Completed

NCT Number: NCT04569240

Accuracy of the Dexcom G6 Continuous Glucose Monitoring System Following Cardiac Surgery

This is a prospective longitudinal study to study the agreement between a novel continuous glucose monitoring system (CGMS) versus current blood glucose monitoring.

Subjects in this study will have their blood glucose measured regularly every 1-3 hours with current methods in the cardiovascular intensive care unit (CVICU), and by point of care glucose using the Accuchek Inform II on the regular floors, and the CGMS at the same time will be captured. Subjects will have measurements taken throughout their stay in the CVICU and on the regular floors.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Cleveland Clinic

Cleveland, Ohio, 44195, United States

About this study

Maintaining tight blood glucose control following surgery is imperative to reducing infections and neurologic dysfunction. This requires frequent blood sampling while in the intensive care unit, leading to increased waste of blood and utilizing time and resources to collect samples and wait for results. Additionally, when on the regular floors, frequent point of care fingersticks for glucose levels are needed for therapeutic intervention.

Continuous glucose monitoring systems (CGMS) are now available for ambulatory use in patients with diabetes. The potential of using such a system in the hospital, including during the postoperative critical care setting and upon transfer to the regular floor, may reduce personnel burden, produce rapid results, and decrease blood waste. However, these systems have not yet been validated for hospital use.

This is a prospective longitudinal study to study the agreement between a novel continuous glucose monitoring system (CGMS) versus current blood glucose monitoring.

Subjects in this study will have their blood glucose measured regularly every 1-3 hours with current methods in the CVICU, and by point of care glucose using the Accuchek Inform II on the regular floors, and the CGMS at the same time will be captured. Subjects will have measurements taken throughout their stay in the CVICU and on the regular floors. Agreement and correlation between systems will be calculated, and error will be classified using the Surveillance Error grid. The proportion of errors of at least moderate risk will be calculated.

Subject characteristics including demographics, comorbidities and baseline disease severity will be summarized using means and standard deviations for continuous measures and frequencies and percentages for categorical factors. To evaluate agreement between methods, concordance correlation coefficients will be calculated, using the method that adjusts for repeated measures as described by King and implemented in software by Carrasco. Analysis will be performed overall, and then stratifying by type of current measure, if the type of measurement varies across settings. Correlations will be calculated using the methods described by Bland and Altman and implemented by Bakdash and Marusich. Differences will also be evaluated for errors according to the Surveillance Error Grid. Rates of moderate or more severe errors will be calculated. For each measure above, 95% confidence intervals will be calculated, accounting for the clustering within subject.. Analyses will be performed using SAS software (version 9.4; Cary, NC) and R software (version 3.5; Vienna Austria).

Glucose readings obtained from the CGMS will be compared with

  • blood glucose from the arterial blood gas (ABG) or peripheral/central venous catheter if arterial catheter not in use (standard of care in CVICU) while patients are in the CVICU
  • fingerstick point of care (POC) glucose when patients are on the regular floors

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Age 18 years and above
  • Planned cardiovascular surgery
  • Planned admission to Cleveland Clinic Main Campus J5 or J6 or Q5 cardiovascular ICU (CVICU) post-operatively
  • With or without known diabetes; if with known diagnosis of diabetes, diabetes can be type 1, type 2, or secondary

Exclusion criteria

  • Allergy to the material of the CGMS or the adhesive to be used
  • Skin conditions precluding the use of the CGMS
  • Pregnancy
  • Other conditions that investigators deem inappropriate for the study

Treatment and study plan

Dexcom G6 continuous glucose monitor

Device

Dexcom G6 continuous glucose monitor will be applied to the upper outer arm before cardiovascular surgery. Data will be collected for 10 days or upon discharge from the ICU.

Primary outcomes

  1. Number of Matching Glucose Measurements (CGM Versus Blood Glucose Measurements)

    Time frame: Up to 10 days

    Glucose readings from the Dexcom continuous glucose monitor compared with blood glucose from the arterial blood gas (ABG) or peripheral/central venous catheter

  2. Glucose Readings From Continuous Glucose Monitor (CGM) Versus Point of Care Glucose

    Time frame: up to 10 days

    Glucose readings in mg/dL from the Dexcom continuous glucose monitor (CGM) will be compared with fingerstick point of care glucose levels in mg/dL

Other outcomes

  1. Blood Volume Waste by Arterial Blood Gas

    Time frame: Up to 10 days

    Blood sample amounts in milliliters required for measurement of glucose by arterial blood gas

  2. Blood Volume Waste by Fingerstick

    Time frame: Up to 10 days

    Blood sample amounts in milliliters required for measurement of glucose by fingerstick

  3. Cost Savings:

    Time frame: Up to 10 days

    Cost of CGMS device including sensor, transmitter versus cost of doing an ABG to obtain glucose levels (CVICU stay), and versus the cost of Accu-chek Inform II meter, test strips, and lancets

  4. Time Savings

    Time frame: Up to 10 days

    time it takes to perform and time it takes to have results based on an ABG or fingerstick POC on Accu-check Inform II

Sponsors and collaborators

Lead sponsor

The Cleveland Clinic

Other

Collaborators

  • DexCom, Inc.

Registry information

Important dates

Study start
2021
Primary completion
2024
Study completion
2024
First posted
Sep 29, 2020
Registry last updated
Nov 10, 2025

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.