WaveCrest LAA occlusion device
DeviceLeft Atrial Appendage Occlusion
NCT Number: NCT02239887
Purpose To establish the safety and efficacy of the Coherex WaveCrest Left Atrial Appendage Occlusion System for left atrial appendage (LAA) closure during treatment of non-valvular atrial fibrillation in patients who are at increased risk for embolic stroke and have an ongoing indication for oral anticoagulation.
Looking for future studies?
Notify Me18 year and older
All sexes
Observational
Primary Safety Endpoint The primary safety endpoint is the incidence of major adverse events up to 45 days post-procedure. Major adverse events are defined as life-threatening incidents as determined by the Clinical Events Committee (CEC) and include events such as device embolization; cardiac injury, re-intervention, and/or device-related surgery; bleeding events such as pericardial effusion requiring drainage, cranial bleeding due to any source, gastrointestinal bleeding; or any other event related to the device or the procedure, which requires treatment.
Primary Efficacy Endpoint The primary efficacy endpoint is the incidence of LAA occlusion at 45 and/or 180 days (+ten days) post-procedure. LAA occlusion will be assessed by transesophageal echocardiography (TEE) color flow Doppler and will be defined as absence of flow in the LAA or color flow Doppler jets less than 3mm.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Candidates for this study must meet ALL of the following criteria
Exclusion criteria
Echocardiographic criteria for exclusion:
Left Atrial Appendage Occlusion
Time frame: 45 Days post procedure
The primary safety endpoint is the incidence of major adverse events up to 45 days post-procedure. Major adverse events are defined as life-threatening incidents as determined by the Clinical Events Committee (CEC) and include events such as device embolization; cardiac injury, re-intervention, and/or device-related surgery; bleeding events such as pericardial effusion requiring drainage, cranial bleeding due to any source, gastrointestinal bleeding; or any other event related to the device or the procedure, which requires treatment.
Time frame: 45 and/or 180 days post procedure
The primary efficacy endpoint is the incidence of LAA occlusion at 45 and/or 180 days (+ten days) post-procedure. LAA occlusion will be assessed by transesophageal echocardiography (TEE) color flow Doppler and will be defined as absence of flow in the LAA or color flow Doppler jets less than 3mm.
Time frame: 180 days and one year post-procedure
The secondary safety endpoint is the incidence of serious adverse events at 180 days and one year post-procedure. Serious adverse events are defined as any events that result in death, are life-threatening, require in-patient hospitalization or prolongation of existing hospitalization, or result in permanent or significant disability/incapacity.
Time frame: intra operative
Coherex Medical
Industry
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.