GTU Artificial Urinary Sphincter
DeviceA totally implantable Artificial Urinary Sphincter (AUS) for the treatment of male Stress Urinary Incontinence (SUI)
NCT Number: NCT02288455
The purpose of this investigation is to demonstrate the safety and efficacy of the GTU artificial urinary sphincter device in restoring continence in males who have confirmed urinary stress incontinence for a minimum of 12 months with primary etiology being radical prostatectomy or transurethral resection of the prostate (TURP).
Looking for future studies?
Notify Me21 year and older
Male
Interventional
Not applicable
Urology Centre, Port Macquarie, New South Wales, Australia
Prospective, non-randomized multi-center study testing the safety and efficacy of the GTU artificial urinary sphincter device in males with urinary incontinence.
Up to 20 sites in Europe, Australia & New Zealand
A sample size of 73 subjects is estimated to provide 80% power for efficacy. Accounting for 10% attrition, 82 subjects will be enrolled to provide non-missing data on 73 subjects.
First Subject Enrolment: July 2014 Last Subject Enrolment: September 2015 Last Subject Follow-up: September 2016 Study Primary Endpoint Completion: January 2016
Each subject will be followed for approximately 14 months. Study duration is approximately 27 months.
Subjects will be evaluated at the screening, pre-implant, implant, 6 weeks post implant for device activation, 1-month, 2-month, 3-month, 6-month, and 12-month post device activation.
The Sponsor (GT Urological, LLC) will work with a Contract Research Organization (CRO) to assist in investigative site monitoring for this clinical investigation.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
a. One 24-hour pad weight test ≥300 gm
Exclusion criteria
A totally implantable Artificial Urinary Sphincter (AUS) for the treatment of male Stress Urinary Incontinence (SUI)
Time frame: 3 months
The primary efficacy endpoint is improvement in pad weight defined as at least 60% improvement in pad weight from baseline to 3-months post device activation as measured by the 24-hour pad weight tests.
Time frame: 3 months
Improvement in 1- hour pad weight test results from baseline to 3-month post-device activation visit.
Time frame: 3 months
Improvement in number of pads used per day from baseline to 3-month post-device activation as measured by a 3-day voiding diary.
Time frame: 3 months
Improvement in Quality of Life assessments as measured by:
Time frame: 3 months
Summary of all adverse events
Time frame: 3 months
The primary safety endpoint is a composite of the following major device-related adverse events and/or outcomes at 3 months post-device activation as reported by the investigational site. The components of this composite safety endpoint are:
GT Urological, LLC
Industry
A Prospective, Non-Randomized, Multi-Center Clinical Investigation of the Safety and PeRformancE of GT UroLogIcal, LLC's ArtiFicial Urinary Sphincter (RELIEF II)
Acronym: RELIEFII
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT01779323
Female Urogenital Diseases, Female Urogenital Diseases and Pregnancy Complications
Durham, North Carolina, United States
View Trial DetailsNCT05771168
Female Urogenital Diseases, Female Urogenital Diseases and Pregnancy Complications
Roma, Rome, Italy
View Trial Details