NCT Number: NCT06512077

Clinical Validation of mμSORS for Non-invasive Blood Glucose Detection in Non-diabetic Subjects

This is a single-center, open-label, prospective study. Blood glucose is measured at different time points during oral glucose tolerance test and hypoglucose clamp test in healthy subjects, using both venous plasma and multi-channel microspatial offset Raman scattering spectroscopy (mμSORS). Venous plasma glucose was set as gold standard. The two measurements were collected synchronously so as to calculate the mean absolute relative difference (MARD) and the consensus error grid (CEG). Accuracy of non-invasive blood glucose testing by mμSORS will be in-depth validated specifically in the 3-10 mmol/l glucose range.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Ruijin Hospital

Shanghai, Shanghai Municipality, 200025, China

About this study

In order to clinically verify the accuracy and safety of non-invasive blood glucose measurement by the mμSORS technology specifically in the 3-10 mmol/l glucose range, this study will conduct oral glucose tolerance tests and hypoglucose clamp tests on healthy subjects at different time points during the tests. In oral glucose tolerance tests, there will be 9 points as follows: 0-min and post glucose-load 30-min, 60-min, 90-min, 120-min, 150-min, 180-min, 210-min and 240-min. While in hypoglucose clamp tests, blood glucose will be set at around 3.0-4.0mmol/l for about 1.5-2.0 hours, blood glucose will be detected every 10 minutes.

Who can participate

Healthy volunteers accepted: Yes

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Male or female.
  • Aged>=18 years for OGTT participants or aged between 20 and 40 years for hypoglucose clamp tests participants.
  • Fasting blood glucose (FPG) < 6.1 mmol/L and glycated hemoglobin (HbA1c) < 5.7% during the screening period.
  • There are no scars, obvious pigmentation and other factors that interfere with the detection of the palm skin to be tested.
  • Confirmed as healthy based on the results of physical examination, medical history, vital signs and clinical laboratory tests.
  • Fully understand the nature, significance, possible benefits, possible inconveniences or potential risks of this research, and should also understand the research procedures, be willing to complete the entire research process, and provide written consent form.

Exclusion criteria

  • With diabetes history.
  • Any history of serious clinical diseases such as circulatory system, endocrine system, nervous system, digestive system, respiratory system, hematology, immunology, psychiatry or metabolic abnormality, or any other disease that can interfere with the test results, which the investigator considers significant.
  • Alcohol dependency or drug abuse.
  • Those who have participated in clinical trials of other drugs within 3 months before screening (since the last visit of the previous trial).
  • Those who have used any drugs that affect blood glucose levels (such as insulin, oral hypoglycemic drugs, steroid hormones, thyroxine, etc.) within 28 days before screening;
  • Pregnancy or lactation period.
  • Difficulty in venous blood collection or fainting of needles or blood.
  • Other circumstances that the investigator considers inappropriate to participate in the study.

Treatment and study plan

Multi-channel Microspatial Offset Raman Scattering Spectroscopy (mμSORS) for Noninvasive Blood Glucose Detection

Device

Blood glucose of participants were measured by venous plasma and multi-channel microspatial offset Raman scattering spectroscopy (mμSORS). The two measurements were collected synchronously and analyzed.

Primary outcomes

  1. Mean absolute relative deviation (MARD) of venous plasma glucose and glucose values measured by mμSORS at each time point of OGTT and hypoglucose clamp tests.

    Time frame: 2 months

    Glucose will be measured using both intravenous sampling (plasma) and multi-channel microspatial offset Raman scattering spectroscopy (mμSORS) detection synchronously.

  2. Consensus Error Grid (CEG) for venous plasma glucose and glucose values measured by mμSORS at each time point of OGTT and hypoglucose clamp tests.

    Time frame: 2 months

    Glucose will be measured using both intravenous sampling (plasma) and multi-channel microspatial offset Raman scattering spectroscopy (mμSORS) detection synchronously.

Secondary outcomes

  1. MARD for two measurement methods in different blood glucose ranges.

    Time frame: 2 months

    Plasma glucose ranges 3.9-10.0 mmol/L and <3.9 mmol/L.

  2. Incidence of Treatment-Emergent Adverse Events.

    Time frame: 2 months

    Safety outcomes.

Sponsors and collaborators

Lead sponsor

Shanghai Jiao Tong University School of Medicine

Other

Collaborators

  • Shanghai Photonic View Technology Co., Ltd.

Registry information

Official study title

Clinical Validation of Multi-channel Microspatial Offset Raman Scattering Spectroscopy (mμSORS) for Non-invasive Blood Glucose Detection in Non-diabetic Subjects

Important dates

Study start
2024
Primary completion
2024
Study completion
2024
First posted
Jul 22, 2024
Registry last updated
Jan 28, 2025

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.