A Unique Device for Independent Lung Ventilation

Adequate blood oxygenation and ventilation/perfusion matching should be the main goal of anaesthetic and intensive care management. At present, the sole method of improving gas exchange restricted by ventilation/perfusion mismatching is independent ventilation with two ventilators. Recently, however, a unique device has been developed, enabling ventilation of independent lungs in 1:1, 2:1, 3:1, and 5:1 proportions. The main goal of the study was to evaluate the device's utility, precision and impact on biomechanical lung parameters during thoracic surgery under general anaesthesia with double lumen tube intubation. Secondly- to measure the gas distribution in supine and lateral decubitus position.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

About this study

Anaesthetic management:

One hour before thoracic surgery, the patient-subjects received diazepam 0.15 mg kg-1 as premedication. After arriving at the operating theatre, standard monitoring system practices was applied, including heart rate (HR), systolic arterial pressure (SAP), diastolic arterial pressure (DAP), mean arterial pressure (MAP), and pulse oximetry (SpO2). Moreover, an intra-vein cannula was placed, and the infusion of multi-electrolytic fluid 5-10 ml kg-1 h-1 was started. After pre-oxygenation, atropine 0.5 mg and fentanyl 3 µg kg-1 were given, and the induction of anaesthesia with thiopentone 5-7 mg kg-1 was started. Suxamethonium was administered for neuromuscular blockade, and bronchial intubation with a Robertshow double lumen tube was performed. The left bronchus was intubated for right lung surgery, and the right bronchus was intubated for left lung surgery. Tube placement was checked via auscultation and fiberscope. Anaesthesia was maintained with sevoflurane, additional fentanyl doses were used if needed, and neuromuscular blockade was obtained with vecuronium 0.1 mg kg-1. Additionally, 0.1 mg kg-1 dose of morphine was given subcutaneously for postoperative pain control. Patients were also ventilated with O2 (oxygen) and AIR (air) mixture, using the following settings: volume control intermittent positive pressure ventilation, FiO2 (oxygen fraction) 0.4, Vt (tidal volume) 6-10 ml/kg, and f (frequency) 12-15 /min. Furthermore, end tidal CO2 (carbon dioxide) was monitored due to normocapnia maintenance (4.0-5.3 kPa). At the end of surgery, intercostal blockade was brought about with 0.5% bupivacaine, with 5 ml for each nerve. Finally, the neuromuscular blockade was reversed with neostigmine 0.04 mg kg-1 and with atropine 0.01 mg kg-1.

After anaesthesia stabilisation, the unique control system was used, called the 'tidal volume divider'. This device was placed between the anaesthetic machine and the double lumen tube of the patient. This control system enables conventional ventilation (without any intervention from the control system, as the settings are defined on the ventilator), as well as independent ventilation with division of the tidal volume between the lungs in proportions of 1:1, 2:1, 3:1, and 5:1. It also enables PEEP (selective positive end-expiratory pressure) application to each lung. With independent ventilation, settings such as frequency, tidal volume and inspiration time are defined by ventilator, and this system only controls the direction of tidal volume to each lung (as the control system is a flow divider) by using a differential pneumatic resistor. It also enables dependent and non-dependent lung ventilation in the lateral decubitus positions. Furthermore, the system monitors the expired volume, airway pressure and dynamic compliance of each lung. This device was described and tested on mechanical lung models with a variety of lung model mechanics (compliance, resistance) and ventilation parameters (frequency, tidal volume). It was also tested clinically for its safety during our previous study. The system was invented, developed and patented by a group of Polish engineers from the Nałęcz Institute of Biocybernetics and Biomedical Engineering, the Polish Academy of Sciences, Poland.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • ASA I and II patients who underwent elective thoracic surgery under general anaesthesia, with double lumen tube intubation and one lung ventilation during surgical procedure.

Exclusion criteria

  • All patients with asthma or chronic obstructive pulmonary disease, history of thoracotomy, assessed as ASA III, with difficult airway conditions, kyphoscoliosis or other alterations of chest wall or severe obesity were excluded

Treatment and study plan

independent ventilation

Other

comparison conventional and independent 1:1 ventilation

Primary outcomes

  1. Tidal volume distribution during conventional and independent in 1:1 proportion ventilation in supine, left and right decubitus positions.

    Time frame: up to 12 months

    tidal volume

Secondary outcomes

  1. Peak respiratory pressure measuring during conventional and independent in 1:1 proportion ventilation in supine, left and right decubitus positions.

    Time frame: up to 12 months

    peak respiratory pressure

  2. Dynamic compliance measuring during conventional and independent in 1:1 proportion ventilation in supine, left and right decubitus positions.

    Time frame: up to 12 months

    dynamic compliance

Other outcomes

  1. Impact of supine to left and supine to right decubitus position changes on tidal volume distribution during conventional and independent in 1:1 proportion ventilation.

    Time frame: up to 12 months

    tidal volume

  2. Impact of supine to left and supine to right decubitus position changes on peak respiratory pressure, during conventional and independent in 1:1 proportion ventilation.

    Time frame: up to 12 months

    peak respiratory pressure

  3. Impact of supine to left and supine to right decubitus position changes on dynamic compliance during conventional and independent in 1:1 proportion ventilation.

    Time frame: up to 12 months

    dynamic compliance

Sponsors and collaborators

Lead sponsor

Medical University of Lublin

Other

Registry information

Official study title

1:1 Proportion Ventilation With a Unique Device for Independent Lung Ventilation Using a Double-Lumen Tube in the Supine and Lateral Decubitus Positions. A Pilot Study

Important dates

Study start
2009
Primary completion
2009
Study completion
2010
First posted
Jun 1, 2016
Registry last updated
Jul 12, 2018

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.