NCT Number: NCT01546285

B40 Non-Invasive Blood Pressure Equivalency Study

The purpose of this study is to demonstrate equivalency of the GE Healthcare B40 Patient Monitor noninvasive blood pressure (NIBP) parameter compared to the reference device.

Completed

Looking for future studies?

Notify Me

Key information

Conditions

Age range

3 year–75 year

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Clinimark

Louisville, Colorado, 80027, United States

About this study

Automated non-invasive blood pressure devices have a long history. These devices first made their appearance in the mid-1970's. Since that time, there have been many improvements in the performance of these devices. Specifically, when making changes to the algorithms that control these devices, the goals are: to improve accuracy and precision in the blood pressure (BP) estimates, to reduce the time needed to determine the BP estimates, to increase comfort for the patient, and to work through artifacts and arrhythmias that are often present in the clinical environment.

The purpose of this study is to demonstrate equivalency of the GE Healthcare B40 Patient Monitor noninvasive blood pressure (NIBP) parameter compared to the reference device.

Who can participate

Healthy volunteers accepted: Yes

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Ability to provide written informed consent or have a legally authorized representative including a parent or guardian provide written informed consent
  • Within the age range from older than 3 years old to 75 years old (adult and child)
  • Ability to maintain proper placement of cuff and/or probe and/or sensor

Exclusion criteria

  • Neonates (ages 0-29 days) and infants (30 days to 3 years old) shall be excluded
  • Any subject deemed too unstable, by clinician's discretion, or for another reason deemed unacceptable for study by the clinician
  • Known dysrhythmias (when applicable)
  • Known disease state that compromises circulation to the extremity
  • Vigorous exercise prior to participating in the study
  • Excessive movement or excitability causing false values or no determinations
  • Known allergy to latex when latex products will be in contact with subject
  • Pregnant females

Treatment and study plan

Blood Pressure Readings on B40 Patient Monitor

Device

Simultaneous blood pressure reading on the Investigational B40 patient monitor and DINAMAP PRO1000 patient monitor

Primary outcomes

  1. Difference Between the B40 Monitor and Reference Device DINAMAP PRO1000 on NIBP Measurements

    Time frame: End of each blood pressure reading

    The primary endpoints are the difference in systolic, diastolic and mean blood pressure values between the B40 Monitor and PRO1000. The mean of the difference should be no more than 5mmHg and the standard deviation of the difference should be no more than 8mmHg per the AAMI SP-10 standard.

Sponsors and collaborators

Lead sponsor

GE Healthcare

Industry

Registry information

Important dates

Study start
2012
Primary completion
2012
Study completion
2012
First posted
Mar 7, 2012
Registry last updated
Jun 21, 2013

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.