NCT Number: NCT06850610

Accuracy Of The Owlet OSS 3.0 During Standardized Stable Hypoxia Plateaus

This study tests the accuracy of pulse oximeters in a range of arterial oxygen levels from 100% down to 70%. This is done by comparing the test pulse oximeter readings of young, healthy, adult study participants with blood samples drawn from an artery in the wrist during brief plateaus of progressively lower oxygen saturations. The arterial blood sample is processed in a device called a multi-wavelength hemoximeter and compared to the simultaneous oximeter reading. During each set level of oxygen the participant may be asked to have a motion simulator device apply different types of movement to their hands to assess the pulse oximeter's accuracy during motion. This study will be evaluating the Owlet OSS 3.0 sensor, a component in 2 FDA cleared noninvasive pulse oximeters indicated for infants 1-18 months of age.

Testing should require approximately an hour and 15 minutes of the participant's time.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year–49 year

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Vital Signs Research Group

San Francisco, California, 94107, United States

About this study

Several separate cohorts of study participants will be recruited and enrolled in accordance with laboratory time and staff availability. After informed consent is performed, a local anesthetic will be injected around the radial artery, and a 22-gauge radial artery catheter will be placed. Pulse oximeters/sensors under test will be attached to the subject along with FDA-cleared reference pulse oximeters. ECG electrodes will be placed on the subject to monitor the subject's heart rate. The Owlet Sensor under test placement and serial numbers will be documented for each subject. Subjects will be placed in a seated, semi-reclined position. Subjects will breathe a nitrogen-air-carbon dioxide mixture to produce the desired level of hypoxemia.

Each subject will undergo two cycles of three different hypoxia plateaus in the 90-100%, 80-90% and 70-80% range, with an average of 20-24 data pairs with a maximum range of 17-30 data points. Study subjects may be have a motion protocol applied to test the Owlet Sensor's accuracy during variable frequencies and amplitudes of movement.

The motions applied by the motion simulator will be 1) 2 Hz sine 10mm 2) 3 Hz sine 10mm 3) 4 Hz sine 10mm 4 ) Random motion 30mm peak, 20 to 200 mm / sec These motions are applied with the fingers tapping on the first cycle of hypoxia and rubbing on the second cycle. The subject will not perform the motion during the first blood sample of each plateau but will during the 2-4th samples separated by 30 seconds.

90% of participants will have at least one usable data point below 85% and at least 69% of subjects will have one usable data pair between 70-80%. As this procedure is evaluating a single user device, a new sensor will be applied to each study participant.

The skin tone distribution of subjects will be divided into 3 groups based on the 10 point Monk skin tone scale (MST) and Individual Typology Angle (ITA) measured by an ISO compliant colorimeter at the forehead and at the anatomic location of sensor placement. MST will be assessed with a printed color swatch that has been verified by a colorimeter. Enrollment of subjects must follow these criteria:

  • At least 25% of subjects will be Monk Skin Tone (MST) 1-4 with an ITA greater than 30
  • At least 25% of subjects will be MST 5-7 and have an ITA between -30 and 30
  • At least 25%-of subjects will be MST 8-10 with an ITA less than -30 (with one half of MST 8-10 subjects having an ITA of less than -50)

Who can participate

Healthy volunteers accepted: Yes

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • The subject is in good general health with no evidence of any medical problems.
  • The subject is fluent in both written and spoken English.
  • The subject has provided informed consent and is willing to comply with the study procedures.

Exclusion criteria

  • The subject is obese (BMI>35).
  • The subject has a known history of heart disease, lung disease, kidney or liver disease.
  • Diagnosis of asthma, sleep apnea, or use of CPAP.
  • The subject has diabetes.
  • The subject has a clotting disorder.
  • The subject has a hemoglobinopathy or history of anemia, per subject report or the first blood sample, that in the opinion of the investigator, would make them unsuitable for study participation.
  • The subject has any other serious systemic illness.
  • The subject has a carboxyhemoglobin level greater than 3% (determined during the first sample)
  • Any injury, deformity, or abnormality at the sensor sites that, in the opinion of the investigators' would interfere with the sensors working correctly.
  • The subject has a history of fainting or vasovagal response.
  • The subject has a history of sensitivity to local anesthesia.
  • The subject has a diagnosis of Raynaud's disease.
  • The subject has unacceptable collateral circulation based on exam by the investigator (Allen's test).
  • The subject is pregnant, lactating or trying to get pregnant.
  • The subject is unable or unwilling to provide informed consent, or is unable or unwilling to comply with study procedures.
  • The subject has any other condition, which in the opinion of the investigators' would make them unsuitable for the study.

Treatment and study plan

Pulse oximeter accuracy under hypoxic conditions

Diagnostic Test

This study will specifically evaluate the accuracy of the Owlet pulse oximeter sensor, currently FDA cleared for infants 1-18 months of age in an expanded hypoxia lab trial.

Primary outcomes

  1. Pulse Oximeter Accuracy Measures

    Time frame: From enrollment to completion of 6 plateaus of hypoxia over an approximate 75 minute period.

    • Overall Accuracy: Arms is less than 3%.
    • Non-Disparate Performance Evaluation 1: Among pairwise comparisons of MST groups 1-4, 5-7, and 8-10, the largest difference in SpO2 bias is less than 3.5% for the interval 70% ≤ SaO2≤ 85% and less than 1.5% for 85%< SaO2 ≤ 100%.
    • Non-Disparate Performance Evaluation 2: For a 100-point change in emitter sensor site ITA, the difference in SpO2 bias is less than 3.5% for 70% ≤ SaO2≤ 85% and less than 1.5% for 85%< SaO2 ≤ 100%

Sponsors and collaborators

Lead sponsor

Owlet Baby Care, Inc.

Industry

Registry information

Important dates

Study start
2025
Primary completion
2025
Study completion
2025
First posted
Feb 27, 2025
Registry last updated
Aug 14, 2025

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.