Stanford University Medical Center
Stanford, California, 94305, United States
NCT Number: NCT02237014
Peripheral pulse oximetry allows continuous non-invasive measurement of arterial oxygen saturation, but the gold standard for arterial oxygen saturation is co-oximeter which requires an arterial blood sample. The purpose of this research study is to determine the accuracy of a pulse oximeter with a standard sensor (Masimo LNCS sensor) versus with the study sensors, namely Masimo blue sensor and Nellcor Max-I sensors and compared against co-oximetry. Currently available peripheral oximeters (standard) are inaccurate at low oxygen saturation noted in children with cyanotic heart disease. Hence therapeutic interventions (including surgery and cardiac catheterizations) based solely on peripheral oximetry can be delayed and or inadequate. By doing this study the investigators will be able to establish correct limits of peripheral pulse oximeter when using the standard and the study sensors.
Looking for future studies?
Notify MeUp to 10 year
All sexes
Observational
Stanford, California, 94305, United States
Peripheral pulse oximetry allows continuous non-invasive measurement of arterial oxygen saturation, but the gold standard for arterial oxygen saturation is co-oximeter which requires an arterial blood sample. The purpose of this research study is to determine the accuracy of a pulse oximeter with a standard sensor (Masimo LNCS sensor) versus with the study sensors, namely Masimo blue sensor and Nellcor Max-I sensors and compared against co-oximetry. Currently available peripheral oximeters (standard) are inaccurate at low oxygen saturation noted in children with cyanotic heart disease. Hence therapeutic interventions (including surgery and cardiac catheterizations) based solely on peripheral oximetry can be delayed and or inadequate. By doing this study we will be able to establish correct limits of peripheral pulse oximeter when using the standard and the study sensors.
The investigator hopes to learn the limits of accuracy of currently available and used pulse oximeters. In children with cyanotic heart disease the "blue sensor" has been found in small studies to be more accurate compared to the "standard" pulse oximeter.
This study is important as it will provide information as to which pulse oximeter should be routinely used in children with cyanotic heart disease and to assess which SPO2 even with blue sensor is borderline and therefore the physician will know to obtain arterial blood sample for co-oximeter prior to planning important procedures based on a saturation reading.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Time frame: 2 year
The primary outcome measure is to describe the bias and precision between the Masimo blue sensor and co-oximetry.
Time frame: 2year
The secondary measure was to describe the limits of the blue sensor and standard sensor accuracy as compared with the co-oximeter. If the test sensors consistently have a bias greater than 2 SD from the co-oximeter value at saturations lower than eg., 80 then the recommendation would be that the co-oximetry be used to measure patients saturatioin rather than relying on pulse oximtery alone for clinical decision making
Stanford University
Other
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.