Optometry Research Group (GIO) - Optics Department, University of Valencia
Valencia, Spain
NCT Number: NCT01965288
The aim of this study is to compare the clinical performance of the Biofinity Toric (comfilcon A) contact lens, versus Air Optix for Astigmatism (lotrafilcon B) over 1 month of daily wear.
Looking for future studies?
Notify Me18 year–40 year
All sexes
Interventional
Not applicable
Valencia, Spain
Single masked, open label, randomized, bilateral, crossover study design comparing comfilcon A against lotrafilcon B lenses over 1 month of daily wear for each study lens. Each subject will be randomized to wear the test lens (comfilcon A) or the control lens (lotrafilcon B) for one month of daily wear before repeating the schedule for the second pair without a washout period.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
A person is eligible for inclusion in the study if he/she:
Exclusion criteria
A person will be excluded from the study if he/she:
test lens Pair(comfilcon A) or the control lens Pair(lotrafilcon B)
Other names: Biofinity Toric
test lens Pair(comfilcon A) or the control lens Pair(lotrafilcon B)
Other names: Air Optix
Time frame: Baseline
Participant rating of lens Daily and Comfortable Wearing Time. Collected at baseline for all habitual lenses. (The hours of average comfortable wearing time and average daily wearing time.)
Time frame: Baseline
Participant rating of lens Comfort, Dryness, Handling, Lens Fit Stability and Vision Satisfaction. Collected at baseline for all habitual lenses. (0-10; Comfort, Lens Fit and Satisfaction / 0=very poor,10=excellent; Dryness / 0=very dry, 10=no dryness; Handling / 0=very difficult, 10=very easy for handling)
Time frame: Baseline
Participant rating of vision quality on insertion, during the day, end of day. Collected at baseline for all habitual lenses. (0-100; 0=extremely poor vision totally blurred, 100=excellent vision totally sharp)
Time frame: Baseline
Participant rating of vision stability on insertion, during the day, end of day. Collected at baseline for all habitual lenses. (0-100; 0=totally unstable fluctuating/changing, 100=perfectly stable not fluctuating/changing)
Time frame: Baseline
Participant rating for overall sensation of moistness. Collected at baseline for all habitual lenses. (5-point Likert Scale; Excellent, Good, Average, Below Average)
Time frame: Baseline
Participant rating for overall sensation of smoothness. Collected at baseline for all habitual lenses. (5-point Likert Scale; Excellent, Good, Average, Below Average)
Time frame: Baseline
Participant rating for comfort satisfaction. Collected at baseline for all habitual lenses. (4-point Likert Scale; Completely Satisfied, Somewhat Satisfied, Somewhat Dissatisfied, Completely Dissatisfied)
Time frame: Baseline
Participant rating for dryness satisfaction. Collected at baseline for all habitual lenses. (4-point Likert Scale; Completely Satisfied, Somewhat Satisfied, Somewhat Dissatisfied, Completely Dissatisfied)
Time frame: Baseline
Participant rating for handling satisfaction. Collected at baseline for all habitual lenses. (4-point Likert Scale; Completely Satisfied, Somewhat Satisfied, Somewhat Dissatisfied, Completely Dissatisfied)
Time frame: Baseline
Participant rating for lens fit satisfaction. Collected at baseline for all habitual lenses. (4-point Likert Scale; Completely Satisfied, Somewhat Satisfied, Somewhat Dissatisfied, Completely Dissatisfied)
Time frame: Baseline
Participant rating for vision satisfaction. Collected at baseline for all habitual lenses. (4-point Likert Scale; Completely Satisfied, Somewhat Satisfied, Somewhat Dissatisfied, Completely Dissatisfied)
Time frame: Baseline
Participant rating for overall satisfaction. Collected at baseline for all habitual lenses. (4-point Likert Scale; Completely Satisfied, Somewhat Satisfied, Somewhat Dissatisfied, Completely Dissatisfied)
Time frame: Dispense
Participant rating of comfort upon insertion. Collected at dispense for each lens. (0-10; 10=Can't Feel)
Time frame: Dispense
Participant rating of vision satisfaction upon insertion. Collected at dispense for each lens. (0-10; 10= Very Satisfied)
Time frame: Dispense
Participant rating of Vision Quality with contact lens prescription. Collected at dispense for each lens. (0-100; 0=extremely poor vision totally blurred, 100=excellent vision totally sharp)
Time frame: Dispense
Participant rating of vision stability on insertion. Collected at dispense for each lens. (0-100; 0=extremely poor vision totally blurred, 100=excellent vision totally sharp)
Time frame: Dispense
Assessment of monocular and binocular high and low contrast visual acuity (VA). Collected at dispense for each lens. logMAR (VA).
