Chenla Children's Healthcare
Kratie, Kratié Province, 10109, Cambodia
NCT Number: NCT04311320
Purpose: Using a novel oxygen blender prototype with modified bCPAP to manage children 1 month to 5 years of age hospitalized with respiratory distress due to lower respiratory tract infection (LRTI) in low-resource settings is not associated with clinical failure within 1 hour of use.
Aim: To investigate if this novel oxygen blender is safe to use among children 1 month to 5 years old hospitalized for respiratory distress from lower respiratory tract infections in a low-resource setting.
Looking for future studies?
Notify Me0 day–5 year
All sexes
Interventional
Not applicable
Kratie, Kratié Province, 10109, Cambodia
The study will be a prospective cohort feasibility and safety trial. Patients admitted to Chenla Children's Healthcare who meet eligibility criteria will be asked for consent and then enrolled. Once enrolled, the participant will be treated per hospital protocols and stabilized on bCPAP respiratory support. Once stable for 2 hours, the blender will be incorporated into the bCPAP circuit. At this point, the study nurses will begin recording data for the study as well as providing standard medical care per hospital protocols.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
The device that is being evaluated is a novel low-cost, high access oxygen blender designed to be used in the World Health Organization bCPAP design. It functions by utilizing the Venturi effect, whereby a narrow jet of one fluid (i.e. oxygen) creates a suction effect and entrains surrounding ambient fluid (i.e. room air) into the jet, effectively creating a mixed solution of both fluids. The cost of materials amounts to approximately 1.40 U.S.D. The average time it takes to build a blender and incorporate it into a bCPAP circuit for a new user is approximately 16 minutes. The materials can all be found in a standard hospital, namely 3cc syringes with rubber stops and a 22 gauge hypodermic needle.
Time frame: 1 hour
The primary outcome of this study is the feasibility of the investigational device as measured by clinical failure within 1 hour of installing the blender in the bubble continuous positive airway pressure (bCPAP) circuit. Clinical failure is defined as the occurrence of any of the following patient events: intubation, death, or transfer for higher level of care, or any 2 of the following events; <85% O2 saturation after 30 min of treatment, new signs of respiratory distress, or partial pressure of carbon dioxide greater than 60 mmHg and pH less than 7.2 in capillary blood gas.
Outcome will be reported as the percent of participants who experience clinical failure within 1 hour after blender installation.
Time frame: Approximately 7 days
Oxygen saturation will be measured with pulse oximeter every 6 hours for the length of hospital submission. Outcome will be reported as mean of all measurements per patient. Units are percent.
Time frame: Approximately 7 days
Oxygen concentration of outflow from nasal cannula will be measured with an oxygen analyzer every 6 hours for the length of hospital submission. Outcome will be reported as mean of all measurements per patient. Units are percent oxygen.
Time frame: Approximately 7 days
Flow of oxygen from tank will be measured with the gauge on the oxygen canister. Measurements will be recorded every 6 hours for the length of hospital submission. Outcome will be reported as mean of all measurements per patient in units of liters per minute.
Time frame: Approximately 7 days
CPAP level (as displayed on CPAP device) will be recorded every 6 hours for the length of hospital submission. Outcome will be reported as mean of all measurements per patient. This outcome measure is reported in units of cm of H2O.
Time frame: Approximately 7 days
bCPAP maintenance is defined as the number of times no bubbling in the water column is noted by study staff indicating the need for maintenance. Need for maintenance will be assessed every 2 hours for the length of hospital stay. Number of incidents per patient will be recorded and outcome will be reported as a mean value of all measurements per patient.
Time frame: Approximately 7 days
Blender maintenance is defined as the number of times the blender needs to be repaired or replaced per patient. Need for maintenance will be assessed every 2 hours for the length of hospital stay. Number of incidents per patient will be recorded and outcome will be reported as mean value of all measurements per patient.
Time frame: Approximately 7 days
Outcome will be reported as the length of time (in hours) that respiratory support (i.e. requiring bCPAP and blender) is required during hospital submission.
Time frame: Approximately 7 days
Outcome is reported as number of patients who require transfer to a higher level of care.
Time frame: Approximately 7 days
Outcome is reported as number of patients requiring intubation.
University of Minnesota
Other
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT05987800
Bronchopulmonary Dysplasia, Congenital, Hereditary, and Neonatal Diseases and Abnormalities
Jette, Brussels Capital, Belgium
View Trial DetailsNCT06992492
Lung Diseases, Respiration Disorders
Banī Suwayf, Egypt
View Trial DetailsNCT06168149
Female Urogenital Diseases and Pregnancy Complications, Lung Diseases
Ankara, Yenimahalle, Turkey (Türkiye)
View Trial DetailsNCT05967624
Lung Diseases, Respiration Disorders
Boston, Massachusetts, United States
View Trial Details