NCT Number: NCT03752229

Capillary Blood Volume and Pain Perception Obtained in a Process of Puncturing With Different Safety Lancets

The study was a randomized, single blind, uni-center pivotal study in male and female healthy volunteers, to determine capillary blood volume and pain perception obtained in a process of puncturing with different safety lancets.

A group of three hundred (300) Caucasian male and female Subjects was randomized (aged > 18 to < 65 years old) who were in good physical and mental condition. The Subjects had 4 (four) fingers (two fingers from each hand) lanced each with a different type of safety lancet. Overall, 4 (four) different safety lancets were investigated per each Subject.

Overall, 48 (forty eight) different safety lancets (4 different safety lancets per one Subject) were investigated, the capillary blood volume and pain were measured.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year–65 year

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

BioResearch Group Sp. z o.o.

Nadarzyn, Masovian Voivodeship, 05-830, Poland

Who can participate

Healthy volunteers accepted: Yes

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Willing to sign the informed consent.
  • Male and female volunteers
  • Age between 18 and 65 inclusive
  • Able to communicate well with the Investigator and comply with the requirements of the study
  • Volunteers who are in good physical and mental health

Exclusion criteria

  • Confirmed or suspected malignant cancer
  • Pregnancy or breastfeeding (only for women with childbearing potential based on pregnancy test conducted at Site)
  • History of poor blood circulation.
  • Any skin condition on his or her fingers that prevents blood sampling
  • Anxiety with needles or finger pricks
  • Clotting disorders (including bleeding) in medical history
  • Neuropathy or other condition affecting sensation in the hands.
  • History of blood borne infection (e.g., HIV, hepatitis B or C, syphilis, malaria, babesiosis, brucellosis, leptospirosis, arboviral infections, relapsing fever, T lymphotropic virus Type 1, Creutzfeldt-Jakob disease)
  • Currently participating in another study
  • History of drug or alcohol abuse within the 12 months prior to screening or evidence of such abuse
  • Intake of alcohol within 48 hours prior to the start of the study (self-reporting)
  • Donation or loss of 400 mL or more of blood within 4 weeks prior to the start of the study
  • Any other condition that in the Investigator opinion may negatively influence Subject's participation in the study
  • Intake of medicines that affect blood coagulability (including anticoagulants such as vitamin K, antivirals and anticoagulants such as heparin, aspirin, thrombin inhibitors, vitamin K antagonists).

Treatment and study plan

Acti-Lance Lite

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: HTL-STREFA S.A. safety lancet type 610

Acti-Lance Universal

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: HTL-STREFA S.A. safety lancet type 610

Acti-Lance Special

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: HTL-STREFA S.A. safety lancet type 610

Medlance Plus Super Lite

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: HTL-STREFA S.A. safety lancet type 553

Medlance Plus Lite

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: HTL-STREFA S.A. safety lancet type 553

Medlance Plus Universal

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: HTL-STREFA S.A. safety lancet type 554

Medlance Plus Extra

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: HTL-STREFA S.A. safety lancet type 555

Medlance Plus Special

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: HTL-STREFA S.A. safety lancet type 556

Prolance Micro Flow

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: HTL-STREFA S.A. safety lancet type 430

Prolance Low Flow

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: HTL-STREFA S.A. safety lancet type 430

Prolance Normal Flow

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: HTL-STREFA S.A. safety lancet type 430

Prolance High Flow

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: HTL-STREFA S.A. safety lancet type 430

Prolance Max Flow

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: HTL-STREFA S.A. safety lancet type 430

Prolance Pediatric

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: HTL-STREFA S.A. safety lancet type 430

Haemolance Micro Flow

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: HTL-STREFA S.A. safety lancet type 420

Haemolance Low Flow

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: HTL-STREFA S.A. safety lancet type 420

Haemolnace Normal Flow

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: HTL-STREFA S.A. safety lancet type 420