Time frame: Dispense
Assessment of Lens Fit Performance for centration. Collected at dispense for each lens. (Biomicroscopy; centered or slightly decentered)
Time frame: Dispense
Assessment of Lens Fit Performance for corneal coverage. Collected at dispense for each lens. Corneal coverage assessed in primary gaze: (yes=full corneal coverage at all times, no=incomplete corneal coverage)
Time frame: Dispense
Assessment of Lens Fit Performance for post blink movement. Collected at dispense for each lens. (0-4, 0.5 increments; 0=Insufficient, unacceptable movement, 4= Excessive, unacceptable movement)
Time frame: Dispense
Assessment of Lens Fit Performance for lens orientation in primary position of gaze. Collected at dispense for each lens. (Degree of mislocation relative to lens axis mark.)
Time frame: Dispense
Assessment of Lens Fit Performance for lens marking visibility. Collected at dispense for each lens. (1-3, 1=excellent, 2=average, 3=poor)
Time frame: Dispense
Assessment of Lens Fit Performance for lens rotational stability on blink. Collected at dispense for each lens. (No rotation and 5-10 degrees rotation from axis location mark)
Time frame: Dispense
Assessment of Lens Fit Performance for lens to stabilize in 5-10 min. Collected at dispense for each lens. (Varied less than 5 degrees from lens marking location between 5-10 min)
Time frame: Dispense
Assessment of Lens Fit Performance for overall lens stability. Collected at dispense for each lens. (Excellent or Good)
Time frame: Dispense
Assessment of Lens Fit Performance for lens rotational recovery to original position. Collected at dispense for each lens. Assessed in degree of mislocation relative to original position after manual temporal rotation. (30 deg/10 blinks, 45 deg/60 sec)
Time frame: Dispense
Assessment of Lens Fit Performance for overall lens fit acceptance. Collected at dispense for each lens. Rated perfect or not perfect based on lens fit alone. (0-4; 0=should not be worn, 3=not perfect but OK to dispense, 4=perfect)
Time frame: 2 weeks
Participant rating of lens Daily and Comfortable Wearing Time. Collected at 2 weeks wear for each lens. (The hours of average comfortable wearing time and average daily wearing time.)
Time frame: 4 weeks
Participant rating of lens Daily and Comfortable Wearing Time. Collected at 4 weeks wear for each lens. (The hours of average comfortable wearing time and average daily wearing time.)
Time frame: 2 Weeks
Proportion of subjects using rewetting drops. Collected at 2 weeks for each lens. (Yes, No)
Time frame: 4 Weeks
Proportion of subjects using rewetting drops. Collected at 4 weeks for each lens. (Yes, No)
Time frame: 2 weeks
Participant rating of lens Comfort, Dryness, Handling, Lens Fit Stability and Vision Satisfaction. Collected at 2 weeks wear for each lens. (0-10; Comfort, Lens Fit and Satisfaction / 0=very poor,10=excellent; Dryness / 0=very dry, 10=no dryness; Handling / 0=very difficult, 10=very easy for handling)
Time frame: 4 weeks
Participant rating of lens Comfort, Dryness, Handling, Lens Fit Stability and Vision Satisfaction. Collected at 4 weeks wear for each lens. (0-10; Comfort, Lens Fit and Satisfaction / 0=very poor,10=excellent; Dryness / 0=very dry, 10=no dryness; Handling / 0=very difficult, 10=very easy for handling)
Time frame: 2 weeks
Participant rating of vision quality on insertion, during the day, end of day and night. Collected at 2 weeks wear for each lens. (0-100; 0=extremely poor vision totally blurred, 100=excellent vision totally sharp)
Time frame: 4 weeks
Participant rating of vision quality on insertion, during the day, end of day and night. Collected at 4 weeks wear for each lens. (0-100; 0=extremely poor vision totally blurred, 100=excellent vision totally sharp)
Time frame: 2 weeks
Participant rating of vision stability on insertion, during the day, end of day. Collected at 2 weeks wear for each lens. (0-100; 0=totally unstable fluctuating/changing, 100=perfectly stable not fluctuating/changing)
Time frame: 4 weeks
