NCT Number: NCT05715515
This study is designed as an international prospective, multicentric, clinical study to investigate the performance and usability of the Panbio™ COVID-19/Flu A&B Rapid Panel Professional Use and Self- Test devices for the qualitative detection of COVID-19 antigen, Influenza A antigen and Influenza B antigen in human nasopharyngeal (NP) and mid-turbinate nasal swabs, respectively.
This study is part of the performance evaluation to support the CE conformity assessment procedures.
Looking for future studies?
Notify MeAll sexes
Interventional
Not applicable
Easley, South Carolina, 29640, United States
In total, a minimum of 1531 male and female subjects will be prospectively enrolled at multiple clinical sites. A minimum of 100 lay users will also be asked to interpret randomly assigned mock devices.
After obtaining written informed consent, a study-specific subject identification number (ID) will be assigned to the subject. All eligible subjects will be assigned to both Professional Use arm and the Self-Test arm of the study. Subject demographics and a brief medical history will be collected.
Lay users (self-tester or caregiver) will first collect one mid-turbinate nasal swab from both nostrils, and perform and interpret the Panbio™ Rapid Panel ST device, under the observation of a trained observer. All procedures for testing and result interpretation, including sample collection and extraction, will be conducted following the IFU provided in the kit. Each Panbio™ Rapid Panel ST result will be interpreted firstly by the lay user and followed immediately by the observer who will also photograph the test at the time of interpretation. Each lay user will fill out a user evaluation questionnaire, and a study staff member will complete an observer questionnaire for each lay user. A minimum of 100 lay users will also read mock device tests presented by a study staff member. Mock devices (simulated to represent testing results) will be used for usability testing (result interpretation).
A study staff member will then collect two NP swabs (one swab from each nostril) from all subjects. The first NP swab sample will be used to conduct a Panbio™ Rapid Panel PU test, following the IFU. The test result will be photographed at the time of result interpretation. Each study staff member who performs a Panbio™ Rapid Panel PU test will fill out a user evaluation questionnaire once, to assess the usability of the device.
The second NP swab sample will be eluted in Universal Transport Medium (UTM) provided by Abbott/the Core Laboratory, labelled and stored, according to the laboratory manual. UTM samples will be shipped to the Core Laboratory for testing with RT-PCR protocols for Flu A, Flu B and SARS-CoV-2, according to the laboratory manual.
Each subject's demographic information, education, symptomology data, test results and questionnaires, where relevant, will be recorded on source documentation and transferred to an Electronic Data Capture system.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Subjects from all age groups, able and willing to provide written informed consent or assent, suspected of having a respiratory viral infection consistent with COVID-19 and/or Flu by their healthcare provider within the first five days of the onset of symptoms, and presenting at least two of the following symptoms: fever, headache, tiredness, dry cough, sore throat, runny or stuffy nose, muscle aches, loss of smell, loss of taste or shortness of breath, will be prospectively enrolled.
Exclusion criteria
The Panbio™ COVID-19/Flu A&B Rapid Panel Professional use device is a visual lateral flow assay, for rapid immunochromatographic qualitative detection of COVID-19 antigen, Influenza A antigen and Influenza B antigen in nasopharyngeal swabs, collected by a trained health worker.
The Panbio™ COVID-19/Flu A&B Rapid Panel Self-Test device is a visual lateral flow assay, for rapid immunochromatographic qualitative detection of COVID-19 antigen, Influenza A antigen and Influenza B antigen in self-collected mid-turbinate nasal swabs. The test is designed to be used by lay users
Other names: Panbio™ COVID-19/Flu A&B Rapid Panel Professional Use Product, Panbio™ COVID-19/Flu A&B Rapid Panel Self-Test Device
Time frame: "up to 24 months"
To assess the clinical performance (sensitivity/specificity) of the Panbio™ COVID-19/Flu A&B Rapid Panel in fresh collected mid-turbinate nasal swab specimens from individuals suspected of respiratory viral infection consistent with COVID-19 and/or Flu by their healthcare provider within the first five days of the onset of symptoms in a Near Patient Testing setting, compared to Flu and SARS-CoV-2 RT-PCR tested using NP swab samples. Concordance analyisis of investigational device and reference method. Acceptance criterion: Sensitivity >80%, Specificity >95%
To assess the clinical performance (sensitivity/specificity) of the Panbio™ COVID- 19/Flu A&B Rapid Panel Self-Test device, as performed by lay users (self-testers or caregivers) using fresh collected mid-turbinate nasal self- collected samples, when compared to Flu and SARS-CoV-2 RT- PCR using NP swab samples.
Time frame: "up to 24 months"
To assess concordance between the Panbio™ COVID-19/Flu A&B Rapid Panel Self-Test device using mid-turbinate nasal self-collected samples and the Professional Use device using NP samples collected by a trained health worker.
Time frame: "up to 24 months"
To assess the usability of the Panbio™ COVID-19/Flu A&B Rapid Panel professional use device when used by trained healthcare workers. Usability is assessed by a questionnaire.
Time frame: "up to 24 months"
To assess the usability of the Panbio™ COVID-19/Flu A&B Rapid Panel Self-Test device, as performed by lay users with no laboratory or clinical experience. Usability testing will also comprise of the interpretation of mock devices. User and observer questionnaires will also be utilized to assess usability.
Abbott Rapid Dx
Industry
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT05268939
COVID-19, Coronaviridae Infections
Cullman, Alabama, United States
View Trial DetailsNCT06124638
COVID-19, Coronaviridae Infections
Pelham, Alabama, United States
View Trial DetailsNCT05630365
COVID-19, Coronaviridae Infections
Nashville, Tennessee, United States
View Trial DetailsNCT05354115
COVID-19, Coronaviridae Infections
Easley, South Carolina, United States
View Trial Details