NCT Number: NCT05257564
With the rise of cardiovascular diseases (CVD) and diabetes, the global disease burden is shifting towards non-communicable diseases (NCDs). An increasing number of low- and middle-income countries (LMICs) are currently experiencing the double burden of infectious and non-communicable diseases. In order to facilitate a patient-centred approach to healthcare, there is an urgent need to ensure that primary healthcare (PHC) facilities in LMICs are capable of addressing diagnosis and monitoring of non-communicable diseases at the point-of-care (POC). Important minimum parameters for PHC POC diagnosis and monitoring of cardiometabolic diseases are lipids/lipoproteins, glucose, glycated haemoglobin (HbA1c) and serum creatinine, to address cardiovascular disease, diabetes and chronic kidney disease.
While several technologies of multi-parameter POC devices capable of supporting diagnosis and monitoring of cardiometabolic diseases exist, their quantitative accuracy is often not well evaluated outside of the manufacturer's laboratories and published independent evaluations can be rare, particularly in the settings of intended use. These settings are PHC facilities in varying climatic environments and with staff without specialist laboratory training. Our study aims to evaluate the quantitative accuracy of 2 cardiometabolic POC devices in a setting of intended use and performed by the intended user. (Evaluating the quantitative measurements of glucose, HbA1c, total cholesterol and creatinine as measured in a healthcare setting with point-of-care multiparameter devices compared to a laboratory reference method).
Looking for future studies?
Notify Me20 year and older
All sexes
Interventional
Not applicable
Dharān, Sunsari, 56700, Nepal
Independent performance evaluation of POC devices have shown that many devices, regardless of their intended use, do not perform as well as the manufacturers' claims may suggest. There are a variety of reasons for this, including difference in skills levels of personnel who were involved in generating the data for the manufacturer's claims and who actually perform the test in a clinic, limited evaluation of samples from different geographical locations and thus potential interfering substances or less controlled environmental conditions (e.g. dust, heat and humidity).
Independent performance evaluations to assess clinical accuracy, as well as system usability are important to drive adoption of any technology, and even more so, if the technology is intended to move testing outside of the traditional setting, i.e. away from the central laboratory to the point-of-care. Many devices appear ideally suited for certain settings, however when it comes to actual implementation, the users often discover that the device and workflow do not meet their needs. The choice of a suboptimal device means badly invested resources and may lead to inappropriate use, resulting in fewer reliable tests for patients in the absence of alternatives. Data from real-world evaluations in settings of the intended use can support decision makers to select the right device.
The setting for this study will be a primary healthcare facility and a health post in Nepal, where study participants attend the facility in the context of the "Early detection and management of Chronic Kidney Diseases, Hypertension, Diabetes and Cardiovascular disease in Community in Nepal (KHDC-Nepal) program", conducted by the University Hospital of Geneva and the BP Koirala Institute of Health Science in Nepal.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Fingerstick capillary blood will be collected according to each manufacturer's Instructions for use. The fingerstick capillary blood samples will be used immediately on the point-of-care assays. The venous blood will be collected in specific tubes for plasma (GLU), for whole blood (HbA1c) and for serum (CREP, CHOL). Plasma will be separated by centrifugation at the PHC facility and health post and aliquots will be prepared. Serum will be left to clot prior to centrifugation and serum aliquots will be prepared at the PHC facility and health post. The samples will be stored until transfer to the BPKIHS reference lab.
Time frame: 4 months
Estimates of correlation between quantitative measurements of blood glucose, HbA1c, total cholesterol and creatinine collected from capillary blood and tested with point-of-care multiparameter devices versus quantitative measurements of blood glucose, HbA1c, total cholesterol and creatinine collected from venous blood and tested with a laboratory reference method.
Time frame: 4 months
Estimates of limits of agreement for quantitative measurements of blood glucose, HbA1c, total cholesterol and creatinine collected from capillary blood and tested with point-of-care multiparameter devices versus quantitative measurements of blood glucose, HbA1c, total cholesterol and creatinine collected from venous blood and tested with a laboratory reference method.
Time frame: 4 months
Rate of invalid test results and error types for Tascom (POC device)
Time frame: 4 months
Rate of invalid test results and error types for JanaCare(POC device)
Time frame: 4 months
System usability score for Tascom (POC device) SUS Score (>80.3, Grade A, Excellent) SUS Score (68-80.3, Grade B, Good) SUS Score (68, Grade C, Okay) SUS Score (51-68, Grade D, Poor) SUS Score (<51, Grade F, Awful)
Time frame: 4 months
System usability score for JanaCare(POC device) SUS Score (>80.3, Grade A, Excellent) SUS Score (68-80.3, Grade B, Good) SUS Score (68, Grade C, Okay) SUS Score (51-68, Grade D, Poor) SUS Score (<51, Grade F, Awful)
Foundation for Innovative New Diagnostics, Switzerland
Other
Performance and Usability Evaluation of Cardiometabolic Point-of-care Devices in a Target Use Setting
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT06461273
Behavior, Body Weight
Newark, New Jersey, United States
View Trial DetailsNCT05820295
Arrhythmias, Cardiac, Atrial Fibrillation
New York, United States
View Trial DetailsNCT06094920
Albuminuria, CKD
Almelo, Overijssel, Netherlands
View Trial DetailsNCT05813652
Anxiety Disorders, Arthritis
Hamilton, Ontario, Canada
View Trial Details