Skip to main content
OpenTrials
Completed

NCT Number: NCT04497623

Clinical Evaluation of Ventilador Innovation Product in Colombia in the SARS COVID 19 Pandemic, Unisabana Herons.

The objective of this study is to evaluate the efficacy and safety of the Unisabana-Herons invasive mechanical ventilator designed to provide the basic ventilatory support necessary to preserve the life of patients with respiratory failure and indication of mechanical ventilation, especially for those who suffer from acute respiratory distress syndrome (ARDS) when conventional commercial invasive ventilators are not available in the context of the health emergency due to the COVID-19 epidemic.

The Unisabana-Herons ventilator allows to precisely configure the respiratory rate, tidal volume (or inspired air volume), inspiratory time, the inspiration: expiration ratio, the positive pressure at the end of expiration (PEEP), the inspired fraction of oxygen and inspiratory air flow, parameters that allow managing the respiratory failure associated with COVID-19. The ventilator also monitors peak inspiratory pressures (PIP), mean, PEEP, plateau, and graphs in real time the pressure-time, volume-time, flow-time curves, which allows detecting when one of these is at levels dangerous to induce ventilator trauma (barotrauma and volutrauma) and thus ensure effective and safe ventilation, so as to avoid ventilator-induced lung injury.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year–70 year

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Fundacion Neumologica Colombiana, Bogotá, Bogota D.C., Colombia

Loading trial locations.

About this study

The Unisabana-Herons ventilator is an invasive mechanical ventilator that works on the same principles of invasive positive pressure mechanical ventilators that have existed for 80 years. Although the effectiveness of ventilatory assistance in saving human lives was known since biblical times, the first mechanical ventilators only appeared in 1800 and it was in 1900 when the first positive pressure ventilators were manufactured, which have a turning point in 1940 as a result of the polio epidemic, when invasive positive pressure mechanical ventilators were developed that could be used massively and have evolved to current models. These positive pressure fans completely replaced the first negative pressure models, have abundant support in the scientific literature, and are the most commonly used today.

Since the beginning of the COVID-19 epidemic in Colombia, the University of La Sabana, a multidisciplinary team in order to find solutions to deal with the disease, and its first project, consisted in the design and manufacture of an invasive mechanical ventilator (Ventilator Unisabana-Herons) able to supply the basic ventilatory needs of the patient with severe respiratory failure due to COVID-19 at the time when the installed capacity of classic commercial mechanical ventilators is exhausted. The Unisabana-Herons ventilator was built based on those recommended by INVIMA, the MHRA (UK Medicines and Devices Regulatory Agency) and the FDA, to provide efficient and safe volume controlled ventilation to patients with indications. of mechanical ventilation for respiratory failure according to the ventilatory modes already affected that have strong scientific evidence.

This study seeks to evaluate the effectiveness, usability and safety of the Unisabana-Herons ventilator for the management of patients with an indication for invasive mechanical ventilatory support, admitted to level III and IV university hospitals with Intensive Care services enabled through a cohort study of 5 patients with a 24-hour follow-up.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Patients older than 18 years and younger than 70 years with indication of volume-controlled mechanical ventilation for more than 24 hours.
  • Patients with acute respiratory failure (PaO2 / FiO2 <300) requiring volume-controlled mechanical ventilation. These patients may or may not have COVID-19 (at the current time of the epidemic, it is assumed that every patient with an indication for mechanical ventilation is a possible case of COVID-19).
  • Postoperative patients who require ventilatory support and are expected to need it for more than 24 hours.
  • Patients with traumatic brain injury and indication of mechanical ventilatory support with an expected duration greater than 24 hours
  • Patients with acute intoxication and respiratory depression and indication of mechanical ventilatory support with an expected duration greater than 24 hours

Exclusion criteria

  • Pregnant women
  • Patients with hypotension MAP <65 mmHg
  • Patients with PaO2 / FiO2 <100
  • Cerebral edema in cerebral protection and / or suspected endocranial hypertension
  • SOFA >9
  • For those patients who are already receiving mechanical ventilation, the presence of one or more of the following criteria: PEEP> 8 cmH2O, plateau pressure> 30 cm H2O or FiO2> 70%
  • COVID-19 confirmed by RT-PCR.

Treatment and study plan

Invasive mechanical ventilation using the Unisabana-Herons Ventilator during 24 hours

Device

Recruited patients will receive volume-controlled mechanical ventilation (assist-control mode) using the Unisabana-Herons ventilator for 24 hours. Information on clinical, hemodynamic, and respiratory variables will be recorded from 30 minutes before the start of mechanical ventilation with the Unisabana-Herons ventilator. Recordings will be made in the first 4 hours every 15 minutes in the CFR, except for arterial blood gases which will be taken every 30 minutes. In the following 20 hours, arterial blood gas controls will be taken at hour 12 from the start of the ventilator and at hour 24. Other hemodynamic and respiratory variables will be recorded every hour from hour 4 of ventilator start until hour 24.

Primary outcomes

  1. Improvement or maintenance of the oxygenation level measured by PaO2

    Time frame: 24 hours

    Maintenance: less than 20% drop in PaO2 with respect to the baseline measurements made in the 30 minutes before connecting the Unisabana-Herons ventilator.