Participant rating of vision stability on insertion, during the day, end of day. Collected at 4 weeks wear for each lens. (0-100; 0=totally unstable fluctuating/changing, 100=perfectly stable not fluctuating/changing)
Time frame: 2 weeks
Participant rating for overall sensation of moistness. Collected at 2 weeks wear for each lens. (5-point Likert Scale; Excellent, Good, Average, Below Average, Poor)
Time frame: 4 weeks
Participant rating for overall sensation of moistness. Collected at 4 weeks wear for each lens. (5-point Likert Scale; Excellent, Good, Average, Below Average, Poor)
Time frame: 2 weeks
Participant rating for overall sensation of smoothness. Collected at 2 weeks wear for each lens. (5-point Likert Scale; Excellent, Good, Average, Below Average, Poor)
Time frame: 4 weeks
Participant rating for overall sensation of smoothness. Collected at 4 weeks wear for each lens. (5-point Likert Scale; Excellent, Good, Average, Below Average, Poor)
Time frame: 2 weeks
Participant rating for comfort satisfaction. Collected at 2 weeks wear for each lens. (4-point Likert Scale; Completely Satisfied, Somewhat Satisfied, Somewhat Dissatisfied, Completely Dissatisfied)
Time frame: 4 weeks
Participant rating for comfort satisfaction. Collected at 4 weeks wear for each lens. (4-point Likert Scale; Completely Satisfied, Somewhat Satisfied, Somewhat Dissatisfied, Completely Dissatisfied)
Time frame: 2 weeks
Participant rating for dryness satisfaction. Collected at 2 weeks wear for each lens. (4-point Likert Scale; Completely Satisfied, Somewhat Satisfied, Somewhat Dissatisfied, Completely Dissatisfied)
Time frame: 4 weeks
Participant rating for dryness satisfaction. Collected at 4 weeks wear for each lens. (4-point Likert Scale; Completely Satisfied, Somewhat Satisfied, Somewhat Dissatisfied, Completely Dissatisfied)
Time frame: 2 weeks
Participant rating for handling satisfaction. Collected at 2 weeks wear for each lens. (4-point Likert Scale; Completely Satisfied, Somewhat Satisfied, Somewhat Dissatisfied, Completely Dissatisfied)
Time frame: 4 weeks
Participant rating for handling satisfaction. Collected at 4 weeks wear for each lens. (4-point Likert Scale; Completely Satisfied, Somewhat Satisfied, Somewhat Dissatisfied, Completely Dissatisfied)
Time frame: 2 weeks
Participant rating for vision satisfaction. Collected at 2 weeks wear for each lens. (4-point Likert Scale; Completely Satisfied, Somewhat Satisfied, Somewhat Dissatisfied, Completely Dissatisfied)
Time frame: 4 weeks
Participant rating for vision satisfaction. Collected at 4 weeks wear for each lens. (4-point Likert Scale; Completely Satisfied, Somewhat Satisfied, Somewhat Dissatisfied, Completely Dissatisfied)
Time frame: 2 weeks
Participant rating for lens fit satisfaction. Collected at 2 weeks wear for each lens. (4-point Likert Scale; Completely Satisfied, Somewhat Satisfied, Somewhat Dissatisfied, Completely Dissatisfied)
Time frame: 4 weeks
Participant rating for lens fit satisfaction. Collected at 4 weeks wear for each lens. (4-point Likert Scale; Completely Satisfied, Somewhat Satisfied, Somewhat Dissatisfied, Completely Dissatisfied)
Time frame: 2 weeks
Participant rating for overall satisfaction. Collected at 2 weeks wear for each lens. (4-point Likert Scale; Completely Satisfied, Somewhat Satisfied, Somewhat Dissatisfied, Completely Dissatisfied)
Time frame: 4 weeks
Participant rating for overall satisfaction. Collected at 4 weeks wear for each lens. (4-point Likert Scale; Completely Satisfied, Somewhat Satisfied, Somewhat Dissatisfied, Completely Dissatisfied)
Time frame: 2 Weeks
Assessment of wavefront aberrations. Collected at 2 weeks for each lens. Wavefront measurement (3mm), Scale in microns (µm).
Time frame: 4 Weeks
Assessment of wavefront aberrations. Collected at 4 weeks for each lens. Wavefront measurement (3mm), Scale in microns (µm).
Time frame: 2 Weeks
Assessment of wavefront aberrations. Collected at 2 weeks for each lens. Wavefront measurement (5 mm), Scale in microns (µm).
