CI632D
DeviceCI632 cochlear implant with Slim Modiolar electrode including dexamethasone in the electrode within wells (CI632D)
NCT Number: NCT04750642
An evaluation of Cochlear's cochlear implant electrode array which passively elutes dexamethasone for a defined period of time to help reduce inflammatory responses.
Looking for future studies?
Notify Me18 year and older
All sexes
Interventional
Not applicable
Fiona Stanley Hospital, Murdoch, New South Wales, Australia
The primary objectives of the pivotal evaluation is to show the efficacy of the dexamethasone eluting electrode through reduction in electrode impedances as compared to a standard electrode and the improvement of speech recognition from preoperative baseline. The secondary objectives are to assess the benefit-risk balance of a dexamethasone eluting electrode as similar to a standard electrode by comparison of adverse events and speech outcomes.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
CI632 cochlear implant with Slim Modiolar electrode including dexamethasone in the electrode within wells (CI632D)
CI632 cochlear implant with Slim Modiolar electrode with market approval and does not include dexamethasone
Time frame: Six months postoperative
The impedance was measured across the active electrodes of the cochlear implant for each participant. Reduced impedances indicate less fibrotic obstruction caused by trauma associated with the electrode insertion.
Time frame: Six months postoperative
Scores on the CNC Words in Quiet test were expressed as a percentage of words repeated correctly by the participant in a sound booth. A higher score indicates better speech perception performance. This outcome measure included the CI632D group only.
Time frame: Twelve months postoperative
Adverse events were recorded from baseline to 12 months postoperative.
Time frame: Six months postoperative
Scores on the CNC Words in Quiet test were expressed as a percentage of words repeated correctly by the participant in a sound booth. A higher score indicates better speech perception performance.
Time frame: Six months postoperative
Scores from the AzBio Sentences in Quiet test were expressed as a percentage of words repeated correctly by the participant in a sound booth. A higher score indicates better speech perception performance.
Cochlear
Industry
Cochlear Implant With Dexamethasone Eluting Electrode Array (The CI-DEX Study): Pivotal Study
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT06936449
Bilateral Hearing Loss, Ear Diseases
Englewood, Colorado, United States
View Trial DetailsNCT07081542
Bilateral Hearing Loss, Cochlear Implantation
Kissimmee, Florida, United States
View Trial DetailsNCT00023036
Cytomegalovirus Infection, Cytomegalovirus Infections
Bethesda, Maryland, United States
View Trial DetailsNCT02038972
Ear Diseases, Hearing Disorders
Orlando, Florida, United States
View Trial Details