  2. Improvement or maintenance of the oxygenation level measured by O2 Saturation

    Time frame: 24 hours

    Maintenance: less than 20% drop in SatO2 levels with respect to the baseline measurements made in the 30 minutes before connecting the Unisabana-Herons ventilator.

Secondary outcomes

  1. Improvement or maintenance of adequate levels of carbon dioxide measured by PaCO2

    Time frame: 24 hours

    Maintenance is defined as a change of less than 20% in the levels of this variable with respect to the baseline measurements made in the 30 minutes before connecting the Unisabana-Herons ventilator.

  2. Improvement or maintenance of adequate levels of HCO3

    Time frame: 24 hours

    Maintenance is defined as a change of less than 20% in the levels of this variable with respect to the baseline measurements made in the 30 minutes before connecting the Unisabana-Herons ventilator.

  3. Improvement or maintenance of adequate levels of excess base.

    Time frame: 24 hours

    Maintenance is defined as a change of less than 20% in the levels of this variable with respect to the baseline measurements made in the 30 minutes before connecting the Unisabana-Herons ventilator.

  4. Improvement or maintenance of adequate levels of blood pH

    Time frame: 24 hours

    Maintenance is defined as a change of less than 20% in the levels of this variable with respect to the baseline measurements made in the 30 minutes before connecting the Unisabana-Herons ventilator.

  5. Improvement or maintenance of PaO2/FiO2

    Time frame: 24 hours

    Stability is defined as a decrease of less than 20% in this variable with respect to the baseline measurements made in the 30 minutes before connecting the Unisabana-Herons ventilator.

  6. Improvement or maintenance of SatO2/FiO2

    Time frame: 24 hours

    Stability is defined as a decrease of less than 20% in this variable with respect to the baseline measurements made in the 30 minutes before connecting the Unisabana-Herons ventilator.

  7. Uninterrupted and faultless operation in the period of use of the ventilator

    Time frame: 24 hours

    Present or absent outcome

  8. Inspiratory peak pressure> 35 cm H2O that does not have a clinical explanation other than the ventilator (such as a mucus plug)

    Time frame: 24 hours

    Peak pressure >35 CM H20.

  9. Plateau airway pressure> 30 cm H2O that does not have a clinical explanation other than the ventilator

    Time frame: 24 hours

    Present or absent outcome

  10. VT> 8 cc / kg of ideal weight that does not have a clinical explanation other than the ventilator

    Time frame: 24 hours

    Present or absent outcome

  11. Decrease or increase in respiratory rate, tidal volume, PEEP, peak inspiratory pressure, FiO2, not due to a clinician order (changes not ordered by the clinical team but due to the ventilator variability)

    Time frame: 24 hours

    Present or absent outcome

  12. Pneumothorax (not having an explanation other than ventilatory support, such as the insertion of a central catheter)

    Time frame: 24 hours

    Present or absent outcome

  13. Pneumomediastinum (not having an explanation other than ventilatory support, such as the insertion of a central catheter)

    Time frame: 24 hours

    Present or absent outcome

  14. Subcutaneous emphysema (not having an explanation other than ventilatory support, such as the insertion of a central catheter)

    Time frame: 24 hours

    Present or absent outcome

  15. Hemodynamic deterioration in the hour following the start of the Unisabana-Herons ventilator that requires a 100% increase in the dose of vasopressors and that does not have a clinical explanation other than the ventilator

    Time frame: 24 hours

    Present or absent outcome

  16. Cardiac arrest without a clinical explanation other than the ventilator

    Time frame: 24 hours

    Present or absent outcome

  17. Death without a clinical explanation other than the ventilator

    Time frame: 24 hours

    Present or absent outcome

  18. Elevation of creatinine that does not have a clinical explanation other than the ventilator

    Time frame: 24 hours

    Present or absent outcome

  19. Elevation of BUN that does not have a clinical explanation other than the ventilator

    Time frame: 24 hours

    Present or absent outcome

  20. Digestive bleeding without a clinical explanation other than ventilator

    Time frame: 24 hours

    Present or absent outcome

  21. Stress ulcers (upper gastrointestinal tract) without a clinical explanation other than ventilator

    Time frame: 24 hours

    Present or absent outcome

  22. Pneumonia associated with ventilator.

    Time frame: 24 hours

    Present or absent outcome

  23. Tracheobronchitis associated with ventilator.

    Time frame: 24 hours

    Present or absent outcome

  24. Critical care polyneuropathy that does not have a different explanation for the use of the ventilator and / or muscle relaxants necessary for mechanical ventilation

    Time frame: 24 hours

    Present or absent outcome

  25. Critical care myopathy that does not have a different explanation for the use of the ventilator and / or muscle relaxants necessary for mechanical ventilation

    Time frame: 24 hours

    Present or absent outcome

Sponsors and collaborators

Lead sponsor

Fundación Neumologica Colombiana

Other

Collaborators

  • Clínica Universidad de La Sabana
  • Fundación Cardioinfantil Instituto de Cardiología
  • Universidad de la Sabana

Registry information

Official study title

Clinical Evaluation of Ventilator Innovation Product in Colombia in the SARS Pandemic Covid 19, Unisabana Herons - A Cohort Study - Phase 1

Acronym: SabanaHerons

Important dates

Study start
2020
Primary completion
2020
Study completion
2020
First posted
Aug 4, 2020
Registry last updated
Sep 24, 2020

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.