Skip to main content
OpenTrials
Completed

NCT Number: NCT00977314

Bone Conduction Auditory Performance Via the Tooth for Single-Sided Deafness

The purpose of this study is to determine the safety and efficacy of using bone conduction via the teeth to treat Single-Sided Deafness (SSD).

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year–80 year

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Hearing Resource Center, Redwood City, California, United States

Loading trial locations.

About this study

The Sonitus Bone Conduction Device (BCD) is a bone conduction device for single-sided deafness (SSD). The Sonitus BCD consists of an Oral Appliance (OA), an external microphone component, worn behind the ear (BTE), a calibration interface cable and a PC-controlled calibration software for subject calibration. The Sonitus BCD picks up sounds from a microphone located within the ear canal of the deaf ear, capitalizing on the acoustics of the natural pinna and ear canal. The signal picked up by the microphone is then transmitted wirelessly to a removeable bone conduction oral appliance located on the upper molars. The oral appliance receives the acoustic signal from the BTE and applies an equivalent vibratory signal to the teeth that reaches the skull via bone conduction and routed transcranially to the contralateral cochlea.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Must be >18, < 80 years old
  • Must be fluent in English, as determined by the PI
  • Must not be a member of a vulnerable group (IRB defined)
  • Must remain in geographic area during duration of the study
  • Diagnosis of acquired SSD (Section 3.1), time since onset (≥3 mos)
  • Must have a minimum of 6 posterior teeth remaining in the upper arch (3 teeth per side),third molars are acceptable if healthy, fully erupted and substituting for second molars

Exclusion criteria

  • Must not be current users of devices such as Baha, CROS or TransEar
  • Must not have known active medical causes of SSD:
  • Active middle ear pathology
  • Conductive HL (Otosclerosis, otitis media, otitis externa and others)
  • Sudden hearing loss that is not stable
  • Must not have known medical problems that might be life-threatening or is a contraindication for elective dental or medical procedures
  • Must not have known problems that may interfere with the impression procedure, such as inability to breath through nose (e.g. severe flu, allergies or cold)
  • Must not have allergies to polymers
  • Must not have known dental abnormalities:
  • Temporary crowns or undergoing dental treatment
  • Poor oral hygiene and/or rampant decay
  • Current orthodontics
  • Active caries in one or more of the possible abutment teeth for the device
  • Active moderate to severe periodontal disease around abutment teeth for the device
  • Suspicious oral/facial lesions or swelling of any type
  • Severe pain on palpation on any area of mouth, face or neck
  • Moderate to severe heat sensitivity on any of the upper teeth
  • Subject currently being treated for temporomandibular joint disorder (TMJ), pain, full mouth reconstruction or posterior dental implants
  • Must not have known Audiological conditions:
  • Conductive hearing loss (air-bone gap >10dB HL at more than 3 frequencies)
  • Word recognition scores inconsistent with pure tone averages
  • Fluctuating hearing loss
  • Must not have known psychological factors that interfere with their ability to comply, comprehend, consent and cooperate

Treatment and study plan

The Sonitus Bone Conduction Hearing System

Device

Comparison of HINT scores (Speech front with noise at better ear) at 30 days tested with the device in place verses with the device removed.

SoundBite

Device

SoundBite Hearing System

Device

Primary outcomes

  1. Incidence of Device- and Procedure-related Adverse Events at 30 Days

    Time frame: 30 days

    The safety parameters for the trial were monitored throughout the Evaluation Phase (30 days). The safety checks included:

    Comprehensive Medical evaluation at Enrollment and at Termination Comprehensive Dental evaluation at Enrollment and at Termination, with interim dental checks at each visit in between, if needed Comprehensive Audiological evaluation at Enrollment and Termination.

  2. Efficacy: Ability to Understand Speech in Noise

    Time frame: Day 1, Day 30

    The primary efficacy outcome was a measure of the ability to understand speech in noise while wearing the device compared with not wearing the device. The Hearing in Noise Test (HINT) was utilized for this measure as it is the most widely used test for SSD devices. An improvement in HINT score is indicated as a negative (-) dB value change. A more negative (-dB) value indicates an improvement in understanding speech in noise. An improvement in a HINT score of -1 dB is equivalent to a 10% improvement in the ability to understand speech in noise and is likely of clinical benefit. The scores are calculated as HINT Advantage (aided compared with unaided) which depict the differences of using a device as compared to not wearing a device.

Secondary outcomes

  1. Measure of Benefit of SoundBite Using Abbreviated Profile of Hearing Aid Benefit (APHAB).

    Time frame: 30 days

    The measure of the benefit of the device was assessed using the Abbreviated Profile of Hearing Aid Benefit (APHAB), a 24-item self-assessment inventory in which the amount of difficulty in everyday situations is reported with larger numbers indicating more difficulty. Device benefit is calculated by subtracting the score obtained after using a device from the score obtained before using the device. A software program is utilized to score the APHAB and results are compared a different time points. The APHAB is well characterized and broadly used as a quantifiable measurement of device benefit. The APHAB benefit scores can range from -99 (treatment worse than no treatment) to +99 (treatment better than no treatment)

Sponsors and collaborators

Lead sponsor

Sonitus Medical Inc

Industry

Registry information

Important dates

Study start
2009
Primary completion
2010
Study completion
2010
First posted
Sep 15, 2009
Registry last updated
Oct 17, 2014

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.