Radiometer Turku Oy
Turku, 20750, Finland
NCT Number: NCT06834880
The purpose of this study is to establish the overall (age- and sex independent) at least 95th percentile upper reference limit (URL) of the NTproBNP2 Test Kit on the AQT90 FLEX analyzer.
Looking for future studies?
Notify Me18 year and older
All sexes
Observational
Turku, 20750, Finland
The NTproBNP2 Test is an in vitro diagnostic assay for the quantitative determination of N-terminal pro-B-type natriuretic peptide {NT-proBNP) in EDTA or lithium heparin whole blood or plasma specimens on the AQT90 FLEX analyzer in point of care/ near-patient testing and laboratory settings. It is intended for use as an aid in the diagnosis of heart failure. The test is also intended for use as an aid in the risk stratification of patients with acute coronary syndrome and heart failure.
The NTproBNP2 CAL Cartridge is an in vitro diagnostic reagent intended for the calibration adjustment of the NTproBNP2 Test on the AQT90 FLEX analyzer by establishing points of reference to estimate NT-proBNP values.
The AQT90 FLEX analyzer is an immunoassay instrument based on the quantitative determination of time-resolved fluorescence to estimate the concentrations of clinically relevant markers on whole-blood and plasma specimens to which a relevant anticoagulant has been added. It is intended for use in point of care/ near-patient testing and laboratory settings.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Answered yes to any of the conditions below (Information gathered from a self-reported health questionnaire or results from blood sample tests):
Sample bank with Health Questionnaire
Time frame: Baseline
The primary endpoint of this study is to establish at least 95th percentile of normal concentrations with 95% confidence intervals (CI) for overall samples for the AQT90 FLEX NTproBNP2 Test Kit by analyzing LiHep plasma samples collected from apparently healthy subjects.
Time frame: Baseline
The secondary endpoint would be to establish the 95th percentile for normal concentrations separately for both males and females with the 95% confidence intervals.
Radiometer Medical ApS
Industry
AQT90 FLEX NTproBNP2 Test Kit - 23004 - Reference Interval Study Protocol
Acronym: RI
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT03446651
Cardiovascular Diseases, Heart Diseases
New Haven, Connecticut, United States
View Trial DetailsNCT05806606
Behavior, Cardiovascular Diseases
Hong Kong
View Trial DetailsNCT06847438
Cardiovascular Diseases, Heart Diseases
Atlanta, Georgia, United States
View Trial DetailsNCT07721909
Cardiovascular Diseases, Heart Diseases
Muscat, Oman
View Trial Details