BioResearch Group Sp. z o.o.
Nadarzyn, Masovian Voivodeship, 05-830, Poland
NCT Number: NCT03752229
The study was a randomized, single blind, uni-center pivotal study in male and female healthy volunteers, to determine capillary blood volume and pain perception obtained in a process of puncturing with different safety lancets.
A group of three hundred (300) Caucasian male and female Subjects was randomized (aged > 18 to < 65 years old) who were in good physical and mental condition. The Subjects had 4 (four) fingers (two fingers from each hand) lanced each with a different type of safety lancet. Overall, 4 (four) different safety lancets were investigated per each Subject.
Overall, 48 (forty eight) different safety lancets (4 different safety lancets per one Subject) were investigated, the capillary blood volume and pain were measured.
Looking for future studies?
Notify Me18 year–65 year
All sexes
Interventional
Not applicable
Nadarzyn, Masovian Voivodeship, 05-830, Poland
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: HTL-STREFA S.A. safety lancet type 610
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: HTL-STREFA S.A. safety lancet type 610
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: HTL-STREFA S.A. safety lancet type 610
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: HTL-STREFA S.A. safety lancet type 553
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: HTL-STREFA S.A. safety lancet type 553
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: HTL-STREFA S.A. safety lancet type 554
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: HTL-STREFA S.A. safety lancet type 555
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: HTL-STREFA S.A. safety lancet type 556
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: HTL-STREFA S.A. safety lancet type 430
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: HTL-STREFA S.A. safety lancet type 430
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: HTL-STREFA S.A. safety lancet type 430
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: HTL-STREFA S.A. safety lancet type 430
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: HTL-STREFA S.A. safety lancet type 430
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: HTL-STREFA S.A. safety lancet type 430
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: HTL-STREFA S.A. safety lancet type 420
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: HTL-STREFA S.A. safety lancet type 420
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: HTL-STREFA S.A. safety lancet type 420
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: HTL-STREFA S.A. safety lancet type 420
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: HTL-STREFA S.A. safety lancet type 420
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: HTL-STREFA S.A. safety lancet type 420
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: HTL-STREFA S.A. safety lancet type 520
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: HTL-STREFA S.A. safety lancet type 520
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: HTL-STREFA S.A. safety lancet type 450
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: HTL-STREFA S.A. safety lancet type 450
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: HTL-STREFA S.A. safety lancet type 450
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: HTL-STREFA S.A. safety lancet type 545
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: HTL-STREFA S.A. safety lancet type 545
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: HTL-STREFA S.A. safety lancet type 546
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: HTL-STREFA S.A. safety lancet type 547
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: HTL-STREFA S.A. safety lancet type 548
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: HTL-STREFA S.A. safety lancet type 549
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: Owem Mumford lancet
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: Owem Mumford lancet
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: Owem Mumford lancet
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: Medicore lancet
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: Medicore lancet
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: Arkray lancet
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: Arkray lancet
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: Medipurpose lancet
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: Medipurpose lancet
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: Medipurpose lancet
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: Sterilance lancet
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: Dynarex lancet
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: Sterilance lancet
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: Dynarex lancet
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: Ypsomed lancet
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: Promismed lancet
Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.
Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.
Other names: Cambridge Sensors lancet
Time frame: The blood volume was noted after 20 seconds
The amount of capillary blood volume collected after a single lancing of the fingertip.
Time frame: The blood volume was noted until the end of bleeding or up to 2 minutes time
The amount of capillary blood volume collected after a single lancing of the fingertip.
Time frame: directly after lancing
The intensity of pain perceived by the patient was assessed with the use of the numeric Visual Analog Scale (VAS). The Subject marked on the scale the number from 0 ("no pain") to 10 ("worst imaginable pain") that represented their perception of pain. A higher score indicated greater pain intensity.
Time frame: after 30 minutes ± 5 min after lancing
The intensity of pain perceived by the patient was assessed with the use of the numeric Visual Analog Scale (VAS). The Subject marked on the scale the number from 0 ("no pain") to 10 ("worst imaginable pain") that represented their perception of pain. A higher score indicated greater pain intensity.
HTL-Strefa S.A.
Industry
A Randomized, Single Blind, Uni-centre Study in Healthy Volunteers, to Determine Capillary Blood Volume and Pain Perception Obtained in a Process of Puncturing With Different Safety Lancets
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT06967714
Behavior, Healthy
Brussels, Belgium
View Trial DetailsNCT06336109
Healthy, Mental Disorders
Aalborg, North Denmark, Denmark
View Trial DetailsNCT06330402
Healthy, Neurologic Manifestations
Aalborg, North Denmark, Denmark
View Trial DetailsNCT04299490
Agnosia, Apnea
Baltimore, Maryland, United States
View Trial Details