Completed

NCT Number: NCT02762565

Clinical Utility of a Low-Cost Hand-Held Breast Scanner

Early detection of breast cancer improves the survival rate and makes treatment less costly. This study would measure the accuracy a low-cost hand-held commercially viable device, iBE, for the detection of clinically relevant findings in the breast using the results of current mammography as a comparison. This research will have no impact on clinical decision making.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Abramson Cancer Center of the University of Pennsylvania

Philadelphia, Pennsylvania, 19104, United States

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • 18 years of age and older
  • Women and men with symptomatic breast lump (either by palpation or imaging) OR
  • Asymptomatic women presenting to the imaging center for a screening mammogram.
  • Signed Informed Consent

Exclusion criteria

  • Patients under 18 years of age
  • Patients who previously participated in this study and are returning to the Women's Imaging Center for follow-up diagnostic tests

Treatment and study plan

PEFS system (Product Name: Intelligent Breast Exam™ or iBE™

Device

Primary outcomes

  1. Number of Quadrants With Detected Breast Lesions

    Time frame: 3 months

    The iBE results were evaluated by definitive quadrants. The definitive quadrants break down the superimposed clock into 4 equal sectors (e.g. 12-3, 3-6, 6-9, 9-12).

  2. Sensitivity of iBE Breast Lesion Detection

    Time frame: 1 day

    the number of quadrants with positive breast lesion detected by iBE confirmed by mammogram and ultrasound, 4 quadrants per breast per participant. 77 quadrants were labeled positive for breast lesion by iBE, 66 were confirmed positive by mammogram or ultrasound allowing calculation of sensitivity of the the iBE breast lesion detection

  3. Specificity of iBE Breast Lesion Detection

    Time frame: 1 day

    the number of quadrants characterized normal breast by iBE confirmed by mammogram and ultrasound, 4 quadrants per breast per participant

Sponsors and collaborators

Lead sponsor

Abramson Cancer Center at Penn Medicine

Other

Registry information

Important dates

Study start
2015
Primary completion
2016
Study completion
2016
First posted
May 5, 2016
Registry last updated
Aug 16, 2021