Nantes University Hospital
Nantes, Loire-Atlantique, 44093, France
NCT Number: NCT06032169
The upper arm is sometimes inaccessible for arterial pressure (AP) measurement. Placing the automatic cuff at the ankle is a common alternative. However, whether AP measurement at the ankle is reliable is uncertain. Furthermore, it is unknown whether it is necessary to place the patient in a horizontal position to improve the reliability of the measurement, or if leaving the patient in default position (often semi-recumbent) is acceptable. The aim of this study is to answer these questions. Additionally, investigators will evaluate whether a smartphone application that allows for the simple measurement of AP (by gently placing a finger on the camera lens) is a valuable alternative to upper arm measurements.
Looking for future studies?
Notify Me18 year and older
All sexes
Observational
Nantes, Loire-Atlantique, 44093, France
A question that arises relatively frequently is "when the upper arm is inaccessible, where should the investigators place the automatic cuff to measure AP?". Indeed, surgical intervention involving the upper limb, a wound, a fracture, or vascular access (temporary catheter or arteriovenous fistula) often makes the arm unsuitable for monitoring AP with an automatic cuff. The arterial catheter is not devoid of risks (including the loss of precious time for its insertion) and is rarely an appropriate alternative, at least in the first instance. Thus, the inaccessibility of the upper arm legitimates the frequent placement of the automatic cuff at the ankle. However, the automatic cuff has only been validated for placement at the upper arm, not at the ankle. Surprisingly, the reliability of ankle AP measurements obtained by this positioning of the cuff at the ankle has been little studied: only two studies, using arterial catheter as the reference method, are available, one of which was in the specific population of patients with obesity. These studies reported that the measurement error at the ankle was excessive, but confirmation is necessary.
Patients, especially in acute care settings, are rarely in a strictly horizontal supine position. For various reasons, the critically ill patient is rather in a semi-recumbent position, meaning that the angle between the trunk and the horizontal axis is between 30 and 60°. Thus, the cuff placed at the ankle is no longer at the level of the heart and more precisely of the phlebostatic axis. This could modify the hydrostatic pressure and lead to overestimation of AP compared to a measurement of AP at heart level (on the arm or invasively, for example). Indeed, the impact on AP measurement of the height of the upper arm relative to the heart has already been demonstrated: the vertical displacement of the upper arm changes the measured AP value. Does the angle of elevation of the trunk significantly modify the AP measured at the ankle? If so, and since the measurement at the ankle in the horizontal supine position (0°) tends to underestimate AP, do the different sources of measurement error "cancel out" in the semi-recumbent position? In what position of the trunk is it preferable to measure AP at the ankle? The answers to these questions are uncertain. Indeed, the only study that investigators are aware of (a pilot study) and that specifically addressed this issue did not provide clear answers. This is important since AP is a crucial parameter of monitoring and therefore of therapeutic decision-making. It is remarkable that in acute care settings, little to no attention is paid to trunk elevation when measuring BP at the ankle, and of course, there is no automation of patient positioning (e.g., in a horizontal position) prior to each automatic measurement at the ankle.
Therefore, it appears important to evaluate the accuracy and precision of non-invasive ankle measurements of AP and the impact of patient position on it. This will be the main objective of the study. If these ankle measurements prove to be neither accurate nor precise, having an alternative would be desirable. It is noteworthy that when the arm is inaccessible, the finger often is. AP measurement at the finger is now possible with a simple smartphone. This finger measurement requires calibration with one or two automatic cuff measurements, which should be possible in most cases (the arm is frequently inaccessible for prolonged AP monitoring, but still allows for one or two isolated measurements). So, which is the better alternative site for AP measurement, the finger or ankle? AP measurement using a smartphone has never been evaluated from this perspective. This will be a secondary objective of the study.
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Patients of a surgical intensive care unit having an arterial catheter undergoing noninvasive measurements of AP in the two studied positions (semi-recumbent or horizontal).
Time frame: 15 minutes
The measurement error will be considered acceptable if the current international standard, the AAMI/ESH/ISO standard, is met: mean error (bias) ≤ 5.0 mmHg and its standard deviation ≤ 8.0 mmHg.
The reference AP will be measured using an arterial catheter, taking advantage of the fact that many critically ill patients already have one.
Time frame: 15 minutes
The AAMI/ESH/ISO standard will indicate if the reliability is acceptable: mean error (bias) ≤ 5.0 mmHg and its standard deviation ≤ 8.0 mmHg.
Time frame: 15 minutes
The AAMI/ESH/ISO standard will indicate if the reliability is acceptable: mean error (bias) ≤ 5.0 mmHg and its standard deviation ≤ 8.0 mmHg.
Time frame: 15 minutes
The AAMI/ESH/ISO standard will indicate if the reliability is acceptable: mean error (bias) ≤ 5.0 mmHg and its standard deviation ≤ 8.0 mmHg.
Time frame: 15 minutes
Via a dedicated error grid, the risk associated with measurement errors was classified into one of the 5 risk levels ranging from "no risk" to "dangerous risk". Distribution across the different risk levels was compared by Fisher exact tests.
Time frame: 15 minutes
The ability of averaged noninvasive measurements to detect hypotension (invasive mean Arterial Pressure <65 mmHg, systolic Arterial Pressure <90 mmHg), hypertension (invasive mean Arterial Pressure >100 mmHg, systolic Arterial Pressure >140 mmHg), and a significant therapy-induced change in invasive mean Arterial Pressure (>10%) was determined through area under the receiver operating characteristic curve (AUCROC) analysis.
Nantes University Hospital
Other
NON-INVASIVE MEASUREMENT OF ARTERIAL PRESSURE AT THE ANKLE: IMPACT OF POSITION, SEMI-RECUMBENT OR HORIZONTAL. A SINGLE-CENTER PROSPECTIVE STUDY.
Acronym: 2PAC
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT06556914
Infections, Inflammation
Samsun, Ilkadım, Turkey (Türkiye)
View Trial DetailsNCT07151716
Confusion, Delirium
Beijing, Beijing Municipality, China
View Trial DetailsNCT04014920
Body Weight, Intensive Care Unit
Montpellier, Herault, France
View Trial DetailsNCT04964492
Cardiovascular Diseases, Hemodynamics
Amiens, France
View Trial Details