The purpose of this clinical investigation is to evaluate the safety and feasibility of the ColoSeal ICD System in a prospective, multicenter, single-arm study. The ColoSeal ICD System is intended to be used to protect a damaged segment of colon such as a surgical anastomosis, anastomosis leak, or perforation from contact with fecal flow for up to 14 days. The device will be evaluated in adult patients with rectal and rectosigmoid cancer undergoing a resection with a colorectal anastomosis. Subjects will undergo scheduled resection per the surgeon's preferred technique. Following completion of the colon anastomosis, a leak check will be performed prior to insertion of the ColoSeal ICD Device to confirm no leak is present. The ICD Device will be inserted, advanced and positioned with the ICD Delivery System transanally. The anchor portion is positioned > 5 cm above and ideally 10-20 cm above (more proximal in GI tract) the area requiring protection from fecal flow (the anastomosis). The anchor portion of the device is designed to be placed in healthy bowel and not at the region of damaged bowel where protection is required. After positioning, deployment, and anchoring of the ICD Device the Delivery System is removed. The external portion of the device is further anchored to the subject's skin using sutures and an optional extension tubing may be attached. A collection bag is attached to the end of the device to collect output. The ICD Device will remain in place for 10+/-3 days post-operatively. An anastomosis leak test will be performed prior to device removal. Once no leak is confirmed, the ICD Device is removed from the subject.