RayOne EMV Toric IOL
DeviceImplantation of intraocular lens.
Other names: Model RAO210T
NCT Number: NCT05985304
The objectives of the clinical investigation are to determine the safety and performance of the RayOne EMV Toric (Model RAO210T) following unilateral implantation and approximately 6 months (120 to 180 days) of post-operative assessment, as a randomized comparison to aspheric monofocal control for low cylinder (1.50 D). This patient population with cataracts and low corneal cylinder (+1.00 to +1.50 D) will have the natural crystalline lens removed and replaced with an intraocular lens in the capsular bag.
Looking for future studies?
Notify Me22 year and older
All sexes
Interventional
Not applicable
Vold Vision, Fayetteville, Arkansas, United States
This study is a prospective, multicenter, randomized, active controlled, masked (assessor and subject) pivotal investigation for unilateral implantation of low cylinder (1.50 D) toric IOL. It compares the Rayner RAO210T toric IOL to the Rayner RAO600C aspheric monofocal IOL. Subjects who sign the ICF are considered enrolled in the study. After signing the ICF, subjects will be screened for eligibility. Inclusion and Exclusion Criteria must be applied prior to subject randomization. Subjects who meet all protocol-specified eligibility criteria will be assigned to the randomized controlled study arms if the study eye's estimated IOL cylinder power using Barrett Toric Calculator is 1.50 D. Randomized subjects receive either the Rayner RAO210T Toric IOL (1.50 D low cylinder) or the Rayner RAO600C aspheric monofocal IOL in the study eye.
Up to 295 adult subjects will be enrolled (consented) assuming a 15% screen failure rate, then up to 250 subjects will be randomized, of which approximately 125 subjects will be randomized to receive low cylinder (1.50 D) RayOne EMV Toric (Model RAO210T) in one eye and approximately 125 subjects randomized to receive the RAO600C aspheric monofocal IOL in one eye, to complete 120-180 days follow-up for at least 100 subjects in each group.
All subjects who meet inclusion criteria will be assigned to the randomized controlled study arms to receive either the RAO210T toric IOL (1.50 D low cylinder) or the RAO600C aspheric monofocal IOL in the study eye according to a 1:1 ratio. Randomization will be stratified by site.
Up to 11 U.S. sites will be encouraged to enroll a minimum of 20 subjects. No site will enroll more than 25% of the subjects enrolled in the study.
If a subject has significant cataract in both eyes, it is recommended that cataract surgery is performed in one eye before the subject is enrolled in the study. Once a subject has been enrolled, it is recommended that the fellow eye does not undergo cataract surgery (except for a YAG capsulotomy) throughout the duration of the study. At screening, if both eyes qualify for the study, the eye to undergo cataract surgery and IOL implantation is the eye with worse pre-operative BCDVA. If pre-operative BCDVA is the same for each eye, the right eye will be the study eye.
Subjects will complete 6 study visits in approximately 9 months. Subject participation is calculated as the difference between the time of the pre-operative visit to completion of Visit 4 (120 to 180 days post-operative).
Standard clinical trial methods will be used to minimize bias, such as the use of site personnel performing manifest refraction and visual acuity assessments masked to subject treatment assignment, masking of subjects in the randomized controlled investigation evaluating low cylinder power, standardized test procedures, common Investigator training and common inclusion and exclusion criteria.
In order to minimize bias, measures will be taken to mask the site personnel performing post-operative manifest refraction and visual acuity assessments, as to the subject's treatment assignment until after the final database lock. Every attempt should be made to have the same masked site personnel perform the same masked post-operative assessments for an individual subject throughout the subject's study participation.
Subjects will be masked to their IOL assignment in the randomized controlled investigation evaluating low cylinder power. All material which may indicate the subjects' assignment, e.g., packaging, documents, etc., will be removed from any areas where subjects and/or masked site personnel may see them. Unmasked personnel will further be instructed to scrupulously avoid conversation and communication with masked personnel, subjects and all other persons regarding subjects' assignments, outcomes, clinical courses, and all other information potentially relevant to the study and its conduct.
Study Groups:
Sample Size Determination:
Subjects will be enrolled with the goal of completing a total of 200 subjects through Visit 4 within the randomized groups, of whom 100 were implanted with the low cylinder RayOne EMV Toric IOL (Model RAO210T) and 100 were implanted with the RAO600C monofocal IOL.
Effectiveness For effectiveness, 100 subjects (eyes) randomized to the Toric (test) group (lowest cylinder power +1.50D) and 100 subjects (eyes) randomized to the Monofocal (control) group completing Visit 4 yields over 98% statistical power to reject H0e in favor of H1e and conclude the Toric IOL has statistically significantly lower mean residual manifest cylinder at Visit 4 compared to the Monofocal IOL, using a one-sided t-test with an alpha level of 0.025 and assuming a true mean difference of 0.4 D and a standard deviation of 0.7 D in both groups.
