Skip to main content
OpenTrials
Completed

NCT Number: NCT05139693

Accuracy of Pulse Oximeters With Profound Hypoxia

The aim of this project is to test the accuracy of pulse oximeters during mild, moderate and severe hypoxia. This is done by comparing the reading of the pulse oximeter during brief, steady state hypoxia with a gold-standard measurement of blood oxyhemoglobin saturation (arterial blood sample processed in a laboratory hemoximeter). The data obtained is submitted by pulse oximeter manufacturers to the FDA for device approval.

Completed

Looking for future studies?

Notify Me

Key information

Conditions

Age range

18 year–49 year

Sex eligibility

All sexes

Study type

Observational

Primary location

Hypoxia Research Laboratory

San Francisco, California, 94143, United States

About this study

The study will include at least 10 subjects (up to 14 if needed to reach the 200 necessary data points to meet the ISO 80601-2-61:2017).

Per FDA guidance, at least 2, or 15% of the subjects will have dark skin.

Who can participate

Healthy volunteers accepted: Yes

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • The subject is male or female, aged ≥18 and <50.
  • The subject is in good general health with no evidence of any medical problems.
  • The subject is fluent in both written and spoken English.
  • The subject has provided informed consent and is willing to comply with the study procedures.

Exclusion criteria

  • The subject is obese (BMI>30).
  • The subject has a known history of heart disease, lung disease, kidney or liver disease.
  • Diagnosis of asthma, sleep apnea, or use of CPAP.
  • Subject has diabetes.
  • Subject has a clotting disorder.
  • The subject a hemoglobinopathy or history of anemia, per subject report or the first blood sample, that in the opinion of the investigator, would make them unsuitable for study participation.
  • The subject has any other serious systemic illness.
  • The subject is a current smoker.
  • Any injury, deformity, or abnormality at the sensor sites that in the opinion of the investigators' would interfere with the sensors working correctly.
  • The subject has a history of fainting or vasovagal response.
  • The subject has a history of sensitivity to local anesthesia.
  • The subject has a diagnosis of Raynaud's disease.
  • The subject has unacceptable collateral circulation based on exam by the investigator (Allen's test).
  • The subject is pregnant, lactating or trying to get pregnant.
  • The subject is unable or unwilling to provide informed consent, or is unable or unwilling to comply with study procedures.
  • The subject has any other condition, which in the opinion of the investigators' would make them unsuitable for the study.

Treatment and study plan

Respiree

Device

The aim of this project is to test the accuracy of pulse oximeters during mild, moderate and severe hypoxia. This is done by comparing the reading of the pulse oximeter during brief, steady state hypoxia with a gold-standard measurement of blood oxyhemoglobin saturation (arterial blood sample processed in a laboratory hemoximeter). The data obtained is submitted by pulse oximeter manufacturers to the FDA for device approval

Primary outcomes

  1. To obtain a reading from the Respiree device that corresponds to the associated blood sample or a reference oximeter.

    Time frame: 1 hour per subject

    The number of subjects and the number of comparisons (paired pulse oximeter readings and arterial saturation values) is determined by current FDA guidance requirements. This is a minimum of 200 data points and 10 subjects. In the course of this type of study, some subjects may drop out, some readings can be lost due to motion or other interference and occasionally some do not consent.

Sponsors and collaborators

Lead sponsor

Respiree Pte Ltd

Industry

Registry information

Important dates

Study start
2021
Primary completion
2021
Study completion
2021
First posted
Dec 1, 2021
Registry last updated
Apr 29, 2022

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.