Time frame: 4 Weeks
Assessment of wavefront aberrations. Collected at 4 weeks for each lens. Wavefront measurement (5 mm), Scale in microns (µm).
Time frame: 2 Weeks
Assessment of monocular and binocular high and low contrast visual acuity (VA). Collected at 2 weeks for each lens. logMAR (VA).
Monocular High Contrast Visual Acuity (MHCVA), Monocular Low Contrast Visual Acuity (MLCVA), Binocular High Contrast Visual Acuity (BHCVA), Binocular Low Contrast Visual Acuity (BLCVA)
Time frame: 4 Weeks
Assessment of monocular and binocular high and low contrast visual acuity (VA). Collected at 4 weeks for each lens. logMAR (VA).
Monocular High Contrast Visual Acuity (MHCVA), Monocular Low Contrast Visual Acuity (MLCVA), Binocular High Contrast Visual Acuity (BHCVA), Binocular Low Contrast Visual Acuity (BLCVA)
Time frame: 2 weeks
Assessment of Lens Fit Performance for lens orientation in primary position of gaze. Assessed at 2 weeks wear for each lens. (Degree of mislocation relative to lens axis mark.)
Time frame: 4 weeks
Assessment of Lens Fit Performance for lens orientation in primary position of gaze. Assessed at 4 weeks wear for each lens. (Degree of mislocation relative to lens axis mark.)
Time frame: 2 weeks
Assessment of Lens Fit Performance for lens rotational recovery to original position. Collected at 2 weeks wear for each lens. Assessed in degree of mislocation relative to original position after manual temporal rotation. (30 deg/10 blinks, 45 deg/60 sec)
Time frame: 4 weeks
Assessment of Lens Fit Performance for lens rotational recovery to original position. Collected at 2 weeks wear for each lens. Assessed in degree of mislocation relative to original position after manual temporal rotation. (30 deg/10 blinks, 45 deg/60 sec)
Time frame: 2 weeks
Assessment of lens front surface deposits. Collected at 2 weeks wear for each lens. (Front surface deposits observed, 0-4, 0=clean, 4=deposits ≥0.5)
Time frame: 4 weeks
Assessment of lens front surface deposits. Collected at 4 weeks wear for each lens. (Front surface deposits observed, 0-4, 0=clean, 4=deposits ≥0.5)
Time frame: 2 weeks
Assessment of Lens Fit Performance for centration. Collected at 2 weeks for each lens. Proportion of contact lenses fitted where centration was centered or slightly decentered. (Biomicroscopy)
Time frame: 4 weeks
Assessment of Lens Fit Performance for centration. Collected at 4 weeks for each lens. Proportion of contact lenses fitted where centration was centered or slightly decentered. (Biomicroscopy)
Time frame: 2 weeks
Assessment of Lens Fit Performance for corneal coverage. Collected at 2 weks for each lens. Corneal coverage assessed in primary gaze: (yes=full corneal coverage at all times, no=incomplete corneal coverage)
Time frame: 4 weeks
Assessment of Lens Fit Performance for corneal coverage. Collected at 4 weeks for each lens. Corneal coverage assessed in primary gaze: (yes=full corneal coverage at all times, no=incomplete corneal coverage)
Time frame: 2 weeks
Assessment of Lens Fit Performance for post blink movement. Collected at 2 weeks for each lens. (0-4, 0.5 increments; 0=Insufficient, unacceptable movement, 4= Excessive, unacceptable movement)
Time frame: 4 weeks
Assessment of Lens Fit Performance for post blink movement. Collected at 4 weeks for each lens. (0-4, 0.5 increments; 0=Insufficient, unacceptable movement, 4= Excessive, unacceptable movement)
Time frame: 2 weeks
Assessment of Lens Fit Performance for lens marking visibility. Collected at 2 weeks for each lens. (1-3, 1=excellent, 2=average, 3=poor)
Time frame: 4 weeks
Assessment of Lens Fit Performance for lens marking visibility. Collected at 4 weeks for each lens. (1-3, 1=excellent, 2=average, 3=poor)
Time frame: 2 weeks
Assessment of Lens Fit Performance for lens rotational stability on blink. Collected at 2 weeks for each lens. (No rotation and 5-10 degrees rotation from axis location mark)
Time frame: 4 weeks
Assessment of Lens Fit Performance for lens rotational stability on blink. Collected at 4 weeks for each lens. (No rotation and 5-10 degrees rotation from axis location mark)
Time frame: 2 weeks
Assessment of Lens Fit Performance for lens to stabilize in 5-10 min. Collected at 2 weeks for each lens. Varied less than 5 degrees from lens marking location between 5-10 min.
