment with respect to the proximal segment (i.e., P = M · w). Depending on the direction of the moment and the direction of the angular velocity, the power can be positive or negative. If the signs for the moment and angular velocity are both positive or both negative, the power is positive. If the signs for the moment and angular velocity are different, the power is negative. The unit of measurement for this variable is W/kg = watts/kilogram.
Time frame: After approximate 4 week acclimation period
A gait cycle is the period of time for one stride, that is, the time from one event (usually initial foot contact) to the next occurrence of the same event with the same foot. For each leg, the gait cycle can be divided into a stance phase and a swing phase. This variable is measuring the angle of knee flexion during the swing phase.
Time frame: After approximate 4 week acclimation period
A gait cycle is the period of time for one stride, that is, the time from one event (usually initial foot contact) to the next occurrence of the same event with the same foot. For each leg, the gait cycle can be divided into a stance phase and a swing phase. This variable is expressed as a percentage of the gait cycle.
Time frame: After approximate 4 week acclimation period
The PEQ is a self-administered questionnaire composed of nine scales computed from forty-two items (ambulation, appearance, frustration, perceived response, residual limb health, social burden, sounds, utility, well being). Each question uses a visual analog scale format, scored as a continuous numerical variable measured as the distance in millimeters from the left endpoint of the measured from the left (0-100). The 0 side of the scale is very negative (terrible) and the 100 side of the scale is very positive (excellent). Each scale is reported separately, with a higher score indicating more satisfaction with the prosthesis itself or quality of life.
Mayo Clinic
Other
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT07652619
Transtibial Amputation, Transtibial Amputation - Unilateral
Raanana, Israel
View Trial DetailsNCT07385859
Activities of Daily Living, Body Image
Samsun, Atakum, Turkey (Türkiye)
View Trial DetailsNCT07067203
Amputation, Hamstring Muscle Strength
Giza, Egypt
View Trial DetailsNCT06599060
Transtibial Amputation - Unilateral
Beykoz, Istanbul, Turkey (Türkiye)
View Trial Details