DaVita Clinical Research Germany
Düsseldorf, D-40201, Germany
NCT Number: NCT04395131
The study will be a prospective, interventional, non-randomized, single-center study in 20 ESRD patients on hemodialysis. The study will compare performance of the Clearum HS dialyzer to typical values obtained with other commercially available high flux dialyzers. In addition, the Clearum HS dialyzer will be compared to a Fresenius FX80 dialyzer for a baseline (control) comparison. This will be a post-market study. The duration of the trial is 2 weeks for the FX80 control dialyzer + 6 weeks with the Clearum HS dialyzer used for 3x weekly high flux hemodialysis.
Looking for future studies?
Notify Me18 year and older
All sexes
Interventional
Not applicable
Düsseldorf, D-40201, Germany
The study will compare performance to typical values obtained with commercially available dialyzers used for high flux hemodialysis and comparing the Clearum HS dialyzer to the baseline values of a control FX80 Dialyzer.
The duration of the trial is 2 weeks for the FX80 control dialyzer + 6 weeks with the Clearum HS dialyzer used for 3x weekly high flux hemodialysis.
The sampling, reported in Table 1 below, will be done at mid-week sessions (pre and post session sampling) and consists of a blood draw for baseline dialysis with the FX80 weeks 1 and 2, or Clearum HS dialyzer at weeks 3, 5 and 8 for uremic toxins: urea, creatinine, phosphate, B2-microglobulin, albumin, sodium, potassium, calcium, magnesium, platelets and leucocytes, and platelets. In addition, a presession sample will be taken for coagulation parameter TAT, hematocrit and inflammatory markers IL6 and C reactive protein at baseline for FX80 at mid-week sessions during weeks 1 and 2 and during weeks 3, 5 and 8 for the Clearum HS dialyzer.
Instantaneous B2M clearance sampling will be done at week 7 during the Clearum HS dialyzer phase and consists of a blood sample prior to the dialyzer and immediately after the dialyzer at two time points, 15-30 min and at 120-150 min, respectively.
Blood rest will be done by comparing the dialyzer to a set of 5 illustrations and giving a score of 1 (worst) to 5 (best) that best correlates the dialyzer with the illustration, at the end of the dialysis session after blood has been returned to the patients (both FX80 and Clearum HS) on the sampling days.
The patients will be treated for 2 weeks with a standard dialyzer (FX80) and an additional 6 weeks with the Clearum HS dialyzer (experimental period). Both hemodialyzers will be used for 3 times a week. Expected duration for the study including enrolment will be 3 months.
The study duration will be approximately 3-4 months. Each patient will receive 6 hemodialysis sessions with FX80 and 18 sessions with Clearum HS hemodialyzer, and the treatment time will be between 3.5 and 4.5 hours
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
the hemodialyzer will be used 3x week for 3.5-4.5 hours with a Qb of at least 350 ml/min
Time frame: End of dialysis treatment
The primary objective of this study is to evaluate performance of small-solute (toxin) removal, represented by urea, under standard high flux hemodialysis conditions with a mean urea reduction ratio (URR) of 65% or greater.
Time frame: 8 weeks, 3 sessions per week
Evaluate of all adverse events during the study period
Time frame: 5 sampling at mid-week session of week 1, 2, 3, 5 and 8
Evaluate performance of the B2M reduction ratio
Time frame: 1 sampling at mid-week session of week 7
Evaluate the instantaneous clearance of B2M for the Clearum HS dialyzer at 15-30 min and at 120-150 min at session the mid-week session of week 2 with the Clearum dialyzer.
Time frame: 5 sampling at mid-week session of week 1, 2, 3, 5 and 8
Evaluate the reduction ratio of albumin during the hemodialysis session with the Clearum dialyzer and any trends in albumin concentration over a 6-week period
Time frame: Kinetic over 6 week period
Observe the trends in inflammatory markers CRP over a 6-week period
Time frame: Kinetic over 6 week period
Observe the trends in inflammatory markers IL6 over a 6-week period
Time frame: End of dialysis treatment
Evaluate coagulation parameters as assessed by Thrombin-anti-thrombin (TAT) on the dialyzer after blood restitution at the end of the dialytic session for sessions where sampling occurs
Time frame: End of dialysis treatment
Evaluate on a qualitative photo visual score ranging from poor to very good (1-5) for the residual blood remaining in the dialyzer after rinse back at the end of the session
Mozarc Medical US LLC
Industry
Clearum High Flux Hemodialyzer: Safety and Performance Clinical Study
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT07518628
Behavior, Chronic Disease
Nicosia, Cyprus
View Trial DetailsNCT05154773
Chronic Disease, Chronic Kidney Diseases
Buffalo, New York, United States
View Trial DetailsNCT04466865
Behavior, Chronic Disease
Denver, Colorado, United States
View Trial DetailsNCT02259413
Behavior, Chronic Disease
Winnipeg, Manitoba, Canada
View Trial Details