Time frame: 4 weeks
Assessment of Lens Fit Performance for lens to stabilize in 5-10 min. Collected at 4 weeks for each lens. Varied less than 5 degrees from lens marking location between 5-10 min.
Time frame: 2 weeks
Assessment of Lens Fit Performance for overall lens stability. Collected at 2 weeks for each lens. (Excellent or Good.
Time frame: 4 weeks
Assessment of Lens Fit Performance for overall lens stability. Collected at 4 weeks for each lens. (Excellent or Good.
Time frame: 2 weeks
Assessment of Lens Fit Performance for overall lens fit acceptance. Collected at 2 weeks for each lens. Rated perfect or not perfect based on lens fit alone. (0-4; 0=should not be worn, 3=not perfect but OK to dispense, 4=perfect)
Time frame: 4 weeks
Assessment of Lens Fit Performance for overall lens fit acceptance. Collected at 4 weeks for each lens. Rated perfect or not perfect based on lens fit alone. (0-4; 0=should not be worn, 3=not perfect but OK to dispense, 4=perfect)
Time frame: Baseline
Assessment of ocular health. Collected at baseline after removal of habitual lenses. Proportion of eyes with grades 0-1 (No eyes graded >1) (Biomicroscopy, 0-4, ½ grades; 0=none: no injection present, 4=severe)
Time frame: 2 weeks
Assessment of ocular health. Collected at 2 weeks after removal of lenses. Proportion of eyes with grades 0-1 (No eyes graded >1) (Biomicroscopy, 0-4, ½ grades; 0=none: no injection present, 4=severe)
Time frame: 4 weeks
Assessment of ocular health. Collected at 4 weeks after removal of lenses. Proportion of eyes with grades 0-1 (No eyes graded >1) (Biomicroscopy, 0-4, ½ grades; 0=none: no injection present, 4=severe)
Time frame: Baseline
Assessment of ocular health. Collected at baseline after removal of habitual lenses. Proportion of eyes with grades 0-1 (No eyes graded >1) (Biomicroscopy, 0-4, ½ grades; 0=none: no injection present, 4=severe)
Time frame: 2 weeks
Assessment of ocular health. Collected at 2 weeks after removal of lenses. Proportion of eyes with grades 0-1 (No eyes graded >1) (Biomicroscopy, 0-4, ½ grades; 0=none: no injection present, 4=severe)
Time frame: 4 weeks
Assessment of ocular health. Collected at 4 weeks after removal of lenses. Proportion of eyes with grades 0-1 (No eyes graded >1) (Biomicroscopy, 0-4, ½ grades; 0=none: no injection present, 4=severe)
Time frame: Baseline
Assessment of ocular health. Collected at baseline after removal of habitual lenses. Proportion of eyes with grades 0-1 (No eyes graded >1) (Biomicroscopy, 0-4, ½ grades; 0=none: no injection present, 4=severe)
Time frame: 2 weeks
Assessment of ocular health. Collected at 2 weeks after removal of lenses. Proportion of eyes with grades 0-1 (No eyes graded >1) (Biomicroscopy, 0-4, ½ grades; 0=none: no injection present, 4=severe)
Time frame: 4 weeks
Assessment of ocular health. Collected at 4 weeks after removal of lenses. Proportion of eyes with grades 0-1 (No eyes graded >1) (Biomicroscopy, 0-4, ½ grades; 0=none: no injection present, 4=severe)
Time frame: Baseline
Assessment of ocular health. Collected at baseline after removal of habitual lenses. Proportion of eyes with grades 0-1 (No eyes graded >1) (Biomicroscopy, 0-4, ½ grades; 0=none: no injection present, 4=severe)
Time frame: 2 weeks
Assessment of ocular health. Collected at 2 weeks after removal of lenses. Proportion of eyes with grades 0-1 (No eyes graded >1) (Biomicroscopy, 0-4, ½ grades; 0=none: no injection present, 4=severe)
Time frame: 4 weeks
Assessment of ocular health. Collected at 4 weeks after removal of lenses. Proportion of eyes with grades 0-1 (No eyes graded >1) (Biomicroscopy, 0-4, ½ grades; 0=none: no injection present, 4=severe)
Time frame: Baseline
Investigators' objective assessment of ocular health. Collected at baseline after removal of habitual lenses. Proportion of eyes with grades 0-1 (No eyes graded >1) (Biomicroscopy, 0-4, ½ grades; 0=none: no injection present, 4=severe)
Time frame: 2 weeks
Assessment of ocular health. Collected at 2 weeks after removal of lenses. Proportion of eyes with grades 0-1 (No eyes graded >1) (Biomicroscopy, 0-4, ½ grades; 0=none: no injection present, 4=severe)
Time frame: 4 weeks
