NCT Number: NCT02041702

Accent Cardiac MRI Study

The aim of this study was to assess safety and efficacy of SJM MRI conditional pacing system. The study hypothesized that cardiac MRI scan would not affect performance of SJM MRI conditional pacing system.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Zhejiang Greentown Cardiovascular Disease Hospital, Hangzhou, China

Loading trial locations.

About this study

Accent Cardiac MRI study was a prospective, randomized ,multi-center study conducted in Asia. Subjects indicated for dual-chamber pacemaker and who signed study consent form were considered enrolled and post implant were randomized to either Cardiac MRI Scan Group or Control Group.

Clinical evaluation and device interrogation were performed at baseline, pre- and post-MRI scan, and one month post-MRI for all subjects. At 9-12 weeks post-implant, patients in Cardiac MRI Scan Group underwent an elective non-diagnostic cardiac MRI scan at 1.5 Tesla (T) while patients in Control group underwent device interrogation and clinical evaluation twice with a 45-minute waiting period in between. The safety endpoint was freedom from MRI-related complications while efficacy endpoint was change in right atrial/ ventricular (RA/RV) capture threshold and sensing amplitude from pre-MRI scan to one month post-MRI scan.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Have an approved indication per American College of Cardiology(ACC)/ American Heart Association(AHA)/ Heart Rhythm Society(HRS)/ European Society of Cardiology(ESC) guidelines for implantation of a pacemaker
  • Will receive a new dual chamber pacemaker and leads
  • Be willing to undergo an elective MRI scan without complete sedation, that is ,under consciousness during MRI scan
  • Be able to provide informed consent for study participation
  • Be willing and able to comply with the prescribed follow-up tests and schedule of evaluations
  • Is not contraindicated for an MRI scan (per the pre-MRI safety screening form)
  • Age≥18 years
  • Subjects must be able to undergo pectoral implantation

Exclusion criteria

  • Having an existing pacemaker or Implantable Cardioverter Defibrillator (ICD) .A new pacemaker and lead is required for enrollment
  • Permanent Atrial Fibrillation/Flutter
  • Are medically indicated for an MRI scan at the time of enrollment
  • Have an existing active implanted medical device, e.g. neurostimulator, infusion pump etc
  • Have a non-MRI compatible device or material implanted
  • Have a lead extender or adaptor
  • Be unable to fit in MRI bore; will come into contact with the magnet façade inside the MRI bore
  • Are currently participating in a clinical investigation that includes an active treatment arm
  • Are allergic to dexamethasone sodium phosphate(DSP)
  • Are pregnant or planning to become pregnant during the duration of the study
  • Have a life expectancy of less than 12 months due to any condition
  • Subjects with exclusion criteria required by local law (e.g. age)
  • Are unable to comply with the follow up schedule

Treatment and study plan

Non-Diagnostic Cardiac MRI scan

Diagnostic Test

Primary outcomes

  1. Freedom From MRI Scan-related Complications

    Time frame: MRI Visit ,1 Month Post MRI Visit

    Number of subjects who were free from MRI scan-related complications

  2. Change in Right Atrial (RA) Capture Threshold @0.5 Millisecond (ms) From Pre-MRI Scan to 1 Month After MRI Scan

    Time frame: MRI Visit ,1 Month Post MRI Visit

    Number of subjects who experienced an increase in RA capture threshold @ 0.5 ms at 1-Month post MRI scan visit ≤ 0.5 Volt (V) compared to pre-MRI scan value collected at MRI scan visit

  3. Change in Right Ventricular (RV) Capture Threshold @0.5ms From Pre-MRI Scan to 1 Month After MRI Scan

    Time frame: MRI Visit, 1 Month Post MRI Visit

    Number of subjects who experienced an increase in RV capture threshold @ 0.5 ms at 1-Month post MRI scan visit ≤ 0.5 V compared to pre-MRI scan value collected at MRI scan visit

  4. Change in RA Sensing Amplitude From Pre-MRI Scan to 1 Month After MRI Scan

    Time frame: MRI Visit, 1 Month Post MRI Visit

    Number of subjects who experienced a decrease in RA sensing amplitude ≤ 50% and at least 1.5 mV at 1-Month post MRI scan visit compared to pre-MRI scan value collected at MRI scan visit

  5. Change in RV Sensing Amplitude From Pre-MRI Scan to 1 Month After MRI Scan

    Time frame: MRI Visit, 1 Month Post MRI Visit

    Number of subjects who experienced a decrease in RV sensing amplitude ≤ 50% and at least 5 mV at 1-Month post MRI scan visit compared to pre-MRI scan value collected at MRI scan visit

Sponsors and collaborators

Lead sponsor

Abbott Medical Devices

Industry

Registry information

Official study title

A Clinical Evaluation of the Safety and Efficacy of St. Jude Medical (SJM) Magnetic Resonance Imaging (MRI) Conditional Pacing System in Patients Undergoing Cardiac MRI

Acronym: Accent Cardiac

Important dates

Study start
2014
Primary completion
2016
Study completion
2016
First posted
Jan 22, 2014
Registry last updated
Aug 21, 2019