Skip to main content
OpenTrials
Completed

NCT Number: NCT01763775

Clinical Trial of Transtek Body Fat Analyzer (GBF-1251-B & Other 3 Models)

The clinical protocol of the clinical testing of this device:

1. Objective of the test: To verify the functions and efficiency of devices. 2. Test methods and procedures: Comparison Test. 3. Device Under Test (DUT): Transtek Body Fat Analyzer, Model: GBF-1251-B, BF-1255-B, BF-1256-B, and GBF-1257-B. 4. Comparison device: Transtek, Glass Body Fat Analyzer, GBF-950-D. 5. Study endpoints: DUT and the comparison device are substantial equivalence. 6. Statistical methodology used: Description of statistical methods.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Wuhou District Hospital

Chengdu, Sichuan, 610041, China

About this study

  • Test Purpose:

The aim of clinical test is to collect weight, body fat, total body water, bone mass, and muscle mass reading value which measured by DUT and reference device.

  • Target Subject:

The subject database shall contain at least 25 male and 25 female subjects and all above 18 years old.

  • Test procedures:
  • Record the name, age, gender, and height of each subject.
  • Let subject seated calmly for two minutes before test, dry the feet.
  • Use device to measure weight, body fat, total body water, bone mass, and muscle mass; record reading value.
  • For each subject, repeat 3) to get 3 pairs measurements.
  • Repeat step 1) to 4) for every device (DUT: GBF-1251-B, BF-1255-B, BF-1256-B, and GBF-1257-B; and reference device: GBF-950-D).
  • Repeat step 1) to 5) for each subject.
  • Note:
  • No motion and speaking are allowed during the measurement.
  • All DUT use the same algorithms and contact the patient at the same body locations (feet), and operate on the same frequency.
  • Test environment: Temperature: 22±1℃; Relative humidity: 40~70%.

Who can participate

Healthy volunteers accepted: Yes

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • male,female

Exclusion criteria

  • below 18 years old

Treatment and study plan

Transtek 125X

Device

Measuring weight, total body water, body fat, muscle mass, and bone mass. Measure each participants by Transtek 125X (GBF-1251-B, BF-1255-B, BF-1256-B, GBF-1257-B).

Transtek 950D

Device

Measuring weight, total body water, body fat, muscle mass, and bone mass. Measure each participants by Transtek 950D (GBF-950-D).

Primary outcomes

  1. BMI (Body Mass Index)

    Time frame: 5 Day

    Body Fat, Total Body Water, Muscle Mass, Bone Mass

Secondary outcomes

  1. Weight

    Time frame: 5 Days

    Body Weight

Sponsors and collaborators

Lead sponsor

Leo Wang

Other

Collaborators

  • West China Hospital

Registry information

Official study title

Verify the Functions and Efficiency of Transtek (Trade Mark) Body Fat Analyzer, GBF-1251-B, BF-1255-B, BF-1256-B, and GBF-1257-B

Acronym: BS-BT

Important dates

Study start
2012
Primary completion
2012
Study completion
2012
First posted
Jan 9, 2013
Registry last updated
Jan 9, 2013

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.