NCT Number: NCT02041702
Accent Cardiac MRI Study
The aim of this study was to assess safety and efficacy of SJM MRI conditional pacing system. The study hypothesized that cardiac MRI scan would not affect performance of SJM MRI conditional pacing system.
Looking for future studies?
Notify MeKey information
Conditions
Age range
18 year and older
Sex eligibility
All sexes
Study type
Interventional
Phase
Not applicable
Primary location
Zhejiang Greentown Cardiovascular Disease Hospital, Hangzhou, China
About this study
Accent Cardiac MRI study was a prospective, randomized ,multi-center study conducted in Asia. Subjects indicated for dual-chamber pacemaker and who signed study consent form were considered enrolled and post implant were randomized to either Cardiac MRI Scan Group or Control Group.
Clinical evaluation and device interrogation were performed at baseline, pre- and post-MRI scan, and one month post-MRI for all subjects. At 9-12 weeks post-implant, patients in Cardiac MRI Scan Group underwent an elective non-diagnostic cardiac MRI scan at 1.5 Tesla (T) while patients in Control group underwent device interrogation and clinical evaluation twice with a 45-minute waiting period in between. The safety endpoint was freedom from MRI-related complications while efficacy endpoint was change in right atrial/ ventricular (RA/RV) capture threshold and sensing amplitude from pre-MRI scan to one month post-MRI scan.
Who can participate
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
- Have an approved indication per American College of Cardiology(ACC)/ American Heart Association(AHA)/ Heart Rhythm Society(HRS)/ European Society of Cardiology(ESC) guidelines for implantation of a pacemaker
- Will receive a new dual chamber pacemaker and leads
- Be willing to undergo an elective MRI scan without complete sedation, that is ,under consciousness during MRI scan
- Be able to provide informed consent for study participation
- Be willing and able to comply with the prescribed follow-up tests and schedule of evaluations
- Is not contraindicated for an MRI scan (per the pre-MRI safety screening form)
- Age≥18 years
- Subjects must be able to undergo pectoral implantation
Exclusion criteria
- Having an existing pacemaker or Implantable Cardioverter Defibrillator (ICD) .A new pacemaker and lead is required for enrollment
- Permanent Atrial Fibrillation/Flutter
- Are medically indicated for an MRI scan at the time of enrollment
- Have an existing active implanted medical device, e.g. neurostimulator, infusion pump etc
- Have a non-MRI compatible device or material implanted
- Have a lead extender or adaptor
- Be unable to fit in MRI bore; will come into contact with the magnet façade inside the MRI bore
- Are currently participating in a clinical investigation that includes an active treatment arm
- Are allergic to dexamethasone sodium phosphate(DSP)
- Are pregnant or planning to become pregnant during the duration of the study
- Have a life expectancy of less than 12 months due to any condition
- Subjects with exclusion criteria required by local law (e.g. age)
- Are unable to comply with the follow up schedule
Treatment and study plan
Primary outcomes
-
Freedom From MRI Scan-related Complications
Time frame: MRI Visit ,1 Month Post MRI Visit
Number of subjects who were free from MRI scan-related complications
-
Change in Right Atrial (RA) Capture Threshold @0.5 Millisecond (ms) From Pre-MRI Scan to 1 Month After MRI Scan
Time frame: MRI Visit ,1 Month Post MRI Visit
Number of subjects who experienced an increase in RA capture threshold @ 0.5 ms at 1-Month post MRI scan visit ≤ 0.5 Volt (V) compared to pre-MRI scan value collected at MRI scan visit
-
Change in Right Ventricular (RV) Capture Threshold @0.5ms From Pre-MRI Scan to 1 Month After MRI Scan
Time frame: MRI Visit, 1 Month Post MRI Visit
Number of subjects who experienced an increase in RV capture threshold @ 0.5 ms at 1-Month post MRI scan visit ≤ 0.5 V compared to pre-MRI scan value collected at MRI scan visit
-
Change in RA Sensing Amplitude From Pre-MRI Scan to 1 Month After MRI Scan
Time frame: MRI Visit, 1 Month Post MRI Visit
Number of subjects who experienced a decrease in RA sensing amplitude ≤ 50% and at least 1.5 mV at 1-Month post MRI scan visit compared to pre-MRI scan value collected at MRI scan visit
-
Change in RV Sensing Amplitude From Pre-MRI Scan to 1 Month After MRI Scan
Time frame: MRI Visit, 1 Month Post MRI Visit
Number of subjects who experienced a decrease in RV sensing amplitude ≤ 50% and at least 5 mV at 1-Month post MRI scan visit compared to pre-MRI scan value collected at MRI scan visit
Sponsors and collaborators
Lead sponsor
Abbott Medical Devices
Industry
Registry information
Official study title
A Clinical Evaluation of the Safety and Efficacy of St. Jude Medical (SJM) Magnetic Resonance Imaging (MRI) Conditional Pacing System in Patients Undergoing Cardiac MRI
Acronym: Accent Cardiac
Important dates
- Study start
- 2014
- Primary completion
- 2016
- Study completion
- 2016
- First posted
- Jan 22, 2014
- Registry last updated
- Aug 21, 2019
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Related clinical trials
Published trials that share one or more normalized conditions with this study.
BIOTRONIK Conduction System Pacing With the Solia Lead - Solia CSP S
NCT06540079
Arrhythmia, Sinus, Arrhythmias, Cardiac
Phoenix, Arizona, United States
View Trial DetailsPV Loops With CNT in Multiple Pacing Sites
NCT05345925
A-V Block, Arrhythmias, Cardiac
Krakow, Poland
View Trial DetailsPreload and Autonomic Responses With Cardiac Neuromodulation Therapy (CNT) (CS-06)
NCT04463121
Arrhythmias, Cardiac, Bradycardia
Prague, Czechia
View Trial DetailsImpact of Swaddle Blanket on Gastroesophageal Reflux
NCT02083952
Apnea, Arrhythmias, Cardiac
Ridgewood, New Jersey, United States
View Trial Details