CHU de Nîmes
Nîmes, Gard, 30029, France
NCT Number: NCT04360304
Acute circulatory failure (ACF) is a common cause of admission in intensive care unit (ICU). Echocardiography is a widespread tool nowadays for the initial assessment and the hemodynamic monitoring. An interesting data from this exam is the Left ventricular outflow tract (LVOT) velocity time integral (VTI), reflecting stroke volume, and therefore cardiac output. A new tool for automated recording has been developped on the VENUE GE echograph. This study aims at assessing this automated measurement of LVOT VTI compared with the classic manual method.
Looking for future studies?
Notify Me18 year and older
All sexes
Interventional
Not applicable
Nîmes, Gard, 30029, France
Acute circulatory failure is a dramatically common issue in critical care, affecting nearly 1/3 of the patients admitted in ICU, and associated with a high morbi-mortality. Echocardiography enables a quick point of care assessment of the hemodynamic status of the patient. One key parameter is stroke volume, which can be estimated by LVOT VTI.
In practice, the guidelines recommend measuring LVOT VTI in a 5 chambers cavity window with pulsed wave doppler, trying to get an angle between aortic flow and the doppler signal as cloth as possible to 0°. Than VTI must be calculated from an average of 3 to 5 cycles in sinus rhythm, or 5 to 10 cycles in arrythmia such as atrial fibrillation. The main issue is that this approach is time consuming and tedious, and most of the practitioners choose to measure only the most representative beat (" Best VTI "), likely to be source of a loss of reproducibility.
In order to make it easier, some companies have developed softwares to help the intensivist. Thereby, General Electric has equiped the VENUE echograph with a system that helps first selecting the proper place for sampling, and then automatically acquire, trace, and calculate the average LVOT VTI based on a 4 seconds sample. Finally, this tool gives a VTI trending when several measures are repeated along the patient' stay.
This program has already been tested on a swine model in hemorrhagic shock with encouraging results (better agreement with cardiac output by thermodilution with the automated method than with the manual one). However, feasibility was only 60%.
This human study aims at assessing the interest of automated LVOT VTI measurement compared to the manual standard one when performing a cardiac echography for hemodynamic assessment of patients hospitalised in ICU. Otherwise, a measure of the " Best-VTI " will be done at the same time in accordance with the current practice, in order to assess this approach.
Finally, when a fluid resuscitation will be indicated, a second exam will be done, and agreement of VTI variation between automated and manual method studied.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
SAP < 90mmHg; Urine output < 0,5mL/kg/h during more than 2 hours; Blood lactate level > 2mmol/L; Increase of Norepinephrine doses needed
Exclusion criteria
An echocardiographic exam is realised after inclusion, measuring especially :
For patients that need fluid resuscitation of at least 250mL in less than 10minutes (indication at the discretion of the physician in charge of the patient), a second echocardiography is done according to the same protocol described before.
Time frame: DAY 0
Agreement between the average LVOT VTI obtained by the manual method, and with the automatic one, estimated through calculation of Lin's coefficient and Bland-Altman plot.
Time frame: DAY 0
Time frame: DAY 0
-Number of the event: Impossibility for the device giving a estimation of mean LVOT VTI
Time frame: DAY 0
-Number of the event: < 3 spectral patterns traced in patients in sinus rhythm
Time frame: DAY 0
-Number of the event: < 5 spectral patterns traced in patients in Atrial Fibrilation
Time frame: DAY 0
-Number of the event: One or more automated spectral tracing by the machine, judged as inappropriate by the expert
Time frame: DAY 0
-Number of the event: Mean LVOT VTI obtained thanks to non-consecutive traces
Time frame: DAY 0
Time frame: DAY 0
Time frame: DAY 0
Time frame: DAY 0
Time frame: DAY 0
Centre Hospitalier Universitaire de Nīmes
Other
Assessment of Agreement Between Two Measurement Methods of Left Ventricular Outflow Tract (LVOT) Velocity Time Integral (VTI) : Automatic Versus Manual, in Critically Ill Patients in Acute Circulatory Failure
Acronym: AUTO-VTI
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT02050269
Acute Circulatory Failure, Pathologic Processes
Orléans, France
View Trial DetailsNCT03680963
Acute Circulatory Failure, Pathologic Processes
Argenteuil, France
View Trial DetailsNCT03788772
Acute Circulatory Failure, Infections
Paris, France
View Trial DetailsNCT05847998
Acute Circulatory Failure, Disease
Bron, France
View Trial Details