POLFA (Needle Assembly)
DeviceEach subject receives the same intervention and blood is drawn into both the INA and CNA. The difference is the order in which the CNA or INA is applied.
NCT Number: NCT02476851
This study evaluates whether whole blood transferred through the new POLFA needle assembly meets supernatant hemoglobin acceptability standards.
Looking for future studies?
Notify Me18 year and older
All sexes
Interventional
Not applicable
Haemonetics, Braintree, Massachusetts, United States
The POLFA needle is part of the set's PPD (personal protection device) sampling port assembly that facilitates safer blood transfer between the collection set (i.e., the diversion pouch) and a Vacutainer® for the purpose of viral testing or other testing of the whole blood. The POLFA needle assembly was developed as a means to reduce costs while maintaining functionality, quality and safety. Since the overall functionality of the POLFA needle was designed to be equal to the currently approved Kawasumi needle, a change to the overall device safety and use profile is neither anticipated nor intended.
This trial will serve to generate data sufficient to demonstrate that blood transferred through the POLFA needle results in plasma hemoglobin levels below 100mg/dL.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Eligibility Criteria:
Physical trauma consistent with associated coagulopathy within the last 30 days, Surgery within the last 30 days, Known history of hypercoagulopathy (i.e., Factor V Leiden, Prothrombin G20210A, idiopathic venous thrombotic events, etc.).
Exclusion criteria
- any individual not meeting the above criteria.
Each subject receives the same intervention and blood is drawn into both the INA and CNA. The difference is the order in which the CNA or INA is applied.
Each subject receives the same intervention and blood is drawn into both the INA and CNA. The difference is the order in which the CNA or INA is applied.
Time frame: Following a one-time whole blood donation, samples were measured within 1 month of donation.
To determine whether whole blood transferred to a Vacutainer® through the POLFA needle will have supernatant hemoglobin levels <100 mg/dL within a 95% CI and 95% reliability. The average hemoglobin levels in the blood collected through each assembly is presented.
Haemonetics Corporation
Industry
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT02110576
Transmission, Blood, Recipient/Donor
Denver, Colorado, United States
View Trial DetailsNCT01756976
Transmission, Blood, Recipient/Donor
Pittsburgh, Pennsylvania, United States
View Trial Details