Completed

NCT Number: NCT00636077

Comparative Effects Of Dialysate Flow Rate And Membrane Packing On The Performance Of Dialyzers Used For Hemodialysis

The purposes of this study are to determine if the performance of a dialyzer depends on how tightly the hollow fiber membranes are packed in the housing of the dialyzer (the membrane packing density) and if that dependence is a function of the dialysate flow rate. The study will examine how efficiently three different sized molecules pass through a dialyzer membrane at different dialysate flow rates in dialyzers with different membrane packing densities. Transfer of urea, phosphorus and beta-2-microglobulin from blood to dialysate will be measured during clinical hemodialysis using four different dialyzers, each used at three different dialysate flow rates. The data derived from these measurements may provide insight into the importance of membrane packing density as a design parameter for hemodialyzers and if changing the membrane packing density might provide equivalent performance at a lower dialysate flow rate.

Completed

Looking for future studies?

Notify Me

Key information

Age range

19 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

University of Louisville, Kidney Disease Program

Louisville, Kentucky, 40202-1718, United States

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Stable hemodialysis patients dialyzing through a native fistula or Gore-Tex graft. The access must be capable of delivering a stable blood flow of 400 ml/min.
  • Age older than 18 years.
  • Fluid removal requirement less than 3 liters per treatment.

Exclusion criteria

  • Noncompliance with dialysis regimen.
  • Hematocrit less than 28%.
  • Active infection

Treatment and study plan

Polyflux HD-C4 Big

Device

Three consecutive treatments with the HD-C4 Big. During the third treatment, dialyzer clearances of urea, phosphorus and β2-microglobulin will be determined at a blood flow rate of 400 mL/min and dialysate flow rates of 350 mL/min, 500 mL/min and 800 mL/min.

Other names: Gambro Polyflux HD-C4 Big dialyzer, Gambro Revaclear MAX

HD-C4 Small

Device

Three consecutive treatments with the HD-C4 Small. During the third treatment dialyzer clearances of urea, phosphorus and β2-microglobulin will be determined at a blood flow rate of 400 mL/min and dialysate flow rates of 350 mL/min, 500 mL/min and 800 mL/min.

Other names: Gambro Polyflux HD-C4 Small dialyzer, Gambro Revaclear dialyzer

Optiflux 160NR

Device

Three consecutive treatments with the HD-C4 Small. During the third treatment dialyzer clearances of urea, phosphorus and β2-microglobulin will be determined at a blood flow rate of 400 mL/min and dialysate flow rates of 350 mL/min, 500 mL/min and 800 mL/min.

Other names: Fresenius Optiflux 160NR

Optiflux 200NR

Device

Three consecutive treatments with the HD-C4 Small. During the third treatment dialyzer clearances of urea, phosphorus and β2-microglobulin will be determined at a blood flow rate of 400 mL/min and dialysate flow rates of 350 mL/min, 500 mL/min and 800 mL/min.

Other names: Fresenius Optiflux 200NR

Primary outcomes

  1. Effect of Increased Dialysate Flow Rate on KoA for Urea Between 4 Dialyzers With Different Membrane Packing Densities.

    Time frame: During the third treatment with each dialyzer (one time during each trial period week)

    KoA is a constant that describes the efficiency of a dialyzer in removing urea. KoA is determined by surface area of the membrane, the thickness of the membrane and pore size.

  2. Effect of Increased Dialysate Flow Rate on Whole Blood Urea Clearance Between 4 Dialyzers With Different Membrane Packing Densities.

    Time frame: During the third treatment with each dialyzer (one time during each trial period week)

  3. Effect of Increased Dialysate Flow Rate on KoA for Phosphorus Between 4 Dialyzers With Different Membrane Packing Densities.

    Time frame: During the third treatment with each dialyzer (one time during each trial period week)

  4. Effect of Increased Dialysate Flow Rate on Whole Blood Phosphorus Clearance Between 4 Dialyzers With Different Membrane Packing Densities.

    Time frame: During the third treatment with each dialyzer (one time during each trial period week)

  5. Effect of Increased Dialysate Flow Rate on KoA for b2-microglobulin Between 4 Dialyzers With Different Membrane Packing Densities.

    Time frame: During the third treatment with each dialyzer (one time during each trial period week)

  6. Effect of Increased Dialysate Flow Rate on Whole Blood B2-microglobulin Clearance Between 4 Dialyzers With Different Membrane Packing Densities.

    Time frame: During the third treatment with each dialyzer (one time during each trial period week)

Sponsors and collaborators

Lead sponsor

Vantive Health LLC

Industry

Collaborators

  • Baxter Healthcare Corporation
  • Gambro Renal Products, Inc.
  • University of Louisville

Registry information

Official study title

Effect Of Dialysate Flow Rate On Mass Transfer Coefficient - Area (KoA) In Dialyzers With Different Membrane Packing Densities

Important dates

Study start
2008
Primary completion
2008
Study completion
2008
First posted
Mar 14, 2008
Registry last updated
Jul 14, 2025

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.