Assessment of ocular health. Collected at 4 weeks after removal of lenses. Proportion of eyes with grades 0-1 (No eyes graded >1) (Biomicroscopy, 0-4, ½ grades; 0=none: no injection present, 4=severe)
Time frame: Baseline
Assessment of ocular health. Collected at baseline after removal of habitual lenses. Proportion of eyes with grades 0-1 (No eyes graded >1) (Biomicroscopy, 0-4, ½ grades; 0=none: no injection present, 4=severe)
Time frame: 2 weeks
Assessment of ocular health. Collected at 2 weeks after removal of lenses. Proportion of eyes with grades 0-1 (No eyes graded >1) (Biomicroscopy, 0-4, ½ grades; 0=none: no injection present, 4=severe)
Time frame: 4 weeks
Assessment of ocular health. Collected at 4 weeks after removal of lenses. Proportion of eyes with grades 0-1 (No eyes graded >1) (Biomicroscopy, 0-4, ½ grades; 0=none: no injection present, 4=severe)
Time frame: 2 weeks
Participant preference for their habitual lenses or the first study lenses during the last two weeks with regard to comfort, dryness, handling, vision, lens fit and overall. (Forced choice; habitual, study lenses)
Time frame: 2 weeks
Participant preference for their habitual lenses or the first study lenses during the last two weeks with regard to comfort, dryness, handling, vision, lens fit and overall. (Forced choice; habitual, study lenses)
Time frame: 4 weeks
Participant preference for their habitual lenses or the first study lenses during the last four weeks with regard to comfort, dryness, handling, vision, lens fit and overall. (Forced choice; habitual, study lenses)
Time frame: 4 weeks
Participant preference for their habitual lenses or the first study lenses during the last four weeks with regard to comfort, dryness, handling, vision, lens fit and overall. (Forced choice; habitual, study lenses)
Time frame: 4 weeks
Participants likelihood of switching from habitual lenses to either pair of the study lenses when asked; How likely are they to switch from their habitual lenses to either the first study lenses or the second study lenses?" Collected at 4 weeks for each study pair. (4 point Likert scale; very likely, likely, unlikely, very unlikely)
Time frame: 4 weeks
Participants likelihood of continuing to wear either of the pairs of study lenses when asked; "How likely are they to continue to wearing either the first study lenses or the second study lenses?" Collected at 4 weeks fore each study pair. (4 point Likert scale; very likely, likely, unlikely, very unlikely)
Time frame: 8 weeks
Percentage of participants that answer the question, "What study lens they will most likely recommend to friends, family or colleagues?" Collected at end of study. (Forced choice; First Study Lenses, Second Study Lenses)
Time frame: 8 weeks
Percentage of participants that answer the question, "How likely are they to recommend either the first pair of study lenses or the second pair of study lenses to friends, family or colleagues?" Collected at study end. (4 point Likert scale; very likely, likely, unlikely, very unlikely)
Time frame: 8 weeks
Percentage of participants that answer the question, "Which type of study lens they prefer with regard to comfort, dryness, handling, vision, lens fit and overall performance?" Collected at study end. (Forced choice; habitual lenses, first study lenses, second study lenses)
Time frame: 8 weeks
Percentage of participants that answer the question, "Which type of study lens they prefer with regard to comfort, dryness, handling, vision, lens fit and overall performance?" Collected at study end. (Forced choice; first study lenses, second study lenses, neither)
CooperVision, Inc.
Industry
A Single-Center Study Comparing the Clinical Performance of Biofinity Toric Versus Air Optix for Astigmatism Over 1 Month of Daily Wear.
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT06775509
Astigmatism, Eye Diseases
Tarragona, Spain
View Trial DetailsNCT06784063
Astigmatism, Cataract
Bloomfield Hills, Michigan, United States
View Trial DetailsNCT06749496
Astigmatism, Eye Diseases
Jacksonville, Florida, United States
View Trial DetailsNCT06967129
Astigmatism, Eye Diseases
Mission Viejo, California, United States
View Trial Details