Accuracy Assessment of an Automatic Blood Pressure Measurement Device in Adult Volunteers

The research team has developed an automatic blood pressure monitor (Sphygmo) to be used for the monitoring and diagnosis of pre-eclampsia in pregnant women, particularly in low-resource settings where current monitoring is limited.

90 adult volunteers will be enrolled by researchers at Rice University. The participant will be seated in a comfortable chair with arm at heart level. Arm circumference will be measured and a blood pressure cuff will be placed on the arm. The cuff will be inflated and blood pressure measurements will be taken by a commercially available device and by the Sphygmo device. Blood pressure measurements from both devices will be recorded. The participant's blood pressure will be measured up to 9 times with a waiting period of 45-60 seconds between each measurement.

The results of this study will be used to optimize the blood pressure detection algorithm and thus further develop the device.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Rice University

Houston, Texas, 77030, United States

Who can participate

Healthy volunteers accepted: Yes

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • 18 years of age or older
  • Willing and able to provide informed consent

Exclusion criteria

  • Unable to provide informed consent

Treatment and study plan

Sphygmo: Automatic Blood Pressure Monitor

Device

A team of engineers from Rice University has recently developed Sphygmo, an ambulatory, low-cost blood pressure monitor for use in the diagnosis and management of pre-eclampsia in low-resource hospitals.

GE Dinamap ProCare Automatic Blood Pressure Monitor

Device

This commercially available gold standard blood pressure monitor will be used as a control to assess the accuracy of the Sphygmo device.

Primary outcomes

  1. Accuracy of Blood Pressure Measurement by Sphygmo Relative to the Clinical Standard.

    Time frame: Blood pressure was measured an average of 9 times for each participant during their single visit and the mean of these measurements was recorded and used for analysis. The measurement period lasted approximately 30-45 minutes.

    Mean difference of systolic and diastolic blood pressure readings between Sphygmo device and the gold standard sphygmomanometer.

  2. Percentage of Sphygmo Readings That Were With < 5 mmHg, < 10 mmHg, and < 15 mmHg of the Readings by the Gold Standard Sphygmomanometer.

    Time frame: Blood pressure was measured an average of 9 times for each participant during their single visit and the mean of these measurements was recorded and used for analysis. The measurement period lasted approximately 30-45 minutes.

    The British Hypertension society defines a specific criteria for the accuracy of a sphygmomanometer. Devices are graded according to the cumulative percentage of readings that have an absolute difference between the more favorable observer's mercury sphygmomanometer readings and the test device of < 5 mmHg, < 10 mmHg, and < 15 mmHg. A letter grade of "A" requires that over 60%, 85%, and 95% were achieved in the < 5 mmHg, < 10 mmHg, and <15 mmHg categories, respectively. A letter grade of "B" requires that over 50%, 75%, and 90% were achieved in the < 5 mmHg, < 10 mmHg, and <15 mmHg categories, respectively. The test device achieved a grade of (B/B) with the validation data from this study.

Sponsors and collaborators

Lead sponsor

William Marsh Rice University

Other

Registry information

Important dates

Study start
2015
Primary completion
2015
Study completion
2017
First posted
Oct 17, 2014
Registry last updated
Mar 6, 2020

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.