Skip to main content
OpenTrials
Completed

NCT Number: NCT03099655

Attain Stability™ Quad Clinical Study

The purpose of this clinical study is to evaluate the safety and efficacy of the Attain Stability Quadripolar MRI SureScan Left Ventricular (LV) lead (Model 4798).

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

University of Calgary, Calgary, Canada

Loading trial locations.

About this study

The Attain Stability Quad Clinical Study is a prospective, non-randomized, multi-site, global, investigational Device Exemption (IDE), interventional clinical study. The purpose of this clinical study is to evaluate the safety and efficacy of the Attain Stability™ Quad MRI SureScan LV Lead (Model 4798). This will be assessed through a primary safety and primary efficacy endpoints.

All subjects included in the study will be implanted with a Medtronic market released de novo CRT-P or CRT-D device, compatible market released Medtronic RA and Medtronic RV leads and an Attain Stability Quad MRI SureScan LV Lead (Model 4798).

Up to 471 subjects will be enrolled into the study and up to 471 Attain Stability Quad MRI SureScan LV Lead (Model 4798) implanted, to ensure a minimum effective sample size of 400 Model 4798 leads implanted with 6 months post implant follow up visits (assuming 15% attrition) at up to 56 sites worldwide.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Patient meets CRT implant criteria as determined by local regulatory and/or hospital policy (i.e. US subjects should meet CRT device indications per the HRS/ACC/AHA guidelines)
  • Patient (or legally authorized representative) has signed and dated the study-specific Informed Consent Form
  • Patient is 18 years of age or older, or is of legal age to give informed consent per local and national law
  • Patient is expected to remain available for follow-up visits

Exclusion criteria

  • Patient has had a previous unsuccessful LV lead implant attempt
  • Patient has a previous CRT system or LV lead implanted (for example, transvenous or epicardial)
  • Patient is currently implanted with a recalled (i.e. market-withdrawn, recalled or safety alert) RA and/or RV lead
  • Patient has known coronary venous vasculature that is inadequate for lead placement
  • Patient has unstable angina pectoris or has had an acute myocardial infarction (MI) within the past 30 days
  • Patient has had a coronary artery bypass graft (CABG) or percutaneous transluminal coronary angioplasty (PTCA) within the past 90 days
  • Patient has contraindications for standard transvenous cardiac pacing (e.g., mechanical right heart valve)
  • Patient has had a heart transplant (patients waiting for heart transplants are allowed in the study)
  • Patient has known renal insufficiency that would prevent them from receiving an occlusive venogram during the implant procedure
  • Patient is contraindicated for <1mg dexamethasone acetate
  • Patient is enrolled in any concurrent drug and/or device study that may confound the results of this study
  • Patient has a terminal illness and is not expected to survive more than six months
  • Patient meets exclusion criteria required by local law (e.g. age, pregnancy, breast feeding, etc.)
  • Patient is unable to tolerate an urgent thoracotomy

Treatment and study plan

Attain Stability Quad Left Ventricular Pacing Lead

Device

cardiac stimulation

Other names: 4798 Lead

Primary outcomes

  1. Lead Complication-free Rate at 6 Months

    Time frame: Implant to 6 months post-implant

    Subjects free of Model 4798 lead-related complications at 6 months post-implant

    -Model 4798 lead will be considered safe if the probability of subjects free from Attain Stability Quad lead-related complications at 6-months post implant is greater than 87% (i.e. the one-sided 97.5% lower confidence bound must be greater than 87%). Complications were defined in the protocol as, "An adverse event that includes the following is considered a complication: Results in death, Involves any termination of significant device function, or Requires an invasive intervention". Relationship of complication to the Attain Stability Quad lead was determined by an independent Clinical Events Committee.

  2. Proportion (Reported as a Percent) of Subjects With at Least One Vector Having PCT ≤2.5 V at 0.5 ms Pulse Width and Absence of Phrenic Nerve Stimulation (PNS)

    Time frame: 6 months post-implant

    The Model 4798 LV lead has sixteen (16) programmable pacing vectors. The endpoint for the primary efficacy objective is whether or not there is at least one Model 4798 LV lead pacing vector with pacing capture voltage thresholds less than or equal to 2.5V.

    -Model 4798 lead will be considered effective if the proportion of subjects with at least one Model 4798 LV lead pacing vector with pacing capture voltage thresholds less than or equal to 2.5V at 0.5 ms pulse width (with absence of Phrenic Nerve Stimulation at 5.0 V) at 6 months post-implant is greater than 80%.

  3. Proportion (Reported as a Percent) of Subjects With at Least One Extra Vector Having PCT ≤ 4.0V at 0.5 ms Pulse Width and Absence of Phrenic Nerve Stimulation (PNS)

    Time frame: 6 months post-implant

    The co-primary efficacy endpoint is whether or not a second Model 4798 lead configuration has a pacing capture threshold less than or equal to 4V, excluding the pacing vector that is already counted to the primary efficacy.

    -Model 4798 lead will be considered effective if the proportion of subjects with at least one additional Model 4798 LV lead pacing vector with pacing capture voltage thresholds less than or equal to 4.0 V at 0.5 ms pulse width (with absence of Phrenic Nerve Stimulation at 5.0 V) at 6 months post-implant is greater than 80%.

Secondary outcomes

  1. Implant Success

    Time frame: During procedure

    Number of Attain Stability leads successfully implanted subjects divided by number of subjects undergo Attain Stability lead implant procedure

  2. Implant Duration

    Time frame: During procedure

    Summary statistics including mean and standard deviation will be obtained for total implant procedure time (minutes), fluoroscopy time, and cannulation time for each successful implant procedure.

  3. LV Pacing Capture Threshold (PCT) at 6 Months

    Time frame: Implant to 6 months post-implant

    Summary statistics including mean and standard deviation will be obtained for PCT (volt) measured at 6-month post implant visit for each successful implanted subject with valid pacing data collected at 6 months post implant follow-up visit.

  4. LV Impedance at 6 Months

    Time frame: Implant to 6 months

    Summary statistics including mean and standard deviation will be obtained for Pacing Impedance (Ohm) measured at 6-month post implant visit for each successful implanted subject

  5. Post Implant Lead Failure Rate

    Time frame: Implant to 6 months

    Number of subjects with a given complication from implant through 6 months (183 days) by total number of subjects undergo Attain Stability implant procedure

Sponsors and collaborators

Lead sponsor

Medtronic Cardiac Rhythm and Heart Failure

Industry

Registry information

Important dates

Study start
2017
Primary completion
2018
Study completion
2019
First posted
Apr 4, 2017
Registry last updated
Jan 9, 2020

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.