Skip to main content
OpenTrials
Completed

NCT Number: NCT05806554

Accuracy and Precision of the True Vie I3 Continuous Glucose Monitoring System: an Open Label, Multi-Center Trial

The purpose of the investigation is to evaluate the performance of the I3 CGM according to the FDA's special controls for iCGM.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year–80 year

Sex eligibility

All sexes

Study type

Observational

Primary location

PSHI Praxis GmbH, Mainz, Germany

Loading trial locations.

About this study

Participants will attend a screening visit, 4 clinic visits with Frequent Sampling Testing (FST) at the beginning, 2 in the middle, and end of the I3 CGM sensor wear, and a visit for removal of the I3 CGM. Participants will wear 3 I3 CGM devices for up to 15 days and be instructed to continue their normal daily activities and manage their glucose as usual between in clinic FST per routine care. Participants will not have access to the I3 CGM data for the management of their blood glucose between clinic visits and will continue their usual care including CGM, if they are CGM users.

At the clinic FST visits venous blood samples will be drawn from an intravenous cannula for FST for up to 10 hours for plasma glucose determination. During the clinic visits the investigative site staff will determine the participants' insulin doses and meals and may induce both hypoglycemia and hyperglycemia of up to an hour duration.

Who can participate

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • 1. Patients with T1DM or T2DM diagnosed for at least 6 months. 2. Individual is 18-80 years old, inclusive, at the time of screening. 3. Must be and have been in stable treatment regimen for at least 1 month with a multiple daily insulin dosing regimen (at least two insulin doses with meals daily) or CSII using at least two bolus doses a day with meals, irrespective of delivery device(s).

Exclusion criteria

  • 1. History of skin adhesive tolerance issues in the area of sensor placement. 2. HbA1c > 9%. 3. Insulin meal dosing based on fixed dose regimens. 4. Absence of established correction factor for high glucose. 5. Hematocrit below 10% under the lower limit of the normal range. 6. Body mass index < 18.5 kg/m2. 7. Inadequate intravenous access on arms. 8. Participant has had a hypoglycemic seizure within the past 6 months prior to enrollment.
  • Participant has had an episode of diabetic ketoacidosis (DKA) within the past 6 months prior to enrollment.
  • Participant has a history of a seizure disorder. 11. Pregnancy, planned pregnancy within the study period, or unwillingness to use reliable contraception during the study period.
  • Planned MRI, CT scan or diathermia procedure for the duration of the study.

Treatment and study plan

Continuous glucose monitoring device

Device

Continuous glucose monitoring device

Primary outcomes

  1. Accuracy

    Time frame: 15 days

    percentage of iCGM values in the measuring range less than 70 mg/dL (3.9 mmol/l) within +/- 15 mg/dL (0.83 mmol/l) of the

Sponsors and collaborators

Lead sponsor

Sinocare

Industry

Collaborators

  • Integrated Medical Development

Registry information

Important dates

Study start
2023
Primary completion
2024
Study completion
2024
First posted
Apr 10, 2023
Registry last updated
Nov 25, 2025

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.