NCT Number: NCT03595501
Clinical and analytical tests will be performed based on risk assessment and system specifications to verify that the performance of the investigational device is in accordance with its specifications.
Looking for future studies?
Notify Me3 month and older
All sexes
Observational
New York, 10032, United States
Comparison and precision studies will be conducted in up to 5 clinical sites in Israel and the US, and include:
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
For Residual Samples:
For Prospectively collected samples:
Exclusion criteria
Exclusion criteria
post blood draw and pre sample scan - For whole blood samples:
Exclusion criteria
post sample scan:
Complete blood counts from OLO will be determined from analysis of whole blood samples
Complete blood counts from Predicate will be determined from analysis of whole blood samples
Time frame: 3 months
Reproducibility studies will be conducted using 3 levels of commercial control materials (low, normal and high) to measure all CBC reported parameters. Control material will be run on two instruments at each site. Standard deviation (SD) and coefficient of variation (CV) will be calculated for each measurand for: (1) Between lots/sites; (2) Between instruments; (3) Between days; (4)Between operators/runs; (5) within-run and (6) total variability
Time frame: 3 months
Performance of the OLO device will be measured through repeatability of 20 replicates for 11 residual samples on each site. Samples collected for this study will include 4 within lab reference range, 3 around lower medical decision levels for HGB, PLT and WBC, and 4 around the upper range for RBC, HGB, WBC and PLT to cover all pathological levels and medical decision points. Standard deviation (SD) and coefficient of variation (CV) will be computed for each measurand per sample by site.
Time frame: 3 months
Paird capillary and venous whole blood samples will be collected. The samples will be analyzed in duplicate on the Sight OLO device. Analysis will include Passing-Bablok Regression analysis per parameter (Bland Altman plots, slope, intercept, with 95% confidence intervals, correlation coefficient, and % bias), between: The average of venous whole blood samples scans and average of capillary whole blood sample scans from the same individual on the Sight OLO device
Time frame: 3 months
Evaluate the performance of the OLO device in comparison to values achieved with the predicate. Analysis will include Regression parameters (slope, intercept with 95% CI) between measurement of Sight OLO and measurement on predicate (Correlation coefficient, Bland Altman plots and overall % bias between predicate and Sight OLO device including % bias at medical decision points).
Time frame: 3 months
Establish adult venous and fingerprick reference intervals for the OLO device. The non-parametric method will be used to calculate the lower and upper limits of the reference range.
Sight Diagnostics
Industry
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT02835937
Ambulatory Care, Anemia
Scottsdale, Arizona, United States
View Trial DetailsNCT00472290
Blood Platelet Disorders, Bone Marrow Diseases
View Trial DetailsNCT01300026
Cancer, Chronic Disease
Hackensack, New Jersey, United States
View Trial DetailsNCT02229461
Hematology
Austin, Texas, United States
View Trial Details