NCT Number: NCT03614416

Clinical Trial Testing the BioSerenity Pulse Oximeter (EOXY) Efficacy, Performance and Safety in Healthy Volunteer

The purpose of the study is to test the performance and safety of the Bioserenity pulse oxymeter EOXY. For that, and in agreement with ISO60601-2-61 norm and FDA regulation, measures of heart rate, with a gold-standard, and SaO2 measures generated by blood samples will be compared with Sp02 and heart rate measures that are generated by our EOXY.

Thirteen healthy subjects will participate in the study in total. Three subjects will be enrolled during a primary phase .They will participate at a secondary phase as well as ten others subjects. The purpose of the primary phase is to test the body area where the measurement can be done. The purpose of the secondary phase is to measure SpO2 and HR in desaturation conditions.

Completed

Looking for future studies?

Notify Me

Key information

Conditions

Age range

18 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Université catholique louvain la neuve, centre sportif

Louvain-la-Neuve, 1348, Belgium

About this study

The primary objective is to evaluate and document the SpO2 accuracy of the pulse oximeter equipment, evaluating the performance of the device compared to invasive method blood sampling (SaO2) by CO-Oximeter in healthy subjects during invasive controlled desaturation study at different ranges (100 % to 70% SaO2).

Subject safety during desaturation and saturation period, will be evaluated by reviewing vital sign: heart rate (HR) continuously and blood pressure/temperature between each desaturation plateau.

There are two secondaries objectives, the measurement of the variation of the fraction of inspired oxygen (FiO2), and the validation of the algorithm for data review and analysis.

For the study primary phase, the purpose of this phase is to test the body area where the SpO2 is better detected and define 3 measurement areas. For this, three healthy volunteers will be placed in normal O2 saturation, the SpO2 will be measured using the Bioserenity oximeter (noninvasive method) and will be compared to a gold-standard oximeter CE-marked. The investigator will determine the best area in the body where the SpO2 could be taken for the following subjects.

During a second phase, the measurement of Sp02 and HR with EOXY, will be performed in O2 saturation conditions up to 100% and then in desaturation by plateau from 100 % to 70%. 3 EOXY pulse oximeter recorded the SpO2 at the same time in 3 different areas of the body. As required by the standard, Sp02 measures will be simultaneously compared to SaO2 values measured by blood sampling taken from an indwelling arterial catheter. The HR obtained with EOXY will be compared simultaneously to a gold standard oximeter CE-marked.

As the example in the standard, it was choosen to do the test on the 70-100% range, with 5 plateaus (100%-97%, 97%-92%, 92%-85%, 84%-78%, 77%-70%) and to collect five Sp02 and Sa02 values (blood samples) and five HR measures, per plateau, for each of 13 subjects.

The desaturation will be controlled thanks' the FiO2 measures, that should allow to bring subjects near target levels of desaturation plateau.

Who can participate

Healthy volunteers accepted: Yes

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Healthy subject (Man/ Women) > 18 years old
  • Subject signed informed consent prior to any screening procedure
  • Subject in good health (COHb < 3% , MetHb < 2%, CtHb > 10mg/dL)
  • Healthy adult capable of undergoing controlled hypoxemia to the level called for in the protocol with minimal medical risk

Exclusion criteria

  • Smokers or individuals exposed to high levels of carbon monoxide that result in elevation of carboxy hemoglobin levels
  • Individual subject to conditions that result in elevated levels of methemoglobin
  • Subject who would be placed at undue medical risk associated with any procedure called for in the protocol
  • Subject with open wounds
  • Pregnant woman (negative pregnancy test needed)
  • Subject allergic to silicon, polyamide and silver yarn
  • Subject with cardiorespiratory disorders likely to worsen with thoracic pressure applied by the textile band
  • Subject with mental or motor impairment preventing him from expressing pain
  • Subject with blood or skin disorder that may impacting results
  • Subject with behavioral disorders, too agitated or too aggressive
  • Subject with sensorial disorders, insensible to skin pain
  • Subject susceptible to tension/pressure based headaches

Treatment and study plan

EOXY device

Device

Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard

Gold standard oximeter

Device

Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard

SaO2 sampling

Procedure

Sp02 and HR measures from EOXY device are simulateously compared to gold standard measures : Sa02 sampling measures and pulse oximeter gold standard

Primary outcomes

  1. Percentage SPO2 Accuracy Root-Mean Square (ARMS) at Forehead Position

    Time frame: 1 day

    For each desaturation plateau, SpO2 ARMS are calculated by the difference between the investigational pulse oximeter SpO2 measurement and the co-oximeter's SaO2 measurements

  2. Percentage SPO2 Accuracy Root-Mean Square (ARMS) at Torso Position

    Time frame: 1 day

    For each desaturation plateau, SpO2 ARMS (Accuracy Root-Mean Square) are calculated by the difference between the investigational pulse oximeter SpO2 measurement and the co-oximeter's SaO2 measurements

  3. Percentage SPO2 Accuracy Root-Mean Square (ARMS) at BACK Position

    Time frame: 1 day

    For each desaturation plateau, SpO2 ARMS are calculated by the difference between the investigational pulse oximeter SpO2 measurement and the co-oximeter's SaO2 measurements

  4. HR Accuracy Root-Mean Square (ARMS) at Forehead Position

    Time frame: 1 day

    For each desaturation plateau, HR ARMS are calculated by the difference between the investigational pulse oximeter HR measurement and the HR gold standard measurements

  5. Percentage of HR Accuracy Root-Mean Square (ARMS) at TORSO Position

    Time frame: 1 day

    For each desaturation plateau, HR ARMS are calculated by the difference between the investigational pulse oximeter HR measurement and the HR gold standard measurements

  6. Percentage of HR Accuracy Root-Mean Square (ARMS) at BACK Position

    Time frame: 1 day

    For each desaturation plateau, HR ARMS are calculated by the difference between the investigational pulse oximeter HR measurement and the HR gold standard measurements

Sponsors and collaborators

Lead sponsor

BioSerenity

Industry

Collaborators

  • NAMSA
  • Université Catholique de Louvain

Registry information

Acronym: EOXY

Important dates

Study start
2017
Primary completion
2017
Study completion
2018
First posted
Aug 3, 2018
Registry last updated
Jan 12, 2021

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.