Haemolance High Flow

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: HTL-STREFA S.A. safety lancet type 420

Haemolance Max Flow

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: HTL-STREFA S.A. safety lancet type 420

Haemolance Pediatric

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: HTL-STREFA S.A. safety lancet type 420

MediSafe Solo 29 Gauge

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: HTL-STREFA S.A. safety lancet type 520

MediSafe Solo 23 Gauge

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: HTL-STREFA S.A. safety lancet type 520

ergoLance Micro Flow

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: HTL-STREFA S.A. safety lancet type 450

ergoLance Normal Flow

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: HTL-STREFA S.A. safety lancet type 450

ergoLance High Flow

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: HTL-STREFA S.A. safety lancet type 450

Medlance (light blue)

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: HTL-STREFA S.A. safety lancet type 545

Medlance (blue)

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: HTL-STREFA S.A. safety lancet type 545

Medlance (dark blue)

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: HTL-STREFA S.A. safety lancet type 546

Medlance (yellow)

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: HTL-STREFA S.A. safety lancet type 547

Medlance (orange)

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: HTL-STREFA S.A. safety lancet type 548

Medlance (red)

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: HTL-STREFA S.A. safety lancet type 549

Unistik 3 Extra

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: Owem Mumford lancet

Unistik 3 Comfort

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: Owem Mumford lancet

Unistik Touch

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: Owem Mumford lancet

ReadyLance (blue)

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: Medicore lancet

ReadyLance (orange)

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: Medicore lancet

AssureLance

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: Arkray lancet

AssureLance MicroFlow

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: Arkray lancet

Surgilance

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: Medipurpose lancet

Surgilance Lite (grey)

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: Medipurpose lancet

Surgilance Lite (purple)

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: Medipurpose lancet

SteriLance Lite2

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: Sterilance lancet

SensiLance (pink)

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: Dynarex lancet

SteriLance Press

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: Sterilance lancet

SensiLance (orange)

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: Dynarex lancet

mylife

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: Ypsomed lancet

VeriFine

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: Promismed lancet

Microdot

Device

Finger lancing was undertaken in a standardized way and was performed by medical staff trained from lancing procedure described in the Protocol.

Total blood collection time was recorded in the source documentation and then volume of the blood was measured using a calibrated ruler.

Other names: Cambridge Sensors lancet

Primary outcomes

  1. Volume (microliters) of blood in the capillary tube

    Time frame: The blood volume was noted after 20 seconds

    The amount of capillary blood volume collected after a single lancing of the fingertip.

  2. Volume (microliters) of blood in the capillary tube

    Time frame: The blood volume was noted until the end of bleeding or up to 2 minutes time

    The amount of capillary blood volume collected after a single lancing of the fingertip.

Secondary outcomes

  1. Pain perception

    Time frame: directly after lancing

    The intensity of pain perceived by the patient was assessed with the use of the numeric Visual Analog Scale (VAS). The Subject marked on the scale the number from 0 ("no pain") to 10 ("worst imaginable pain") that represented their perception of pain. A higher score indicated greater pain intensity.

  2. Pain perception

    Time frame: after 30 minutes ± 5 min after lancing

    The intensity of pain perceived by the patient was assessed with the use of the numeric Visual Analog Scale (VAS). The Subject marked on the scale the number from 0 ("no pain") to 10 ("worst imaginable pain") that represented their perception of pain. A higher score indicated greater pain intensity.

Sponsors and collaborators

Lead sponsor

HTL-Strefa S.A.

Industry

Registry information

Official study title

A Randomized, Single Blind, Uni-centre Study in Healthy Volunteers, to Determine Capillary Blood Volume and Pain Perception Obtained in a Process of Puncturing With Different Safety Lancets

Important dates

Study start
2017
Primary completion
2017
Study completion
2018
First posted
Nov 23, 2018
Registry last updated
Nov 26, 2018