Safety For safety, ISO 11979-7 and ANSI Z80.30 specifies that a minimum of 100 subjects should complete a clinical evaluation of an IOL, where a parent IOL has been approved, to obtain appropriate specificity around adverse event and visual acuity rates.
Handling of Missing Data:
Missing data will be imputed for selected endpoints using the methods specified under the analysis descriptions for the endpoints. Where possible, multiple imputation will be used. Except where mentioned in Section 7, missing data will not be imputed.
Multiplicity Considerations:
The overall type I error rate for the toric effectiveness analyses will be controlled at 0.05. The study will be considered successful if all of the primary effectiveness and safety endpoints are met. Thus, no adjustments for multiplicity are necessary.
All primary effectiveness and primary safety endpoints as described below are required to achieve successful outcomes in order to demonstrate study success.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Acceptable methods include at least one of the following: intrauterine (intrauterine device), hormonal (oral, injection, patch, implant, ring), barrier with spermicide (condom, diaphragm), or abstinence.
Exclusion criteria
Implantation of intraocular lens.
Other names: Model RAO210T
Implantation of intraocular lens.
Other names: RAO600C
Removal of natural crystalline lens due to cataracts
Time frame: 120 to 180 days post-operatively
Mean magnitude of residual manifest cylinder (as measured by manifest refractive cylinder).
Time frame: 120 to 180 days post-operatively
The adjusted mean difference of residual manifest cylinder (as measured by manifest refractive cylinder) between Toric and Monofocal groups will be given together with a 95% confidence interval and associated one-sided p-value. This endpoint will be considered successful if the one-sided p-value ≤ 0.025.
Time frame: 120 to 180 days post-operatively
Frequency and percentage of RayOne EMV Toric (Model RAO210T) IOLs with IOL axis misalignment at Visit 4 (as determined by photographic method) less than 10 degrees.
Time frame: 120 to 180 days post-operatively
Frequency and percentage of RayOne EMV Toric (Model RAO210T) IOLs with IOL axis misalignment at Visit 4 (as determined by photographic method) less than 20 degrees.
Time frame: 30 to 60 days and 120 to 180 days post-operatively
Stability of toric IOL axis orientation, expressed as frequency and percentage of RayOne EMV Toric (Model RAO210T) IOLs that rotate ≤ 5 degrees postoperatively between 30 to 60 days (Visit 3) and 120 to 180 days (Visit 4)
Time frame: 120 to 180 days post-operatively
Rates of IOL adverse events through 120 to 180 days post-operatively (Visit 4) compared to the ISO Safety and Performance Endpoint (SPE) rates as described in ISO 11979-7.
Rates of cumulative and persistent adverse events from ISO 11979-7 (2018) Table E.2 will be reported by treatment group (Toric test group, Monofocal control group) and AE type. Rates of adverse events will be compared to the SPE rate from Table E.2. The one-sided p-value from the exact binomial test of each hypothesis will also be provided for each adverse event. If none of these hypothesis test results are statistically significant, this endpoint will be considered successful for the toric IOL.
Time frame: 120 to 180 days post-operatively
The eyes with at least one adverse event not included in the cumulative and persistent adverse events from ISO 11979-7 (2018) Table E.2 through 120 to 180 days post-operatively (Visit 4) will be presented by treatment group (Toric test group, Monofocal control group).
Time frame: 120 to 180 days post-operatively
The percentage of eyes implanted with the Toric test group IOL requiring a secondary surgical intervention for IOL repositioning due to IOL misalignment through 120 to 180 days post-operatively (Visit 4), as well as a two-sided 95% exact binomial confidence interval for the percentage, will be presented.
Time frame: 120 to 180 days post-operatively
The percentage of eyes that achieve a BCDVA at 4 meters of 0.30 logMAR or better at 120 to 180 days post-operatively (Visit 4) will be presented by treatment group (Toric test group, Monofocal control group) for the Safety Set and Best Case Set. If neither of these hypothesis test results are statistically significant, this endpoint will be considered successful for the toric IOL.
Time frame: 120 to 180 days post-operatively
Residual manifest refractive cylinder by subgroups of 0.25 D preoperative keratometric cylinder at 120 to 180 days post-operatively
Time frame: 120 to 180 days post-operatively
Percent reduction in absolute cylinder (as measured by residual manifest cylinder / preoperative keratometric cylinder), at 120 to 180 days post-operatively. Positive numbers represent an decrease in cylinder and negative numbers represent an increase in cylinder.
Rayner Intraocular Lenses Limited
Industry
Clinical Trial to Investigate the Safety and Effectiveness of a Hydrophilic EMV Toric Lens RAO210T in the Correction of Aphakia and Post-operative Corneal Astigmatism
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT04890249
Astigmatism, Cataract
Burwood, New South Wales, Australia
View Trial DetailsNCT03791619
Astigmatism, Cataract
Bakersfield, California, United States
View Trial DetailsNCT04652037
Astigmatism, Behavior
Montreal, Quebec, Canada
View Trial DetailsNCT05510869
Astigmatism, Cataract
View Trial Details