Completed

NCT Number: NCT04245345

Accelerometer Sensing for Micra AV Study

The purpose of the AccelAV Study is to characterize chronic AV synchrony in subjects implanted with MicraTM AV device. This study will be conducted upon market approval of the MicraTM AV Transcatheter Pacing System.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Queen Mary Hospital, Hong Kong, HK

Loading trial locations.

About this study

The Accel AV study is a prospective, single-arm, global, multi-center clinical study to characterize the chronic AV synchrony in subjects implanted with the market released Micra AV system. The study is planned to be conducted in the US and Hong Kong. Overall, the study is expected to be conducted at approximately 20 centers.The expected total study duration is approximately 12-15 months, representing the time necessary to enroll the target sample size and to complete the 3-month follow-up visit.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

Subject will be implanted with a MicraTM (Model MC1AVR1) for an approved indication.

  • Subject has history of AV block.
  • Subject is ≥ 18 years old and as per required local law.
  • Subject (and/or witness as applicable per local regulations) provides signed and dated authorization and/or consent per institution and local requirements.
  • Subject is willing and able to comply with the protocol.

Exclusion criteria

  • Subject currently enrolled or planning to participate in a potentially confounding drug or device trial during the study. Co-enrollment in concurrent trials is only allowed when documented pre-approval is obtained from the Medtronic Clinical Research Specialist.
  • Subject implanted with a MicraTM (Model MC1AVR1) on a non-permanent basis (e.g. CIED infection).
  • Subject is pregnant (if required by local law, women of child-bearing potential must undergo a pregnancy test within seven days prior to MicraTM Model MC1AVR1 implant procedures).
  • Subject meets any exclusion criteria required by local law (age or other).

Treatment and study plan

Accelerometer Sensing for Micra AV Study

Device

Characterize AV synchrony in subjects implanted with Micra AV device. The study will be conducted upon market approval of the Micra AV Transcatheter Pacing System.

Primary outcomes

  1. Atrioventricular (AV) Synchrony During Rest at 1-month Post-implant in Subjects With Persistent 3rd Degree Atrioventrcular Block (AVB) and Normal Sinus Node Function

    Time frame: 1-month post-implant

    Atrioventricular synchrony is defined for each P-wave during the 20-minute resting period. Specifically, for each electrocardiogram (ECG) confirmed P-wave that occurs during the 20-minute resting period at the 1-month visit, the endpoint will be considered met if a ventricular beat (ventricular pace or sensed event) is within 300 ms following an ECG confirmed P-wave. The reported percentage is the average percentage of P-waves considered synchronous across all patients during the 20-minute resting period.

Secondary outcomes

  1. Stability of AV Synchrony During Rest Between 1-month and 3-months Post-implant in Subjects With Persistent 3rd Degree AVB and Normal Sinus Node Function

    Time frame: Between 1-month and 3-months post-implant

    AV synchrony is defined for each ECG confirmed P-wave that occurs during the 20-minute resting period at both the 1-month visit and 3-months visits. For each P-wave the endpoint will be considered met if a ventricular beat (ventricular pace or sensed event) is within 300 ms following an ECG confirmed P-wave. The reported percentages are the average percentage of P-waves considered synchronous across all patients during the 20-minute resting period at each visit (month 1 or month 3).

  2. Percentage of Ambulatory AV Synchrony at 1-month Post-implant in Subjects With Persistent 3rd Degree AVB and Normal Sinus Node Function

    Time frame: 1-month post-implant

    AV synchrony is defined for each P-wave during the the 24-hour Holter monitoring period while the subject went about their activities of daily living. Specifically, for each electrocardiogram (ECG) confirmed P-wave that occurs during the 24-hour Holter period at the 1-month visit, the endpoint will be considered met if a ventricular beat (ventricular pace or sensed event) is within 300 ms following an ECG confirmed P-wave. The reported percentage is the average percentage of P-waves considered synchronous across all patients during the 24-hour Holter monitoring period.

  3. Left Ventricular Outflow Tract Velocity Time Integral

    Time frame: Within 48 hours of procedure

    Characterize the change in stroke volume, as measured by left ventricular outflow tract (LVOT) velocity time Integral (VTI), during Micra AV mediated VDD (atrioventricular synchronous ventricular pacing) pacing and VVI (asynchronous ventricular pacing) pacing in subjects with persistent 3rd degree AV block and normal sinus node function

Sponsors and collaborators

Lead sponsor

Medtronic Cardiac Rhythm and Heart Failure

Industry

Registry information

Acronym: AccelAV

Important dates

Study start
2020
Primary completion
2022
Study completion
2022
First posted
Jan 28, 2020
Registry last updated
Feb 23, 2023