Biolinq External Interferent Protocol

The primary objective of the study is to evaluate the interference effect of acetaminophen (APAP) on the Biolinq MicroArray Intradermal Continuous Glucose Biowearable System during a single in-clinic day on Day 3 of the sensor wear period.

All participants will be asked to wear the Biolinq System and a commercial CGM comparator for three days while taking fingersticks on a commercial Self-Monitoring Blood Glucose Meter. All participants will come back to the clinic on the third day of sensor wear for an in-clinic day which will include blood draws and fingersticks to compare glucose measurements to a Lab Analyzer. At the end of the in clinic day the subjects will have the sensors removed and then exit the study.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Rocky Mountain Clinical Research

Idaho Falls, Idaho, 83404, United States

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

General

  • ≥ 18 years old.
  • Willing and able to provide written signed and dated informed consent.

Diabetes History and Health

  • Diagnosis of diabetes (Type 1, 2 or LADA)
  • Weigh at least 110 lbs (50 kilograms).
  • Be otherwise in good health, as determined by a medical care professional.
  • Willing to refrain from Acetaminophen (APAP) use for 72 hours prior to Biolinq application and for the duration of the study the duration of study enrollment (except for as administered In-Clinic).

Device and Glucose Assessments - Willing to:

  • Wear one (1) Biolinq sensor following the application procedure on the volar forearm for up to 3 days.
  • Wear one (1) commercial CGM on the abdomen for up to 3 days per approved labeling.
  • Participate in one (1) In-Clinic session lasting up to 8 hours of blood draws (anticipated up to 10 hours on site).
  • Perform up to five (5) fingersticks a day with the SMBG device provided during non-in-clinic days.
  • Avoid immersing study devices into water (e.g., no hot tub, SCUBA diving).

Exclusion criteria

General

  • Current participation in another investigational study protocol. (If a subject has recently completed participation in another drug study, the subject must have completed that study at least 30 days prior to being enrolled in this study.) Note: Subjects will not be excluded if enrolled in another observational trial, wherein the subject is in the follow-up phase and no tests/procedures impacting the subject's health are required. Subjects will be excluded if they have been previously enrolled in this study.
  • Work for, are family members with, or live with someone that works for the sponsor or competitor diabetes-related company (includes social media influencers or bloggers).
  • In the investigator's opinion, any reason that may lead to subject non-compliance with study requirements or confound study data.

Health

  • Currently taking Hydroxyurea or chronic use of a medication containing acetaminophen in the last 30 days.
  • Known allergy to medical grade adhesives, acrylic, latex, or isopropyl alcohol.
  • Known contraindication to taking the In-Clinic recommended oral dose of APAP (e.g., cirrhosis, chronic heavy ethanol use, breast feeding).
  • Have dermatological conditions that preclude wearing Biolinq Biowearables (e.g., extensive psoriasis, recent burns, severe sunburn, extensive eczema, extensive scarring, dermatitis herpetiformis, skin lesions, erythema, infection, or other conditions at the discretion of the investigator).
  • For subjects of child-bearing potential, pregnant or not practicing an acceptable form of birth control during the study.

Hematocrit measurement via point-of-care (POC) or laboratory testing that is less than the applicable below-mentioned value:

a. Male: 36.0% b. Female: 33.0% 10. Have donated blood, had significant blood loss, or participated in a study with significant blood sampling (420 cc or more) within 56 days prior to study enrollment or plan to participate in such activities during study wear.

  • Required or scheduled to have diathermy, X-ray, MRI, or CT during study wear.
  • In the investigator's opinion, the subject has a history of concomitant medical condition that could interfere with the study participation or present a risk to the safety and welfare of the subject or study staff. Such historical conditions include but are not limited to:
  • Syncope in past 6 months
  • Severe hypoglycemia (loss of consciousness, seizure, or emergency medical technician assistance within the past 6 months)
  • Diabetic ketoacidosis (DKA) requiring hospital admission in the past 6 months
  • Coagulopathy
  • Chronic infectious disease (e.g., HIV/AIDS, Hepatitis B or C)
  • End stage renal disease and currently managed by dialysis or anticipating initiating dialysis during the study wear period
  • History of congestive heart failure

Treatment and study plan

The Biolinq MicroArray Intradermal Continuous Glucose Biowearable System

Device

All participants will wear 1 Biolinq System on volar forearm.

