Attain Stability Quad Left Ventricular Pacing Lead
Devicecardiac stimulation
Other names: 4798 Lead
NCT Number: NCT03099655
The purpose of this clinical study is to evaluate the safety and efficacy of the Attain Stability Quadripolar MRI SureScan Left Ventricular (LV) lead (Model 4798).
Looking for future studies?
Notify Me18 year and older
All sexes
Interventional
Not applicable
University of Calgary, Calgary, Canada
The Attain Stability Quad Clinical Study is a prospective, non-randomized, multi-site, global, investigational Device Exemption (IDE), interventional clinical study. The purpose of this clinical study is to evaluate the safety and efficacy of the Attain Stability™ Quad MRI SureScan LV Lead (Model 4798). This will be assessed through a primary safety and primary efficacy endpoints.
All subjects included in the study will be implanted with a Medtronic market released de novo CRT-P or CRT-D device, compatible market released Medtronic RA and Medtronic RV leads and an Attain Stability Quad MRI SureScan LV Lead (Model 4798).
Up to 471 subjects will be enrolled into the study and up to 471 Attain Stability Quad MRI SureScan LV Lead (Model 4798) implanted, to ensure a minimum effective sample size of 400 Model 4798 leads implanted with 6 months post implant follow up visits (assuming 15% attrition) at up to 56 sites worldwide.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
cardiac stimulation
Other names: 4798 Lead
Time frame: Implant to 6 months post-implant
Subjects free of Model 4798 lead-related complications at 6 months post-implant
-Model 4798 lead will be considered safe if the probability of subjects free from Attain Stability Quad lead-related complications at 6-months post implant is greater than 87% (i.e. the one-sided 97.5% lower confidence bound must be greater than 87%). Complications were defined in the protocol as, "An adverse event that includes the following is considered a complication: Results in death, Involves any termination of significant device function, or Requires an invasive intervention". Relationship of complication to the Attain Stability Quad lead was determined by an independent Clinical Events Committee.
Time frame: 6 months post-implant
The Model 4798 LV lead has sixteen (16) programmable pacing vectors. The endpoint for the primary efficacy objective is whether or not there is at least one Model 4798 LV lead pacing vector with pacing capture voltage thresholds less than or equal to 2.5V.
-Model 4798 lead will be considered effective if the proportion of subjects with at least one Model 4798 LV lead pacing vector with pacing capture voltage thresholds less than or equal to 2.5V at 0.5 ms pulse width (with absence of Phrenic Nerve Stimulation at 5.0 V) at 6 months post-implant is greater than 80%.
Time frame: 6 months post-implant
The co-primary efficacy endpoint is whether or not a second Model 4798 lead configuration has a pacing capture threshold less than or equal to 4V, excluding the pacing vector that is already counted to the primary efficacy.
-Model 4798 lead will be considered effective if the proportion of subjects with at least one additional Model 4798 LV lead pacing vector with pacing capture voltage thresholds less than or equal to 4.0 V at 0.5 ms pulse width (with absence of Phrenic Nerve Stimulation at 5.0 V) at 6 months post-implant is greater than 80%.
Time frame: During procedure
Number of Attain Stability leads successfully implanted subjects divided by number of subjects undergo Attain Stability lead implant procedure
Time frame: During procedure
Summary statistics including mean and standard deviation will be obtained for total implant procedure time (minutes), fluoroscopy time, and cannulation time for each successful implant procedure.
Time frame: Implant to 6 months post-implant
Summary statistics including mean and standard deviation will be obtained for PCT (volt) measured at 6-month post implant visit for each successful implanted subject with valid pacing data collected at 6 months post implant follow-up visit.
Time frame: Implant to 6 months
Summary statistics including mean and standard deviation will be obtained for Pacing Impedance (Ohm) measured at 6-month post implant visit for each successful implanted subject
Time frame: Implant to 6 months
Number of subjects with a given complication from implant through 6 months (183 days) by total number of subjects undergo Attain Stability implant procedure
Medtronic Cardiac Rhythm and Heart Failure
Industry
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT03446651
Cardiovascular Diseases, Heart Diseases
New Haven, Connecticut, United States
View Trial DetailsNCT05806606
Behavior, Cardiovascular Diseases
Hong Kong
View Trial DetailsNCT06847438
Cardiovascular Diseases, Heart Diseases
Atlanta, Georgia, United States
View Trial DetailsNCT07721909
Cardiovascular Diseases, Heart Diseases
Muscat, Oman
View Trial Details