The purpose of this study is to evaluate the oxygen saturation accuracy performance of Belun Ring Oximeter placed on the index fingers during non-motion conditions
Looking for future studies?
Notify Me18 year–50 year
All sexes
Observational
Louisville, Colorado, 80027, United States
The purpose of this study is to evaluate the SpO2 accuracy performance of Belun Ring Oximeter placed on the index fingers during non-motion conditions over the range of 70-100% SaO2, arterial blood samples, assessed by CO-Oximetry. It is expected that the Accuracy Root Mean Square (ARMS) performance of the oximetry system will meet the required specification of ARMS of 3.5% or less allowed for reflectance technology in non-motion conditions for the range of 70 - 100% SaO2.
The Clinimark proprietary Desaturation Fixture with Automated Data Collection is a single limb blow by system used to deliver medical grade oxygen and nitrogen gas mixtures to induce various hypoxic levels in subjects at a slow steady rate allowing an automatic marking and collection of the Control or secondary Transfer reference Pulse Oximeter and other pulse oximetry systems at 1 second intervals.
The Control Pulse Oximeter, an FDA cleared device, is used to monitor the oxygen saturation levels real time throughout the study for subject safety and to target stable plateaus. This device is used to assess the stability of the data.
Multi-parameter monitor used during the study to observe a subject's vital signs including ECG tracing, heart rate, respiratory rate, end-tidal CO2 with capnograph, secondary monitor for the oxygen concentration being delivered to the subject.
A whole blood analyzer (CO-Oximeter) is used as the reference standard device for obtaining the functional SaO2 value from arterial blood samples obtained during the study.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
The Pulse Oximeter are placed on the left and right Index fingers of each subject of the healthy adult volunteer group to evaluate the SpO2 accuracy performance.
Time frame: 3 days
Subjects were given medical grade mixtures of oxygen and nitrogen to induce stable plateaus across the range of 100% to 70%. The stable plateaus allowed data collection in the following SaO2 ranges 95-100, 90-95, 85-90, 80-85, 75-80, 67-75. In general, 4 to 8 discrete points were collected at each of the levels.
Time frame: 3 days
Data analysis follows ISO80601-2-61:2011, Annex EE and Pulse Oximeters - Premarket Notifications Submissions [510(k)s] Guidance For Industry and Food and Drug Administration Staff (issued: March 4, 2013)
Belun Technology Company Limited
Other
Accuracy Validation of Belun SpO2 Pulse Oximeter FDA Submission Study
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT07720960
Erythropoiesis, Hypoxia
Elche, Alicante, Spain
View Trial DetailsNCT05767385
Cardiovascular Abnormalities, Cardiovascular Diseases
San Francisco, California, United States
View Trial DetailsNCT01542177
Digestive System Diseases, Digestive System Neoplasms
Toronto, Ontario, Canada
View Trial DetailsNCT05354648
Arterial Occlusive Diseases, Arteriosclerosis
View Trial Details