NCT Number: NCT01418079
A cerebral oximeter is a device that uses light to measure the amount of oxygen within the brain. It is similar to the device that measures the level of oxygen in the tip of the finger, known as a pulse oximeter. The cerebral oximeter consists of a sensor placed on the forehead that both emits and detects the amount of light absorbed. This study will determine how accurate the device is by comparing the displayed value on the monitor with blood samples taken simultaneously from the arterial blood in the wrist and venous blood in the neck. In order to test the device over a suitable range, the level of oxygen within the blood will be reduced in a controlled manner by reduction of the inspired oxygen concentration. This is the equivalent of ascending to an altitude of 16,000 feet. The study will be conducted in healthy volunteers.
Looking for future studies?
Notify Me18 year–50 year
All sexes
Interventional
Phase 2
Durham, North Carolina, 27705, United States
This is a calibration and validation study of a near-infrared spectroscopy (NIRS) device designed to measure the cerebral tissue oxygen saturation non-invasively. This is achieved by comparing NIRS-derived cerebral tissue oxygen saturation with a calculated value derived from simultaneous arterial and jugular venous blood samples.
At present the FDA have adopted the standards published in 2005 by the International Organization for Standardization (ISO), entitled ISO 9919. This is a set of technical specifications and guidelines for pulse oximeters, which share certain technical similarities to cerebral oximeters. In particular, Annex EE details the conduct of a controlled desaturation study for the calibration of pulse oximeter equipment. Specifically, the fraction of inspired oxygen delivered to test subjects is varied to achieve a series of targeted steady state saturation periods over a range of arterial oxygen saturation of 70 - 100%.
While cerebral oximeters differ from pulse oximeters in terms of the what is being measured (brain tissue versus arterial blood) the FDA have maintained the requirement to examine data from human volunteer studies in which the arterial oxygen saturation ranges from 70 - 100%. Several FDA-approved cerebral oximeters were validated in a similar manner.
The device controlling the inspired gas concentration is the RespirAct, which permits precise reduction in the arterial oxygen saturation while also maintaining the arterial carbon dioxide level at a precise level.
The study consists of 2 sequences:
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Reduction in blood oxygen saturation by sequential reduction in inspired gas composition. Steps are of 6 minutes duration with reduction in pulse oximeter oxygen saturation from 100 to 70%.
Other names: PMS-3000 cerebral oximeter
Time frame: approximately 4 months
Oximeter measurements are compared to co-oximeter measurements following each subject study. The total sample of subjects is needed to complete the analysis.
O2 MedTech, Inc.
Industry
Accuracy of the O2 MedTech PMS-3000 Monitor for the Non-invasive Measurement of Cerebral and Tissue Oxygenation Using Phase-modulated Infrared Spectroscopy
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT06068257
Breast Cancer, Breast Diseases
Orlando, Florida, United States
View Trial DetailsNCT06702423
Healthy
London, United Kingdom
View Trial DetailsNCT07223164
Healthy
Irvine, California, United States
View Trial DetailsNCT05595902
Healthy
Kadıköy, Istanbul, Turkey (Türkiye)
View Trial Details