Uppsala University Hopsital, Dep. of Anesthesia and Intensive Care
Uppsala, Uppsala County, 751 85, Sweden
NCT Number: NCT01506284
The aims of the present study are:
1. To evaluate the effect of the induction of anaesthesia and paralysis in terms of changes in oscillatory mechanics parameters 2. To evaluate the mechanical properties of the respiratory system in terms of input and transfer oscillatory impedance in response to PEEP changes
Looking for future studies?
Notify Me18 year–80 year
All sexes
Observational
Uppsala, Uppsala County, 751 85, Sweden
It has been recently shown that respiratory system reactance (Xrs) obtained by the forced oscillation technique (FOT) at 5 Hz is more reliable than dynamic compliance for assessing lung collapse and the effects of lung RMs in a porcine ALI model ( Ref.1,2).
Specifically, Xrs (and its derived variable CX5, the oscillatory compliance at 5 Hz) identifies the minimum positive end-expiratory pressure (PEEP) level required to maintain lung recruitment with high sensitivity and specificity. Moreover, it has been recently demonstrated that Xrs may be used to identify the lowest level of PEEP able to prevent atelectasis and that PEEP setting strategy based on maximizing Xrs is able to limit lung injury compared to oxygenation-based approach in a porcine lavage model of lung injury. ( Ref.3)
Recently, at the biomedical engineering department of Politecnico di Milano measurements of chest wall displacement have been successfully performed by means of an optical sensor realized using a laser self-mixing interferometer (LIR). The advantage of this approach is that it is contact-less, that by deflecting the laser been it is possible to scan any region of the chest wall surface and that it allows to measure also low-frequency vibrations.
Protocol
At each stage the following measurements will be performed:
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
FOT: The stimulating signal is generated by an A/D-D/A board and amplified by a power amplifier that drives a loudspeaker the output of which is connected to inspiratory line of the ventilator. Pressure and flow are measured at the inlet of the endotracheal tube by a piezoresistive pressure transducer and a mesh-type heated pneumotachograph coupled with a differential pressure transducer.
LIR: The scanner unit is made by three laser interferometers, the beams of which are deflected on the chest wall surface by a mirror moved by a stepper motor triggered on the respiratory pattern of patient.
Time frame: baseline and 5 minutes of ventilation
FOT measurements:
Respiratory system impedance will be computed from the flow and pressure signals measured at the inlet of the tracheal tube. A composite waveform including 5, 11 and 19 Hz will be used as a stimulating signal generated by an A/D-D/A board and amplified by a power amplifier that drives a loudspeaker the output of which is connected to inspiratory line of the ventilator.
Time frame: baseline and 5 minutes of ventilation
Laser interferometry (LIR):
The movement of 18 points of the chest wall will be measured by a self-mixing interferometer. The measurement points will be chosen along three vertical lines (the midline and the two parasternal lines) and will be equally distributed from the clavicles to the anterior superior iliac spines. The chest wall movement will be measured in response to two different stimulating waveforms: a composite waveform including 5, 11 and 19 Hz, a sinusoidal signal at 100 Hz.
Time frame: baseline and 5 minutes of ventilation
Arterial blood gas measurement (ABG)
Time frame: baseline and 5 minutes of ventilation
Uppsala University Hospital
Other
Assessment of Airways Mechanical Properties by Forced Oscillatory Technique (FOT) and Laser Interferometry (LIR) During Anesthesia.
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT04879290
Anesthesia, Atelectasis
Caen, France
View Trial DetailsNCT05672329
Anesthesia, Apnea
Bern, Canton of Bern, Switzerland
View Trial DetailsNCT05345743
Anesthesia, Atelectasis
Amsterdam, North Holland, Netherlands
View Trial DetailsNCT04566419
Anesthesia, Gynecologic Cancer
Rome, Italy
View Trial Details