Primary outcomes

  1. Number of Samples Where YSI Values Matched the Biolinq Sensor's Color During the First Hour Post Acetaminophen (APAP) Intake.

    Time frame: First hour post Acetaminophen (APAP) intake

    This measures how often (percentage), within the first hour after taking APAP, the Biolinq sensor's color matches the YSI blood sugar readings. BLUE on the sensor means a glucose level between 70 and 180 mg/dL, which matches YSI values from 55 to 207 mg/dL. YELLOW means a sensor reading between 181 and 400 mg/dL, matching YSI values from 154 to 460 mg/dL.

Other outcomes

  1. Number of Participants With Adverse Device Effects

    Time frame: 3 days

    Overall adverse device effect rate. It was pre-specified in the protocol to report adverse event data irrespective of arm randomized for placement of sensor.

  2. Number of Samples Where YSI Values Matched the Biolinq Sensor's Color During the Second Hour Post Acetaminophen (APAP) Intake.

    Time frame: Second hour post Acetaminophen (APAP) intake

    Percentage of correctly displayed color in the second hour post APAP intake; BLUE on the sensor means a glucose level between 70 and 180 mg/dL, which matches YSI values from 55 to 207 mg/dL. YELLOW means a sensor reading between 181 and 400 mg/dL, matching YSI values from 154 to 460 mg/dL.

  3. Number of Samples Where YSI Values Matched the Biolinq Sensor's Color During the Third Hour Post Acetaminophen (APAP) Intake.

    Time frame: Third hour post Acetaminophen (APAP) intake

    Percentage of correctly displayed color in the third hour post APAP intake; BLUE on the sensor means a glucose level between 70 and 180 mg/dL, which matches YSI values from 55 to 207 mg/dL. YELLOW means a sensor reading between 181 and 400 mg/dL, matching YSI values from 154 to 460 mg/dL

  4. Number of Samples Where YSI Values Matched the Biolinq Sensor's Color During the Fourth Hour Post Acetaminophen (APAP) Intake.

    Time frame: Fourth hour post Acetaminophen (APAP) intake

    Percentage of correctly displayed color in the fourth hour post APAP intake; BLUE on the sensor means a glucose level between 70 and 180 mg/dL, which matches YSI values from 55 to 207 mg/dL. YELLOW means a sensor reading between 181 and 400 mg/dL, matching YSI values from 154 to 460 mg/dL

  5. Number of Samples Where YSI Values Matched the Biolinq Sensor's Color During the Fifth Hour Post Acetaminophen (APAP) Intake.

    Time frame: Fifth hour post Acetaminophen (APAP) intake

    Percentage of correctly displayed color in the fifth hour post APAP intake; BLUE on the sensor means a glucose level between 70 and 180 mg/dL, which matches YSI values from 55 to 207 mg/dL. YELLOW means a sensor reading between 181 and 400 mg/dL, matching YSI values from 154 to 460 mg/dL

  6. Number of Samples Where YSI Values Matched the Biolinq Sensor's Color During the Sixth Hour Post Acetaminophen (APAP) Intake.

    Time frame: Sixth hour post Acetaminophen (APAP) intake

    Percentage of correctly displayed color in the sixth hour post APAP intake; BLUE on the sensor means a glucose level between 70 and 180 mg/dL, which matches YSI values from 55 to 207 mg/dL. YELLOW means a sensor reading between 181 and 400 mg/dL, matching YSI values from 154 to 460 mg/dL

  7. Mean of Maximum Bias

    Time frame: Six hours post Acetaminophen (APAP) intake

    The mean of maximum bias (Biolinq minus YSI glucose values) of less than 10 mg/dL computed from maximum biases between the difference in the maximum bias one-hour pre-intake and maximum bias at every hour up to 6 hours post-intake.

Sponsors and collaborators

Lead sponsor

Biolinq Inc.

Industry

Registry information

Official study title

Acetaminophen Interferent Evaluation of the Biolinq MicroArray Intradermal Continuous Glucose Biowearable System

Important dates

Study start
2024
Primary completion
2024
Study completion
2024
First posted
Aug 19, 2024
Registry last updated
Feb 13, 2026

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.