Babson Diagnostics
Austin, Texas, 78753, United States
NCT Number: NCT05186311
This study designed to evaluate the safety and effectiveness of a newly developed sample collection device. Participants will be enrolled and samples will be collected by representative future users of the device. Comparator venous and conventional capillary tubes will be collected by skilled healthcare professionals with phlebotomy experience. Samples will be tested and average differences between the investigational device compared to a marketed comparator will be evaluated. Visual observations will be captured and device safety will be assessed.
Looking for future studies?
Notify Me18 year and older
All sexes
Interventional
Austin, Texas, 78753, United States
The current method for collection of a large volume of capillary blood (up to 600 µL) involves optional warming of the hand or finger prior to procedure, lancing of the fingertip, squeezing the fingertip by hand, and dripping the blood into an open non-sterile tube. The capillary blood collection process is considered cumbersome and lacks standardization in how much pressure is applied in squeezing of the finger which may lead to poor sample quality or insufficient blood volume. Common sample quality issues for capillary blood collection include hemolysis and dilution of sample with interstitial fluid, potentially causing bias in test results. Open collection also carries a risk of blood exposure for healthcare workers and potential risk of infection at the finger puncture site. A new device, BD Microtainer® Easy Collect Capillary Blood Collection System for the collection of capillary blood was created to standardize the current collection process and improve sample quality.
This study designed to evaluate the safety and effectiveness of the BD Microtainer® Easy Collect Capillary Tube SST™ and BD Microtainer® Easy Collect Capillary Tube EDTA devices. This study will be conducted in two parts, one for the SST product and the other for the EDTA product. A minimum of 100 participants, and a maximum of 150 participants per part will be enrolled at a minimum of three various ancillary healthcare site settings representative of the intended use environment such as retail pharmacy site, patient service center, and/or urgent care clinic. Samples will be collected by representative future users of the BD Microtainer® Easy Collect Capillary Blood Collection System who may not have prior blood collection experience. Comparator venous and conventional capillary tubes will be collected by skilled healthcare professionals with phlebotomy experience. Samples will be collected, clotted (serum samples only) and centrifuged (serum samples only) before transportation to the laboratory where they will be tested.
Average differences between the investigational device compared to a marketed comparator will be evaluated. Visual observations for the SST products will be captured which include hemolysis as measured by Plasma Free Hemoglobin (PFH); collection volume; and duration of sample collection. Visual Observations for the EDTA products will be captured which include frequency of occurrence of instrument flags; frequency of occurrence of platelet clump instrument flags; frequency of instrument platelet clump flags with no platelet clump; evaluate whole blood collection volume; and duration of sample collection. Lastly, device safety will be assessed by device and procedure related adverse events.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Test Blood Sample Collection
Other names: BD MiniDraw™ Capillary System
Time frame: within 4 hours of sample collection
To demonstrate clinical equivalence of the SST test device as compared to Venous Comparator Serum and Capillary Comparator for chemistry analytes. This comparison was focused on the analytes with a unit of g/dL
Time frame: within 2 hours of sample collection
To demonstrate clinical equivalence of the EDTA test device as compared to their relevant comparators venous comparator and capillary comparator. This comparison was focused on the analytes with a unit of percentage
Time frame: within 4 hours of sample collection
To demonstrate clinical equivalence of the SST test device as compared to Venous Comparator Serum and Capillary Comparator for chemistry analytes. This comparison was focused on the analytes with a unit of µIU/mL
Time frame: within 4 hours of sample collection
To demonstrate clinical equivalence of the SST test device as compared to Venous Comparator Serum and Capillary Comparator for chemistry analytes. This comparison was focused on the analytes with a unit of mg/dL
Time frame: within 4 hours of sample collection
To demonstrate clinical equivalence of the SST test device as compared to Venous Comparator Serum and Capillary Comparator for chemistry analytes. This comparison was focused on the analytes with a unit of mmol/L
Time frame: within 4 hours of sample collection
To demonstrate clinical equivalence of the SST test device as compared to Venous Comparator Serum and Capillary Comparator for chemistry analytes. This comparison was focused on the analytes with a unit of ng/mL
Time frame: within 4 hours of sample collection
To demonstrate clinical equivalence of the SST test device as compared to Venous Comparator Serum and Capillary Comparator for chemistry analytes. This comparison was focused on the analytes with a unit of U/L
Time frame: within 2 hours of sample collection
To demonstrate clinical equivalence of the EDTA test device as compared to their relevant comparators Venous Comparator and Capillary Comparator. This comparison was focused on the analytes with a unit of 10^3/µL
Time frame: within 2 hours of sample collection
To demonstrate clinical equivalence of the EDTA test device as compared to their relevant comparators Venous Comparator and Capillary Comparator. This comparison was focused on the analytes with a unit of 10^6/µL
Time frame: within 2 hours of sample collection
To demonstrate clinical equivalence of the EDTA test device as compared to their relevant comparators Venous Comparator and Capillary Comparator. This comparison was focused on the analytes with a unit of fL
Time frame: within 2 hours of sample collection
To demonstrate clinical equivalence of the EDTA test device as compared to their relevant comparators Venous Comparator and Capillary Comparator. This comparison was focused on the analytes with a unit of g/dL
Time frame: within 2 hours of sample collection
To demonstrate clinical equivalence of the EDTA test device as compared to their relevant comparators Venous Comparator and Capillary Comparator. This comparison was focused on the analytes with a unit of pg
Time frame: within 4 hours of sample collection
To evaluate sample hemolysis as measured by Plasma Free Hemoglobin (PFH) for BD Microtainer® Easy Collect Capillary Tube SST™ and all comparator tube types: capillary and venous
Time frame: within 4 hours of sample collection
To evaluate serum collection volume of the Test Device SST tube
Time frame: within 4 hours of sample collection
To evaluate duration of sample collection for Test Device SST tube tube and capillary comparator.
Time frame: within 4 hours of sample collection
Evaluation of the frequency of occurrence of all instrument flags (both asterisks and IP messages) in the Test Device EDTA and all comparator tube types: venous and capillary tubes
Time frame: within 4 hours of sample collection
To evaluate whole blood collection volume for the Test Device EDTA
Time frame: within 4 hours of sample collection
To evaluate duration of sample collection for the Test Device EDTA and capillary comparator
Becton, Dickinson and Company
Industry
A Prospective Multi-center Study to Evaluate Clinical Equivalence (Method Comparison) Between the BD MiniDraw™ SST™ Capillary Blood Collection Tube and BD MiniDraw™ EDTA Capillary Blood Collection Tube to Respective Comparators
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT03531866
Chronic Disease, Disease Attributes
Durham, North Carolina, United States
View Trial DetailsNCT02550171
Chronic Disease, Disease Attributes
Manhasset, New York, United States
View Trial DetailsNCT07063823
Celiac Disease, Digestive System Diseases
Stanford, California, United States
View Trial DetailsNCT04436731
Aneurysm, Body Weight
Columbia, Missouri, United States
View Trial Details