Guardian™ Sensor (3)
DeviceUse two Guardian™ Sensor (3)s each connected to a Guardian™ Connect transmitter for approximately 7 days.
NCT Number: NCT03710083
The purpose of this study is to demonstrate the performance and safety of the Guardian™ Sensor (3) over 170 hours (7 days) when inserted in the abdomen used in subjects age 14 - 75 years with the Guardian™ Connect System.
Looking for future studies?
Notify Me14 year–75 year
All sexes
Interventional
Not applicable
The Seventh Medical Center of PLA General Hospital, Beijing, China
This study is a multi-center, prospective single-arm design without controls. Up to 72 subjects will be enrolled in order to have approximately 60 subjects complete the study. Three investigational centers in China will be used during this study.
During the study, each subject will be randomly assigned to one day of the Yellow Springs Instrument (YSI™*) frequent sample testing (FST) (Day 1, 3-5, or 7).Subjects will wear two Guardian™ Sensor (3)s each connected to a Guardian™ Connect transmitter for approximately 7 days (one which will be paired to the Guardian Connect app and the other will function as a glucose recorder.).
The Guardian Sensor (3)s will be worn in the abdomen area and self-inserted by the subject on same side or opposite sides.
On the evening prior to FST, subjects will be asked to fast for approximately 12 hours and adjust their insulin and medications according to routine care (for example as they would do for fasting lipid panel). Subjects may fast for shorter period of time based on investigator discretion.
The subject should be in fasting status upon arrival hospital to start FST process. The feeding protocol may be modified based on investigator discretion. The duration of FST will be approximately 7 hours.
During the study, subjects will continue with their current diabetes regimen independent of the study devices. Subjects will be instructed by the investigational center that they are not to use the study devices (except for the CONTOUR™* study meter) for the management of their diabetes. The CONTOUR™* study meter may be used for treatment decisions and calibration of Guardian™ Sensor (3).
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Use two Guardian™ Sensor (3)s each connected to a Guardian™ Connect transmitter for approximately 7 days.
Time frame: 168 Hours
Percentage of Guardian™ Sensor (3) Values that are within 20% agreement of Yellow Springs Instrument (YSI) plasma glucose values (±20 mg/dL (1.1 mmol/L) when reference blood glucose (YSI) less than or equal to (≤) 80 mg/dL (4.4 mmol/L) ) during YSI frequent sampling testing (FST) days.
Time frame: 168 Hours
A mean rate in Zone A+B of Consensus Error Grid between Guardian Sensor (3) values and YSI™* plasma glucose values during YSI™* FST days was evaluated. Consensus Error Grid (or Parkes error grid) compared the paired primary sensor and YSI reference glucose values. Zone A is defined in the Parkes error grid as the zone of "clinical accurate measurements with no effect on clinical action", Zone B as "altered clinical action with little or no effect on clinical outcome". Ideal situation is 100% in Zone A + B.
Time frame: 168 Hours
A mean rate in Zone A+B of Clarke Error Grid between Guardian Sensor (3) values and YSI™* plasma glucose values during YSI™* FST days was evaluated. Clarke Error Grid compared the paired primary sensor and YSI reference glucose values. Zone A are "clinically accurate in that they would lead to clinically correct treatment decisions". Zone B "would lead to benign or no treatment". Ideal situation is 100% in Zone A + B.
Time frame: 168 Hours
Mean absolute relative difference (MARD) between Guardian Sensor (3) values and YSI™* plasma glucose values during YSI™* FST days was evaluated. MARD = Mean of ((Absolute difference of YSI reference and Sensor glucose values / YSI reference glucose values) * 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple FST days were pooled together for reporting purpose.
Medtronic MiniMed, Inc.
Industry
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT00038727
Body Weight, CVD
Rockville, Maryland, United States
View Trial DetailsNCT06029517
Arterial Occlusive Diseases, Arteriosclerosis
Buffalo, New York, United States
View Trial DetailsNCT01934660
Cardiovascular Disease, Cardiovascular Diseases
Bethesda, Maryland, United States
View Trial DetailsNCT07728162
Diabetes Mellitus, Disease Attributes
Bolu, Turkey (Türkiye)
View Trial Details