CB-17-08 CADe
DeviceCB-17-08 Augmented Endoscopy System with Computer Aided Detection (CADe) function
NCT Number: NCT03954548
The aim of this study is to assess the performance of the CB-17-08 to help endoscopists find potential mucosal polyps during the colonoscopy procedure, without significant noise disturbing the endoscopist attention, nor negative interference with the lesions detection than with the standard colonoscopy alone: the study will investigate whether the use of the device provides an increase in the number of adenomas per colonoscopy as compared to standard colonoscopy. The study will also evaluate the safety of the CB-17-08, assessing if the use of the system increases the total number of excisions without a commensurate number of adenomas as compared to standard colonoscopy.
Looking for future studies?
Notify Me45 year and older
All sexes
Interventional
Not applicable
ASL Roma 1 (Presidio Nuova regina Margherita), Roma, Italy
Study procedure:
Patients will be randomized (1:1) to one of the arms by an IWRS system. Randomization will be stratified by study endoscopist, to balance the number of patients randomized to either of the study groups (1:1) within the same endoscopist, by age (from to 45 to <65 years old; ≥65 years old) and by reason for colonoscopy (screening, surveillance <3 years from previous colonoscopy, surveillance 3-10 years from previous colonoscopy). Each endoscopist shall not enrol more than 90 patients.
The following study visits are foreseen for each patient:
Screening Visit A screening visit is performed at the investigational site. During this visit out-patients scheduled for colonoscopy will be informed about the aims, procedures, benefits and possible risks of the study prior to signing the informed consent form for inclusion in the study. Their medical history will be recorded as well as eventual clinical or laboratory examinations according to the local standard of care preparation for a colonoscopy, ) and the date for the colonoscopy procedure to be performed at the investigational site will be scheduled.
Tandem Colonoscopy Visit Each eligible patient returns to the clinic to undergo a same-day, back-to-back tandem colonoscopy examination performed by the same experienced endoscopist. The randomized allocation will determine whether patients will undergo standard high-definition white light colonoscopy with CB-17-08 immediately followed by standard high definition white light colonoscopy or standard high definition white light colonoscopy followed by standard high definition white light colonoscopy with CB-17-08.
Bowel preparation will be done according to the usual standard of care protocols of the individual sites. The quality of bowel preparation will be assessed during colonoscopy using the Boston Bowel Preparation Scale (BBPS). Sedation according to the sites best experience and standard procedures will be delivered to the patient by the endoscopist or an anaesthesiologist.
Subjects' vital signs (blood pressure [BP], hearth rate [HR] and oxygen saturation [SpO2]) will be measured and monitored prior to, during and at the end of the tandem colonoscopy procedure.
The endoscopist will be instructed to adhere to their usual withdrawal technique and to spend a minimum of 6 minutes withdrawing and examining the colonic mucosa. Time to reach the caecum and time to withdrawal from caecum to exit will be recorded for each colonoscopy. Clean withdrawal time, i.e. withdrawal time excluding the time spent for procedures or washings (if any), will be recorded. At least 6 minutes of clean withdrawal time will be required for all colonoscopies, in accordance to the current ASGE guideline. Withdrawal time, total procedure time, and time for pauses to allow polypectomies and biopsies to be performed will be recorded. Repeated examination of any of the colon segments (e.g. right colon) in normal modality or in retroflexion is not permitted.
Each colorectal polyp detected during the first procedure, as well as each polyp detected during the second procedure, will be immediately removed or biopsied and will be sent to the central histology laboratory for characterization. When a mucosal polyp is detected, its estimated size and morphological appearance according to Paris classification will be reported by the endoscopist on the CRF, as well as the anatomical location inside the colonic districts. On the basis of histological examination, polyps will be categorized according to revised Vienna classification and serrated lesion classification.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Fecal Occult Blood Test (FOBT) or Fecal Immunochemical Test (FIT) positive patients will not be excluded from the study.
CB-17-08 Augmented Endoscopy System with Computer Aided Detection (CADe) function
Time frame: Day 1
AMR, defined as the number of histologically confirmed adenomas and carcinomas detected during the second colonoscopy divided by the total number of histologically confirmed adenomas and carcinomas detected during the first and second colonoscopy combined.
Time frame: Day 1
PMR, defined as the number of histologically confirmed polyps detected during the second colonoscopy divided by the total number of histologically confirmed polyps detected during the first plus second colonoscopy combined.
Diminutive (<5 mm) hyperplastic polyps of the rectosigmoid region will not be accounted for in the endpoint analysis.
Cosmo Artificial Intelligence-AI Ltd
Industry
Prospective, Randomized, Multicenter, Tandem Study Evaluating the Safety and Effectiveness of the CB-17-08 Augmented Endoscopy System for the Detection of Mucosal Colorectal Polyps in Adult Patients Undergoing Screening or Surveillance Colonoscopy for CRC.
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT02970513
Bacterial Infections, Bacterial Infections and Mycoses
Clermont-Ferrand, France
View Trial DetailsNCT06068257
Breast Cancer, Breast Diseases
Orlando, Florida, United States
View Trial DetailsNCT07725692
Anastomotic Leak, Colon Cancer
Alexandria, Please Select, Egypt
View Trial DetailsNCT07180303
Colonic Diseases, Colorectal Cancer
Liyang, Jiangsu, China
View Trial Details